Rapid antigen detection tests for diagnosis of group A streptococcal pharyngitis:: Comparative evaluation of sensitivity and practicability of 16 in vitro diagnostic medical devices performed in July 2002 by the French health products safety agency (Afssaps) as part of its market control mission

被引:12
|
作者
Charlier-Bret, N [1 ]
Boucher, B
Poyart, C
Quesne, G
Bingen, E
Doit, C
Ho, CT
Deschênes, M
Maisonneuve, P
机构
[1] UECM, DIV, DEDIM, Afssaps, F-93000 St Denis, France
[2] Grp Hosp Necker Enfants Malad, Paris, France
[3] Hop Robert Debre, F-75019 Paris, France
来源
PATHOLOGIE BIOLOGIE | 2004年 / 52卷 / 08期
关键词
Streptococcus pyogenes; pharyngitis; tonsillitis; rapid antigen detection test;
D O I
10.1016/j.patbio.2004.07.013
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Within the scope of its health products control mission, the French Health Products Safety Agency (Afssaps) collaborating with two expert's sites, has assessed the 16 tests available on the French market in 2002 for rapid diagnosis of the Streptococcus A tonsillitis. The purpose of this study was to verify the reliability and rapidity of these tests and to give some information to the users about their analytical criteria and practicability characteristics. The analytical study has been performed on a same panel of four reference strains of Streptococcus pyogenes dilutions to determine the limit of detection of all the reagents in the same condition of methodology. The limit of detection has been calculated with the results expressed in colony forming unit by ml (CFU/ml). The practicability study has permitted to analyze the quality of the presentation, the easiness of the final reading and of performing tests. A score has been established for each rapid test. A classification of the analytical sensitivity (limit of detection) and practicability (score) of these 16 devices has been established. The limit of detection of the reagents giving the best results allows the detection of the lowest bacterial concentration of the panel which is 105 CFU/ml. Regarding practicability, the results suggest that, the immunochromatographic strip methods have the best score in a view with the use by a non medical laboratory. (C) 2004 Publie par Elsevier SAS.
引用
收藏
页码:438 / 443
页数:6
相关论文
empty
未找到相关数据