摘要
High-resolution elastographic assessment of the cornea can greatly assist clinical diagnosis and treatment of various ocular diseases. Here, we report on the first noncontact depth-resolved micro-scale optical coherence elastography of the cornea achieved using shear wave imaging optical coherence tomography (SWI-OCT) combined with the spectral analysis of the corneal Lamb wave propagation. This imaging method relies on a focused air-puff device to load the cornea with highly-localized low-pressure short-duration air stream and applies phase-resolved OCT detection to capture the low-amplitude deformation with nano-scale sensitivity. The SWI-OCT system is used here to image the corneal Lamb wave propagation with the frame rate the same as the OCT A-line acquisition speed. Based on the spectral analysis of the corneal temporal deformation profiles, the phase velocity of the Lamb wave is obtained at different depths for the major frequency components, which shows the depthwise distribution of the corneal stiffness related to its structural features. Our pilot experiments on ex vivo rabbit eyes demonstrate the feasibility of this method in depth-resolved micro-scale elastography of the cornea. The assessment of the Lamb wave dispersion is also presented, suggesting the potential for the quantitative measurement of corneal viscoelasticity. (C) 2014 Optical Society of America
引用
收藏
页码:3807 / 3821
页数:15
相关论文
共 50 条
  • [1] Noncontact Depth-Resolved Micro-Scale Corneal Elastography
    Wang, Shang
    Larin, Kirill V.
    OPTICAL ELASTOGRAPHY AND TISSUE BIOMECHANICS II, 2015, 9327
  • [2] Optical coherence elastography to evaluate depth-resolved elasticity of tissue
    Yang, Chenming
    Xiang, Zhen
    Li, Zhongliang
    Nan, Nan
    Wang, Xiangzhao
    OPTICS EXPRESS, 2022, 30 (06): : 8709 - 8722
  • [3] Noncontact Quantitative Optical Coherence Elastography of the Cornea
    Larin, Kirill
    Singh, Manmohan
    Han, Zhaolong
    Vantipalli, Srilatha
    Aglyamov, Salavat
    Twa, Michael D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [4] Noncontact depth-resolved quantification of riboflavin/UV-A and rose-bengal/green light cross-linking of rabbit cornea by optical coherence elastography
    Larin, Kirill
    Singh, Manmohan
    Li, Jiasong
    Han, Zhaolong
    Liu, Chih-Hao
    Wu, Chen
    Raghunathan, Raksha
    Aglyamov, Salavat
    Vantipalli, Srilatha
    Twa, Michael D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [5] Possible depth-resolved reconstruction of shear moduli in the cornea following collagen crosslinking (CXL) with optical coherence tomography and elastography
    Regnault G.
    Kirby M.A.
    Wang R.K.
    Shen T.T.
    O’Donnell M.
    Pelivanov I.
    Biomedical Optics Express, 2023, 14 (09) : 5005 - 5021
  • [6] Compression optical coherence micro-elastography of the cornea
    Singh, Manmohan
    Nair, Achuth
    Aglyamov, Salavat
    Larin, Kirill
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [7] Depth-resolved microscopy of cortical hemodynamics with optical coherence tomography
    Srinivasan, Vivek J.
    Sakadzic, Sava
    Gorczynska, Iwona
    Ruvinskaya, Svetlana
    Wu, Weicheng
    Fujimoto, James G.
    Boas, David A.
    OPTICS LETTERS, 2009, 34 (20) : 3086 - 3088
  • [8] Depth-Resolved Dynamics of Aceto-Whitening in Rabbit Cornea Studied by 1300 nm Optical Coherence Tomography
    Khoo, P.
    Mittar, S.
    Kasaragod, D.
    Lu, Z.
    Matcher, S. J.
    SOVREMENNYE TEHNOLOGII V MEDICINE, 2015, 7 (01) : 52 - 59
  • [9] Depth-resolved dynamics of aceto-whitening in rabbit cornea studied by 1300 nm optical coherence tomography.
    Khoo, P.
    Mittar, S.
    Kasaragod, D.
    Lu, Z.
    Matcher, S. J.
    BIOMEDICAL APPLICATIONS OF LIGHT SCATTERING VIII, 2014, 8952
  • [10] Robust, accurate depth-resolved attenuation characterization in optical coherence tomography
    Li, Kaiyan
    Liang, Wenxuan
    Yang, Zihan
    Liang, Yanmei
    Wan, Suiren
    BIOMEDICAL OPTICS EXPRESS, 2020, 11 (02): : 672 - 687