Magnetic enhancement for the analysis of scintillation crystals by radio frequency glow discharge mass spectrometry

被引:4
|
作者
Yang, Weifeng [1 ,4 ]
Zhuo, Shangjun [1 ,4 ]
Zhu, Zhenli [2 ]
Qian, Rong [1 ,4 ]
Chen, Qiao [3 ]
Dong, Jiangli [1 ,4 ]
机构
[1] Chinese Acad Sci, Natl Ctr Inorgan Mass Spectrometry Shanghai, Shanghai Inst Ceram, Shanghai 200050, Peoples R China
[2] China Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Hubei, Peoples R China
[3] Univ Sussex, Sch Life Sci, Dept Chem, Brighton BN1 9QJ, E Sussex, England
[4] Univ Chinese Acad Sci, Ctr Mat Sci & Optoelect Engn, Beijing 100864, Peoples R China
基金
中国国家自然科学基金;
关键词
QUANTIFICATION; ELEMENTS;
D O I
10.1039/d0ja00481b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A ring magnet enhancement method was developed and utilized for boosting the signal intensity of the analysis of scintillation crystals by radio frequency glow discharge mass spectrometry (rf-GD-MS). Three scintillation crystals of BGO, LYSO and GGAG were applied to observe the performance of the ring magnet method. The investigations suggested that the signal intensities obtained for Bi-209, Lu-175 and Gd-158 with the ring magnet were 7810-27 600 times those obtained without a magnet and were enhanced by a factor of 1.9-2.3 compared with the previously reported stacked magnet and array magnet methods. The results also suggested that the signal intensity was enhanced as the rf power increases from 25 W to 70 W (with a frequency of 13.56 MHz) and a sharp rise up to 4.8 MPa and then decreased with the increase of discharge pressure from 2.2 to 8.4 MPa. The validation results with two NIST samples indicated that the relative errors were within 13% and the lowest detection limit could reach as low as 0.0023 mu g g(-1) (Mg-24, with 40 W and 4.8 MPa) for the ring magnet. The enhancement mechanism of this method was further proposed in which electrons could undergo distinct Larmor precession with a ring magnet, which could increase the collisions between the electrons and other particles to increase the ionization efficiency. Moreover, this work not only shed light on the possible enhancement mechanism by the ring magnet but also provided an effective determination method for the artificial crystals.
引用
收藏
页码:932 / 937
页数:6
相关论文
共 50 条
  • [1] Radio frequency glow discharge mass spectrometry for the characterization of bulk polymers
    Shick, CR
    DePalma, PA
    Marcus, RK
    ANALYTICAL CHEMISTRY, 1996, 68 (13) : 2113 - 2121
  • [2] RADIO-FREQUENCY GLOW DISCHARGE ATOMIC EMISSION AND MASS SPECTROMETRY
    Kawaguchi, Hiroshi
    Tanaka, Tomokazu
    Fukaya, Hiroyasu
    ANALYTICAL SCIENCES, 1991, 7 : 537 - 540
  • [3] Signal Enhancement with Stacked Magnets for High-Resolution Radio Frequency Glow Discharge Mass Spectrometry
    Wei, Juan
    Dong, Jiangli
    Zhuo, Shangjun
    Qian, Rong
    Fang, Yuanxing
    Chen, Qiao
    Patel, Ekbal
    ANALYTICAL CHEMISTRY, 2017, 89 (02) : 1382 - 1388
  • [4] Direct polymer analysis by radio frequency glow discharge spectrometry
    Marcus, RK
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (09) : 1271 - 1277
  • [5] MAGNETIC ENHANCEMENT OF AN IONIZATION SOURCE FOR GLOW-DISCHARGE MASS-SPECTROMETRY
    BENTZ, BL
    HARRISON, WW
    ANALYTICAL CHEMISTRY, 1982, 54 (09) : 1644 - 1646
  • [6] Radio-frequency glow discharge spectrometry: A critical review
    Winchester, MR
    Payling, R
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2004, 59 (05) : 607 - 666
  • [7] Effects of Helium Addition to Radio-Frequency Argon Glow-Discharge Mass Spectrometry
    KRISS, P. O. Box 102, Yusong, Taejon 305-606, Korea, Republic of
    不详
    Anal. Sci., 3 (459-464):
  • [8] Influence of discharge parameters on the resultant sputtered crater shapes for a radio frequency glow discharge mass spectrometry source
    Gibeau, TE
    Marcus, RK
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (12) : 1303 - 1311
  • [9] Solid phase microextraction as sample introduction technique for radio frequency glow discharge mass spectrometry
    Gorecki, T
    Belkin, M
    Caruso, J
    Pawliszyn, J
    ANALYTICAL COMMUNICATIONS, 1997, 34 (10): : 275 - 277
  • [10] Effects of helium addition to radio-frequency argon glow-discharge mass spectrometry
    Woo, JC
    Moon, DM
    Tanaka, T
    Matsuno, M
    Kawaguchi, H
    ANALYTICAL SCIENCES, 1996, 12 (03) : 459 - 464