Micro-CT Analysis of Microgap at a Novel Two-Piece Dental Implant Comprising a Replaceable Sleeve In Vitro

被引:4
|
作者
Kapishnikov, Sergey [1 ]
Gadyukov, Adir [2 ]
Chaushu, Gavriel [3 ,4 ]
Chaushu, Liat [5 ]
机构
[1] Weizmann Inst Sci, Dept Chem Res Support, EM Unit, MicroCT Lab, Rehovot, Israel
[2] New Generat Technol, Implant B, Nazareth, Israel
[3] Rabin Med Ctr, Dept Oral & Maxillofacial Surg, Petah Tiqwa, Israel
[4] Tel Aviv Univ, Maurice & Gabriela Goldschleger Sch Dent Med, Tel Aviv, Israel
[5] Tel Aviv Univ, Maurice & Gabriela Goldschleger Sch Dent Med, Dept Periodontol & Dent Implantol, Tel Aviv, Israel
关键词
cyclic loading; dental implant; micro-CT; microgap; static loading; ABUTMENT CONNECTIONS; BACTERIAL LEAKAGE;
D O I
10.11607/jomi.8563
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Microcomputed tomography (micro-CT) is a relatively new modality to investigate mechanical deformations. The purpose of this study was to assess the microgap at the implant-sleeve connection of a new two-piece dental implant with a replaceable sleeve. Materials and Methods: Implants were assembled with 25-degree angulated abutments. Micro-CT was used to assess implant-sleeve connection gaps under the following mechanical conditions: (1) unloading; (2) compressive 10,000 cyclic loading with 400 N; (3) static compressive load of 200 N or 400 N for 24 hours. Results: The mean gap in the unloaded sample was 2.9 +/- 0.9 mu m. The mean gap difference after cyclic compressive load was 0.3 +/- 0.15 mu m, demonstrating a negligible effect for the cyclic loading. Under static compressive load, there was no increase in microgap size at 200 N. At 400 N, a significant (P < .05) increase was noted. While the mean values increased by 1.9 mu m, the most pronounced significant increase in mean microgap was noted in the direction of force application (5.1 +/- 2.14 mu m), while a significant decrease in mean microgap (1.2 +/- 1.47 mu m) was noted on the opposite side. Conclusion: The mechanical behavior of the implant-sleeve connection under static and dynamic loads was found to be within the previously reported range of implant dentistry.
引用
收藏
页码:451 / 459
页数:9
相关论文
共 14 条
  • [1] Fatigue Evaluation of a New Two-Piece Dental Implant, Having a Replaceable Titanium Sleeve
    Chaushu, Liat
    Chaushu, Gavriel
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2021, 15 (02):
  • [2] A two-piece dental implant comprising a replaceable thin titanium sleeve may be a new approach to re-osseointegration following peri-implant disease
    Liat, Chaushu
    Roni, Shabat
    Adir, Gadyukov
    Gavriel, Chaushu
    MEDICAL HYPOTHESES, 2018, 121 : 103 - 105
  • [3] In vitro synchrotron-based radiography of micro-gap formation at the implant-abutment interface of two-piece dental implants
    Rack, A.
    Rack, T.
    Stiller, M.
    Riesemeier, H.
    Zabler, S.
    Nelson, K.
    JOURNAL OF SYNCHROTRON RADIATION, 2010, 17 : 289 - 294
  • [4] Micro-CT Structure Analysis on Dental Implants: Preliminary In Vitro Trial
    Galletti, Fulvia
    D'Angelo, Tommaso
    Fiorillo, Luca
    Lo Giudice, Paola
    Irrera, Natasha
    Rizzo, Giuseppina
    Cervino, Gabriele
    PROSTHESIS, 2024, 6 (06): : 1437 - 1447
  • [5] RESEARCH AND EDUCATION Accuracy of a custom two-piece surgical guide for all-on-four dental implant placement: An in vitro study
    Elnashoukaty, Hassan M.
    ElDakkak, Sherif
    Abdelhakim, Ahmed
    JOURNAL OF PROSTHETIC DENTISTRY, 2023, 130 (01): : 101.e1 - 101.e9
  • [6] Quantitative analysis of metal artefacts of dental implant in CBCT image by correlation analysis to micro-CT: A microstructural study
    Min, Chang-Ki
    Kim, Kyoung-A
    DENTOMAXILLOFACIAL RADIOLOGY, 2021, 50 (03)
  • [7] Finite Element Analysis of the Stresses Induced by One- Piece and Two-Piece Dental Implants Supporting All-on-4 Implant-Supported Prostheses
    El Homossany, Mahmoud El-Moutassim Bellah
    Askar, Mohamed Gamal
    Mohamed, Ahmed Mostafa Abdelfattah
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (04) : 763 - 770
  • [8] Long-term inflammatory outcome after placement of a novel two-piece dental implant in a dog model experimental peri-implantitis
    Liat, Chaushu
    Haim, Tal
    CLINICAL ORAL INVESTIGATIONS, 2024, 28 (10)
  • [9] Comparative 3D micro-CT and 2D histomorphometry analysis of dental implant osseointegration in the maxilla of minipigs
    Bissinger, Oliver
    Probst, Florian Andreas
    Wolff, Klaus-Dietrich
    Jeschke, Anke
    Weitz, Jochen
    Deppe, Herbert
    Kolk, Andreas
    JOURNAL OF CLINICAL PERIODONTOLOGY, 2017, 44 (04) : 418 - 427
  • [10] The reliability of cone-beam computed tomography to assess bone density at dental implant recipient sites: a histomorphometric analysis by micro-CT
    Gonzalez-Garcia, Raul
    Monje, Florencio
    CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (08) : 871 - 879