International comparisons of He-Ne lasers stabilized with 127I2 at λ ≈ 633 nm -: Part VII:: Comparison of NORAMET 127I2-stabilized He-Ne lasers at λ ≈ 633 nm

被引:9
|
作者
Viliesid, M
Gutierrez-Munguia, M
Galvan, CA
Castillo, HA
Madej, A
Hall, JL
Stone, J
Chartier, A
Chartier, JM
机构
[1] Ctr Nacl Metrol, Queretaro 76900, Mexico
[2] Natl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, Canada
[3] Univ Colorado, Joint Inst Lab Astrophys, Boulder, CO 80309 USA
[4] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
[5] Bur Int Poids & Mesures, F-92312 Sevres, France
关键词
D O I
10.1088/0026-1394/37/4/9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper reports the seventh set of results of a series of grouped laser comparisons from national laboratories undertaken by the Bureau International des Poids et Mesures (BIPM) at the request of the Comite Consultatif pour la Definition du Metre (CCDM, now the Consultative Committee for Length, CCL), during the periods July 1993 to September 1995 and March to July 1997. The results of this comparison, involving six lasers from three countries of North America and from the BIPM, meet the goals set by the CCDM in 1992 and adopted by the Comite International des Poids et Mesures (CIPM) the same year. The standard uncertainty (1 sigma) of the frequency of the He-Ne laser stabilized on the saturated absorption of I-127(2) at lambda approximate to 633 nm is reduced to a level of 12 kHz (2.5 x 10(-11)) when the lasers compared meet the recommended values of the parameters. The lasers were first compared with the BIPMP3 laser, with all the lasers set to the parameter values normally used in each laboratory; the results then ranged from -10.1 kHz to 19.0 kHz. After checking and correcting the values of all the parameters, the range was reduced to -10.6 kHz to +2.6 kHz. Under the latter conditions, the average frequency difference of the group of lasers, with respect to the BIPM4 laser, was +0.3 kHz with a standard uncertainty (1 sigma) of 5.3 kHz. The best relative frequency stabilities, with Allan standard deviations of about 6.2 x 10(-12), 5.1 x 10(-13) and 1.4 x 10(-13), were observed with sampling times of 1 s, 100 s and 1000 s, respectively. The overall best value was 9.1 x 10(-14) for a sampling time of 10 000 s.
引用
收藏
页码:317 / 322
页数:6
相关论文
共 50 条
  • [1] Comparison of 127I2-stabilized He-Ne lasers at 633 nm between the MIKES and the BIPM
    Hu, J
    Riski, K
    Chartier, A
    Picard, S
    METROLOGIA, 1997, 34 (05) : 417 - 419
  • [2] Comparison between the 127I2 stabilized He-Ne lasers wavelength standards at 633 nm
    Qian, Jin
    Liu, Xiu-Ying
    Huang, Xiao-Rong
    Zhang, Zhi-Quan
    Shi, Chun-Ying
    Tan, Hui-Ping
    Xin, Ming
    Wang, Han-Ping
    Cai, Shan
    Liu, Zhong-You
    Zhang, Xiao-Ping
    Jiliang Xuebao/Acta Metrologica Sinica, 2008, 29 (SUPPL.): : 127 - 130
  • [3] COMPARISON OF HE-NE LASERS FROM THE LMM AND THE IMGC STABILIZED ON 127I2 AT 633-NM
    DEVICENTE, J
    SANCHEZPEREZ, AM
    BERTINETTO, F
    METROLOGIA, 1994, 30 (05) : 503 - 506
  • [4] International comparisons of He-Ne lasers stabilized with 127I2 at λ≈633 nm (July 1993 to September 1995) Part IV:: Comparison of Western European lasers at λ≈633 nm
    Darnedde, H
    Rowley, WRC
    Bertinetto, F
    Millerioux, Y
    Haitjema, H
    Wetzels, S
    Pirée, H
    Prieto, E
    Pérez, MM
    Vaucher, B
    Chartier, A
    Chartier, JM
    METROLOGIA, 1999, 36 (03) : 199 - 206
  • [5] International comparisons of He-Ne lasers stabilized with 127I2 at λ≈633 nm (July 1993 to September 1995) Part I:: General
    Chartier, JM
    Chartier, A
    METROLOGIA, 1997, 34 (04) : 297 - 300
  • [6] International comparisons of He-Ne lasers stabilized with 127I2 at λ633 nm (July 1993 to September 1995). Part I: General
    Bur Int des Poids et Mesures, Sevres, France
    Metrologia, 4 (297-300):
  • [7] International comparisons of He-Ne lasers stabilized with 127I2 at λ≈633 nm (July 1993 to September 1995) Part II:: Second comparison of Northern European lasers at λ≈633 nm
    Stahlberg, B
    Ikonen, E
    Haldin, J
    Hu, J
    Ahola, T
    Riski, K
    Pendrill, L
    Karn, U
    Henningsen, J
    Simonsen, H
    Chartier, A
    Chartier, JM
    METROLOGIA, 1997, 34 (04) : 301 - 307
  • [8] International comparisons of He-Ne lasers stabilized with 127I2 at λ≈633 nm (July 1993 to September 1995) Part V:: Comparison of Asian-Pacific and South African lasers at λ ≈ 633 nm
    Brown, N
    Jaatinen, E
    Suh, H
    Howick, E
    Xu, G
    Veldman, I
    Chartier, A
    Chartier, JM
    METROLOGIA, 2000, 37 (02) : 107 - 113
  • [9] International comparisons of He-Ne lasers stabilized with 127I2 at λ≈633nm (July 1993 to September 1995) -: Part III:: Second comparison of Eastern European lasers at λ≈633nm
    Navratil, V
    Fodreková, A
    Gàta, R
    Blabla, J
    Balling, P
    Ziegler, M
    Zeleny, V
    Petrû, F
    Lazar, J
    Veselá, Z
    Gliwa-Gliwinski, J
    Walczuk, J
    Bánréti, E
    Tomanyiczka, K
    Chartier, A
    Chartier, JM
    METROLOGIA, 1998, 35 (06) : 799 - 806
  • [10] COMPARISON BETWEEN THE I-127(2) STABILIZED HE-NE LASERS AT 633 NM AND 612 NM OF THE BIPM AND THE IMGC
    BERTINETTO, F
    CORDIALE, P
    PICOTTO, GB
    CHARTIER, JM
    FELDER, R
    GLASER, MG
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1983, 32 (01) : 72 - 76