Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis

被引:72
|
作者
Moreno-Montanes, Javier [1 ]
Olmo, Natalia [1 ]
Alvarez, Aurora [1 ]
Garcia, Noelia [1 ]
Zarranz-Ventura, Javier [1 ]
机构
[1] Univ Navarra Clin, Dept Ophthalmol, Pamplona 31080, Spain
关键词
NERVE-FIBER LAYER; THICKNESS MEASUREMENTS; NONGLAUCOMATOUS EYES; TIME-DOMAIN; INSTRUMENTS; ACCURACY; DAMAGE; HEAD;
D O I
10.1167/iovs.08-2988
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare the retinal nerve fiber layer (RNFL) evaluation using Cirrus optical coherence tomography (OCT) and Stratus OCT in glaucoma diagnosis. METHODS. One hundred thirty normal and 86 patients with glaucoma were included in this prospective study. The signal strengths of the OCTs were evaluated. The sensitivities and specificities of global RNFL average thickness were compared in the four quadrants and in each clock hour sector. Receiver operating characteristic (ROC) curves, areas under the ROC (AUC), and the likelihood ratio (LR) were plotted for RNFL thickness. Agreement between the OCTs was calculated by using the Bland-Altman method and kappa (kappa) coefficient. RESULTS. Twenty-three percent of all cases examined with Stratus OCT and 1.9% examined with Cirrus OCT had a signal strength below 6 (P = 0.01). In cases with signal strengths >= 6, the mean signal strength was higher with Cirrus OCT than with Stratus OCT (P = 0.01). The RNFL measurements by Cirrus were thicker than those of Stratus OCT (P < 0.05). The AUCs were 0.829 for Stratus and 0.837 for Cirrus OCT (P = 0.706) for global RNFL average. LRs were similar in both OCTs in global RNFL classification but varied in quadrants. The widths of the limits of agreement varied between 42.16 and 97.79 mu m. There was almost perfect agreement (kappa = 0.82) in the average RNFL classification. CONCLUSIONS. Cirrus OCT has better scan quality than Stratus OCT, especially in glaucomatous eyes. In cases with good-quality scans, the sensitivity and specificity, and AUCs were similar. The best agreement was in the global average RNFL classification. The widths of limits of agreements exceed the limits of resolution of the OCTs. (Invest Ophthalmol Vis Sci. 2010; 51:335-343) DOI:10.1167/iovs.08-2988
引用
收藏
页码:335 / 343
页数:9
相关论文
共 50 条
  • [1] Cirrus High-definition Optical Coherence Tomography Versus Spectral Optical Coherence Tomography/Scanning Laser Ophthalmoscopy in the Diagnosis of Glaucoma
    Koh, Kyung Min
    Jin, Sunyoung
    Hwang, Young Hoon
    CURRENT EYE RESEARCH, 2014, 39 (01) : 62 - 68
  • [2] Effect of myopic refractive error on the glaucoma diagnostic ability of cirrus high-definition optical coherence tomography
    Sung, Mi Sun
    Yoon, Jin Ha
    Heo, Hwan
    Park, Sang Woo
    ACTA OPHTHALMOLOGICA, 2015, 93 (03) : e236 - e237
  • [3] Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography
    Park, Seong Bae
    Sung, Kyung Rim
    Kang, Sung Yong
    Kim, Kyung Ri
    Kook, Michael S.
    ARCHIVES OF OPHTHALMOLOGY, 2009, 127 (12) : 1603 - 1609
  • [4] Spectral-domain cirrus high-definition optical coherence tomography is better than time-domain stratus optical coherence tomography for evaluation of macular pathologic features in uveitis
    Gupta, Vishali
    Gupta, Pawan
    Singh, Ramandeep
    Dogra, Mangat R.
    Gupta, Amod
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (06) : 1018 - 1022
  • [5] Macular thickness in healthy eyes using Cirrus high-definition optical coherence tomography
    Jamalzadeh, Afagh
    Foumani, Mohammadrasoul Sabourinejad
    Kazeminezhad, Ehsan
    Hafezi, Vahideh
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2023, 51 (09): : 1023 - 1024
  • [6] Epidermal Segmentation in High-Definition Optical Coherence Tomography
    Li, Annan
    Cheng, Jun
    Yow, Ai Ping
    Wall, Carolin
    Wong, Damon Wing Kee
    Tey, Hong Liang
    Liu, Jiang
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 3045 - 3048
  • [7] Optical Coherence Tomography Angiography Compared With Optical Coherence Tomography for Detection of Early Glaucoma With High Myopia
    Chang, Pei-Yao
    Wang, Jiun-Yi
    Wang, Jia-Kang
    Huang, Tzu-Lun
    Hsu, Yung-Ray
    FRONTIERS IN MEDICINE, 2022, 8
  • [8] Optical coherence tomography angiography compared with optical coherence tomography for detection of glaucoma progression with high myopia
    Chang, Pei-Yao
    Wang, Jiun-Yi
    Wang, Jia-Kang
    SCIENTIFIC REPORTS, 2025, 15 (01):
  • [9] Optical Coherence Tomography in Glaucoma Diagnosis
    Brinkmann, C. K.
    OPHTHALMOLOGE, 2015, 112 (08): : 624 - 625
  • [10] Errors in neuroretinal rim measurement by Cirrus high-definition optical coherence tomography in myopic eyes
    Hwang, Young Hoon
    Kim, Yong Yeon
    Jin, Sunyoung
    Na, Jung Hwa
    Kim, Hwang Ki
    Sohn, Yong Ho
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2012, 96 (11) : 1386 - 1390