Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR

被引:12
|
作者
Hippelainen, Eero [1 ,2 ]
Makela, Teemu [2 ,3 ]
Kaasalainen, Touko [3 ]
Kaleva, Erna [4 ]
机构
[1] Univ Helsinki, Helsinki Univ Hosp, HUS Med Imaging Ctr, Clin Physiol & Nucl Med, POB 340, FI-00029 Helsinki, Finland
[2] Univ Helsinki, Dept Phys, Helsinki, Finland
[3] Univ Helsinki, HUS Med Imaging Ctr, Radiol, Helsinki Univ Hosp, Helsinki, Finland
[4] Tampere Univ Hosp, Radiat Therapy Unit Lahti, Lahti, Finland
来源
EJNMMI PHYSICS | 2017年 / 4卷
关键词
Myocardial perfusion imaging; SPECT; Gated CT; IQ-SPECT; LEHR; VENTRICULAR VOLUME; PROGNOSTIC VALUE; TIME; PROTOCOL; SYSTEM;
D O I
10.1186/s40658-017-0187-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EFTrue). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. Results: The CT verification showed that the phantom EF settings were repeatable and accurate with the EFTrue being within 1% point from the manufacture's nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom's volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. Conclusions: The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR
    Eero Hippeläinen
    Teemu Mäkelä
    Touko Kaasalainen
    Erna Kaleva
    EJNMMI Physics, 4
  • [2] A phantom study comparing left ventricular ejection fraction values from myocardial perfusion LEHR-SPECT and IQ-SPECT
    Snoijs, C. H. P.
    Akkermans, J. B. M.
    Jong de, A. N.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S207 - S208
  • [3] Left ventricular ejection fraction in gated myocardial perfusion, a comparison of SPECT and IQ-SPECT
    de Jong, A. N.
    Akkermans, J. B. M.
    Van Asperen, E. A.
    Snoijs, C. H. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S601 - S601
  • [4] Evaluation of the Left Ventricular Ejection Fraction with Gated IQ-SPECT Myocardial Perfusion Imaging
    Joergensen, Troels
    Hansson, Susanne Haase
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2015, 43 (03) : 193 - 200
  • [5] Dynamic heart phantom ejection fractions with IQ-SPECT system vs. LEHR collimators
    Kaleva, E.
    Hippelainen, E.
    Sippo-Tujunen, I.
    Sipila, O.
    Pitkonen, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S309 - S309
  • [6] Myocardial Perfusion Scintigraphy - Comparison between SPECT LEHR and IQ SPECT
    Correia, M. Rodrigues
    Duarte, A.
    Figueiredo, A.
    Matos, S.
    Calado, D.
    Martins, B.
    Loewenthal, C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S543 - S544
  • [7] Comparison of IQ-SPECT/CT with conventional SPECT/CT for myocardial perfusion imaging
    Novak, Mackenzie
    Grancorvitz, Abigail
    Johnson, Carlyn
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [8] Agreement between left ventricular ejection fraction assessed in patients with gated IQ-SPECT and conventional imaging
    Leva, Lucia
    Matheoud, Roberta
    Sacchetti, Gianmauro
    Carriero, Alessandro
    Brambilla, Marco
    JOURNAL OF NUCLEAR CARDIOLOGY, 2020, 27 (05) : 1714 - 1724
  • [9] Agreement between left ventricular ejection fraction assessed in patients with gated IQ-SPECT and conventional imaging
    Lucia Leva
    Roberta Matheoud
    Gianmauro Sacchetti
    Alessandro Carriero
    Marco Brambilla
    Journal of Nuclear Cardiology, 2020, 27 : 1714 - 1724
  • [10] Myocardial perfusion imaging parameters: IQ-SPECT and conventional SPET system comparison
    Havel, Martin
    Kolacek, Michel
    Kaminek, Milan
    Dedek, Vladimir
    Kraft, Otakar
    Sirucek, Pavel
    HELLENIC JOURNAL OF NUCLEAR MEDICINE, 2014, 17 (03): : 200 - 203