Measurements using a prototype array of plastic scintillator bars for reactor based electron anti-neutrino detection

被引:3
|
作者
Netrakanti, P. K. [1 ]
Mulmule, D. [2 ]
Mishra, D. K. [1 ]
Behera, S. P. [1 ]
Dey, R. [1 ,3 ]
Sehgal, R. [1 ]
Sinha, S. K. [4 ]
Jha, V. [1 ,3 ]
Pant, L. M. [1 ,3 ]
机构
[1] Bhabha Atom Res Ctr, Nucl Phys Div, Mumbai, India
[2] Indian Inst Technol, Dept Phys, Mumbai 400076, India
[3] Homi Bhabha Natl Inst, Anushakti Nagar, Mumbai 400094, India
[4] Bhabha Atom Res Ctr, Res Reactor Serv Div, Mumbai 400085, India
关键词
Anti-neutrino; Plastic scintillators; Reactor monitoring;
D O I
10.1016/j.nima.2021.166126
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report measurement of reactor based electron anti-neutrinos from a prototype array of plastic scintillator bars (mini-ISMRAN) located inside Dhruva research reactor hall, BARC. The detector setup took data for 128 days for reactor on (RON) and 51 days for reactor off (ROFF) condition. A detailed analysis procedure is developed to select the anti-neutrino candidate events based on the energy deposition, number of bars hit as well as topological event selection criteria in position and time. Each of these selection criteria are compared with Monte Carlo based simulations and further an embedding technique is used to estimate the efficiencies from a data driven background study. The obtained anti-neutrino like events in RON condition are 218 +/- 50 (stat) +/- 37 (sys) after background subtraction. The obtained results are compared with theoretical estimation which yields 214 +/- 32 (sys) anti-neutrino events for the RON condition.
引用
收藏
页数:13
相关论文
共 19 条
  • [1] A plastic scintillator array for reactor based anti-neutrino studies
    Mulmule, D.
    Behera, S. P.
    Netrakanti, P. K.
    Mishra, D. K.
    Kashyap, V. K. S.
    Jha, V
    Pant, L. M.
    Nayak, B. K.
    Saxena, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 911 : 104 - 114
  • [2] Development of a Gd loaded liquid scintillator for electron anti-neutrino spectroscopy
    Piepke, AG
    Moser, SW
    Novikov, VM
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 432 (2-3): : 392 - 398
  • [3] Development of a Gd loaded liquid scintillator for electron anti-neutrino spectroscopy
    Piepke, Andreas G.
    Moser, S. Wayne
    Novikov, Vladimir M.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999, 432 (2-3): : 392 - 398
  • [4] A new anti-neutrino detection technique based on positronium tagging with plastic scintillators
    Consolati, G.
    Franco, D.
    Jollet, C.
    Meregaglia, A.
    Minotti, A.
    Perasso, S.
    Tonazzo, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 795 : 364 - 369
  • [5] Comparison of anti-neutrino reactor spectrum models with the Bugey 3 measurements
    Physics Letters. Section B: Nuclear, Elementary Particle and High-Energy Physics, 374 (1-3):
  • [6] Comparison of anti-neutrino reactor spectrum models with the Bugey 3 measurements
    Achkar, B
    Aleksan, R
    Avenier, M
    Bagieu, G
    Bouchez, J
    Brissot, R
    Cavaignac, JF
    Collot, J
    Cousinou, MC
    Cussonneau, JP
    Declais, Y
    Dufour, Y
    Favier, J
    Garciaz, F
    Kajfasz, E
    deKerret, H
    Koang, DH
    Lefievre, B
    Lesquoy, E
    Mallet, J
    Metref, A
    Nagy, E
    Obolensky, M
    Pessard, H
    Pierre, F
    Stutz, A
    Wuthrick, JP
    PHYSICS LETTERS B, 1996, 374 (1-3) : 243 - 248
  • [7] Method of fission product beta spectra measurements for predicting reactor anti-neutrino emission
    Asner, D. M.
    Burns, K.
    Campbell, L. W.
    Greenfield, B.
    Kos, M. S.
    Orrell, J. L.
    Schram, M.
    VanDevender, B.
    Wood, L. S.
    Wootan, D. W.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 776 : 75 - 82
  • [8] A radioxenon detection system using CdZnTe, an array of SiPMs, and a plastic scintillator
    Steven A. Czyz
    Abi T. Farsoni
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 313 : 131 - 140
  • [9] A radioxenon detection system using CdZnTe, an array of SiPMs, and a plastic scintillator
    Czyz, Steven A.
    Farsoni, Abi T.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 313 (01) : 131 - 140
  • [10] Prototype of an array SiPM-based scintillator Compton camera for radioactive materials detection
    Ji-Peng Zhang
    Xiu-zuo Liang
    Jia-le Cai
    Yi Liu
    Xian-chao Huang
    Dao-wu Li
    Xiong Xiao
    Xiao-yu Pang
    Zhi-ming Zhang
    Long Wei
    Lei Shuai
    Radiation Detection Technology and Methods, 2019, 3