PURPOSE. This work proposes shear wave elastography to quantify the elastic anisotropy of the cornea. METHODS. Experiments were conducted on enucleated porcine eyeballs and anesthetized swine. We used the supersonic shear wave imaging (SSI) method implemented on a dedicated 15-MHz rotating linear ultrasound array. This setup allows determining the shear wave speed variations for a set of radial propagation directions. RESULTS. All the results showed a local anisotropy with one main direction of maximal stiffness. The influence of pulsatility was observed in vivo, and electrocardiogram (ECG) gating was consequently performed for all anesthetized swine. On ex vivo corneas, n = 27 acquisitions were performed in the limbus region, where the collagen fibrils are reported to run tangentially to the sclera. A good match was shown between the direction of maximal stiffness and the expected direction of the collagen fibrils. CONCLUSIONS. This preliminary study demonstrates the potential of SSI for the assessment of corneal anisotropy in both ex vivo and in vivo conditions.
引用
收藏
页码:7545 / 7552
页数:8
相关论文
共 50 条
  • [1] Assessment of shear anisotropy using Supersonic Shear Imaging with rotating arrays: in vivo evidence of cornea elastic anisotropy.
    Thu-Mai Nguyen
    Aubry, Jean-Francois
    Touboul, David
    Bercoff, Jeremy
    Tanter, Mickael
    2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, : 1278 - 1280
  • [2] Supersonic Shear Wave Imaging to Assess Arterial Anisotropy: Ex-vivo Testing of the Horse Aorta
    Shcherbakova, Darya
    Swillens, Abigail
    Caenen, Annette
    De Bock, Sander
    Segers, Patrick
    Papadacci, Clement
    Tanter, Mickael
    Pernot, Mathieu
    Saey, Veronique
    Chiers, Koen
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 1541 - 1544
  • [3] High-Resolution Quantitative Imaging of Cornea Elasticity Using Supersonic Shear Imaging
    Tanter, M.
    Touboul, D.
    Gennisson, Jean-Luc
    Bercoff, Jeremy
    Fink, Mathias
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (12) : 1881 - 1893
  • [4] Shear Wave Anisotropy Imaging
    Hsu, SJ
    Palermi, ML
    Nightingale, KR
    McAleavey, SA
    Dahl, JD
    Trahey, GE
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 921 - 924
  • [5] Shear wave anisotropy imaging
    Hsu, SJ
    Palermi, ML
    Nightingale, KR
    McAleavey, SA
    Dahl, JD
    Trahey, GE
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 1090 - 1093
  • [6] Supersonic Shear Wave Imaging to Assess Arterial Nonlinear Behavior and Anisotropy: Proof of Principle via Ex Vivo Testing of the Horse Aorta
    Shcherbakova, D. A.
    Papadacci, C.
    Swillens, A.
    Caenen, A.
    De Bock, S.
    Saey, V.
    Chiers, K.
    Tanter, M.
    Greenwald, S. E.
    Pernot, M.
    Segers, P.
    ADVANCES IN MECHANICAL ENGINEERING, 2014,
  • [7] REAL TIME QUANTITATIVE ELASTOGRAPHY USING SUPERSONIC SHEAR WAVE IMAGING
    Tanter, Mickael
    Pernot, Mathieu
    Montaldo, Gabriel
    Gennisson, Jean-Luc
    Bavu, Eric
    Mace, Emilie
    Nguyen, Thu-Mai
    Couade, Mathieu
    Fink, Mathias
    2010 7TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, 2010, : 276 - 279
  • [8] Quantification of muscle co-contraction using supersonic shear wave imaging
    Raiteri, Brent J.
    Hug, Francois
    Cresswell, Andrew G.
    Lichtwark, Glen A.
    JOURNAL OF BIOMECHANICS, 2016, 49 (03) : 493 - 495
  • [9] SUPERSONIC SHEAR WAVE ELASTOGRAPHY OF IN VIVO PIG KIDNEY: INFLUENCE OF BLOOD PRESSURE, URINARY PRESSURE AND TISSUE ANISOTROPY
    Gennisson, Jean-Luc
    Grenier, Nicolas
    Combe, Christian
    Tanter, Mickael
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2012, 38 (09): : 1559 - 1567
  • [10] Material Characterization of In Vivo and In Vitro Porcine Brain using Shear Wave Elasticity
    Urbanczyk, Caryn
    Palmeri, Mark
    Bass, Cameron R. 'Dale'
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 21 - 24