Performance of Low Air Volume Dry Powder Inhalers (LV-DPI) when Aerosolizing Excipient Enhanced Growth (EEG) Surfactant Powder Formulations

被引:7
|
作者
Boc, Susan [1 ]
Momin, Mohammad A. M. [1 ]
Farkas, Dale R. [2 ]
Longest, Worth [1 ,2 ]
Hindle, Michael [1 ]
机构
[1] Virginia Commonwealth Univ, Dept Pharmaceut, Med Coll Virginia Campus, Richmond, VA 23284 USA
[2] Virginia Commonwealth Univ, Dept Mech & Nucl Engn, Richmond, VA USA
基金
美国国家卫生研究院;
关键词
low volume DPI; excipient enhanced growth; surfactant; powder; spray-dried; IN-VIVO EVALUATION; DEVICE; THERAPY; PRETERM; LUNGS; VITRO;
D O I
10.1208/s12249-021-01998-9
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Efficient delivery of dry powder aerosols dispersed with low volumes of air is challenging. This study aims to develop an efficient dry powder inhaler (DPI) capable of delivering spray-dried Survanta-EEG powders (3-10 mg) with a low volume (3 mL) of dispersion air. A series of iterative design modifications were made to a base low air volume actuated DPI. The modifications included the replacement of the original capsule chamber with an integral dose containment chamber, alteration of the entrainment air flow path through the device (from single-sided (SS) to straight through (ST)), change in the number of air inlet holes (from one to three), varying the outlet delivery tube length (45, 55, and 90 mm) and internal diameter (0.60, 0.89, and 1.17 mm). The modified devices were evaluated by determining the influence of the modifications and powder fill mass on aerosol performance of spray-dried Survanta-EEG powders. The optimal DPI was also evaluated for its ability to aerosolize a micronized powder. The optimized dose containment unit DPI had a 0.21 mL powder chamber, ST airflow path, three-0.60 mm air inlet holes, and 90 mm outlet delivery tube with 0.89 mm internal diameter. The powder dispersion characteristics of the optimal device were independent of fill mass with good powder emptying in one 3 mL actuation. At 10 mg fill mass, this device had an emitted mass of 5.3 mg with an aerosol Dv50 of 2.7 mu m. After three 3 mL actuations, >85% of the spray-dried powder was emitted from the device. The emitted mass of the optimal device with micronized albuterol sulfate was >72% of the nominal fill mass of 10 mg in one 3 mL actuation. Design optimization produced a DPI capable of efficient performance with a dispersion air volume of 3 mL to aerosolize Survanta-EEG powders.
引用
收藏
页数:12
相关论文
共 9 条
  • [1] Performance of Low Air Volume Dry Powder Inhalers (LV-DPI) when Aerosolizing Excipient Enhanced Growth (EEG) Surfactant Powder Formulations
    Susan Boc
    Mohammad A. M. Momin
    Dale R. Farkas
    Worth Longest
    Michael Hindle
    AAPS PharmSciTech, 22
  • [2] Characterization of excipient enhanced growth (EEG) tobramycin dry powder aerosol formulations
    Hassan, Amr
    Farkas, Dale
    Longest, Worth
    Hindle, Michael
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2020, 591
  • [3] Evaluation and modification of commercial dry powder inhalers for the aerosolization of a submicrometer excipient enhanced growth (EEG) formulation
    Son, Yoen-Ju
    Longest, P. Worth
    Tian, Geng
    Hindle, Michael
    EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2013, 49 (03) : 390 - 399
  • [4] Development and Characterization of Excipient Enhanced Growth (EEG) Surfactant Powder Formulations for Treating Neonatal Respiratory Distress Syndrome
    Boc, Susan
    Momin, Mohammad A. M.
    Farkas, Dale R.
    Longest, Worth
    Hindle, Michael
    AAPS PHARMSCITECH, 2021, 22 (04)
  • [5] Development and Characterization of Excipient Enhanced Growth (EEG) Surfactant Powder Formulations for Treating Neonatal Respiratory Distress Syndrome
    Susan Boc
    Mohammad A. M. Momin
    Dale R. Farkas
    Worth Longest
    Michael Hindle
    AAPS PharmSciTech, 22
  • [6] Aerosolization characteristics of dry powder inhaler formulations for the excipient enhanced growth (EEG) application: Effect of spray drying process conditions on aerosol performance
    Son, Yoen-Ju
    Longest, P. Worth
    Hindle, Michael
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2013, 443 (1-2) : 137 - 145
  • [7] AEROSOL PERFORMANCE AND STABILITY CHARACTERISTICS OF SPRAY- DRIED TOBRAMYCIN EXCIPIENT ENHANCED GROWTH INHALATION POWDER FORMULATIONS
    Wei, Xiangyin
    Kaviratna, Anubhav
    Darweesh, Ruba S.
    Longest, P. Worth
    Hindle, Michael
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2017, 30 (04) : A8 - A8
  • [8] REALISTIC IN VITRO LUNG DELIVERY TESTING OF A PEDIATRIC POSITIVE-PRESSURE DRY POWDER DEVICE WITH EXCIPIENT ENHANCED GROWTH (EEG) ANTIBIOTIC FORMULATION
    Farkas, Dale
    Bass, Karl
    Bonasera, Serena
    Hindle, Michael
    Longest, P. Worth
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2021, 34 (05) : A21 - A21
  • [9] Advancement of a high-dose infant air-jet dry powder inhaler (DPI) with passive cyclic loading: Performance tuning for different formulations
    Howe, Connor
    Momin, Mohammad A. M.
    Aladwani, Ghali
    Strickler, Sarah
    Hindle, Michael
    Longest, Worth
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2023, 643