Characterization of an optical multilayer hydrophone for use as broadband ultrasound reference receiver - Comparison with PVDF membrane hydrophones

被引:4
|
作者
Wilkens, V [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
D O I
10.1109/ULTSYM.2002.1193513
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The performance of an optical multilayer hydrophone for ultrasound detection comprising a thin high-finesse dielectric interference filter structure deposited onto a plane glass plate is experimentally investigated. The characterization given covers the frequency response in the range from 1 to 75 MHz, the sensitivity, the stability and the directional response. The results are compared with typical characteristics of PVDF hydrophones. The optical multilayer hydrophone is well suited for use as a broadband reference receiver for secondary hydrophone calibration, in particular because of the constant response in a wide frequency range. In an application example the complex-valued broadband frequency response of a fiber-optic hydrophone is determined using this reference.
引用
收藏
页码:773 / 776
页数:4