Preliminary application of a multi-level 3D printing drill guide template for pedicle screw placement in severe and rigid scoliosis

被引:76
|
作者
Liu, Kun [1 ]
Zhang, Qiang [2 ]
Li, Xin [2 ]
Zhao, Changsong [2 ]
Quan, Xuemin [2 ]
Zhao, Rugang [2 ]
Chen, Zongfeng [3 ]
Li, Yansheng [4 ]
机构
[1] Hosp Shunyi Dist Beijing, Beijing, Peoples R China
[2] Capital Med Univ, Beijing Ditan Hosp, Dept Orthoped, 8 Jingshun East St, Beijing 100015, Peoples R China
[3] Third Mil Med Univ, Chongqing, Peoples R China
[4] Beijing Univ Technol, Coll Electromech Engn, Beijing, Peoples R China
关键词
Computer-assisted surgery; Surgical template; 3D printing; Scoliosis; ADOLESCENT IDIOPATHIC SCOLIOSIS; ACCURACY; SPINE; INSTRUMENTATION;
D O I
10.1007/s00586-016-4926-1
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Accurate implantation of pedicle screw in spinal deformity correction surgeries is always challenging. We have developed a method of pedicle screw placement in severe and rigid scoliosis with a multi-level 3D printing drill guide template. From November 2011 to March 2015, ten patients (4 males and 6 females) with severe and rigid scoliosis (Cobb angle > 70A degrees and flexibility < 30%)were included. Multi-level template was designed and manufactured according to the part (two or three levels) of the most severe deformity. The drill template was then placed on the corresponding vertebral surface. Then, pedicle screws were carefully inserted along the trajectories. The other screws were placed in free hand. After surgery, the positions of the pedicle screws were evaluated by CT scan and graded for validation. 48 screws were implanted using templates, other 104 screws in free hand, and the accuracies were 93.8 and 78.8%, respectively, with significant difference. The deformity correction ratio was 67.1 and 41.2% in coronal and sagittal plane post-operatively, respectively. The average operation time was 234.0 +/- 34.1 min, and average blood loss was 557 +/- 67.4 ml. With the application of multi-level template, the incidence of cortex perforation in severe and rigid scoliosis decreased and this technology is, therefore, potentially applicable in clinical practice.
引用
收藏
页码:1684 / 1689
页数:6
相关论文
共 49 条
  • [1] Preliminary application of a multi-level 3D printing drill guide template for pedicle screw placement in severe and rigid scoliosis
    Kun Liu
    Qiang Zhang
    Xin Li
    Changsong Zhao
    Xuemin Quan
    Rugang Zhao
    Zongfeng Chen
    Yansheng Li
    European Spine Journal, 2017, 26 : 1684 - 1689
  • [2] CLINICAL APPLICATION OF A DRILL GUIDE TEMPLATE FOR PEDICLE SCREW PLACEMENT IN SEVERE SCOLIOSIS
    Li, Xin
    Zhang, Yaoshen
    Zhang, Qiang
    Zhao, Changsong
    Liu, Kun
    ACTA ORTOPEDICA BRASILEIRA, 2017, 25 (02): : 67 - 70
  • [3] A multi-level drill guide template improves the accuracy of pedicle screw placement in lumbar spine
    Ge, D. -W.
    Yang, L.
    Chen, X.
    Tang, J.
    Chen, H. -T.
    Li, H. -J.
    Sui, T.
    Zhuang, Y.
    Zheng, S. -N.
    Cao, X. -J.
    EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 2018, 22 (12) : 3643 - 3648
  • [4] The Preliminary Research of Drill Guide Template Design for Pedicle Screw Placement with a Low-cost 3D Pinter
    Liao, Chao-Yaug
    Cheng, Ching-Jen
    Huang, Wei-Jhen
    Cheng, Che-Ming
    ADVANCES IN APPLIED MECHANICS AND MATERIALS, 2014, 493 : 507 - 510
  • [5] A multi-level rapid prototyping drill guide template reduces the perforation risk of pedicle screw placement in the lumbar and sacral spine
    Merc, Matjaz
    Drstvensek, Igor
    Vogrin, Matjaz
    Brajlih, Tomaz
    Recnik, Gregor
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2013, 133 (07) : 893 - 899
  • [6] A multi-level rapid prototyping drill guide template reduces the perforation risk of pedicle screw placement in the lumbar and sacral spine
    Matjaz Merc
    Igor Drstvensek
    Matjaz Vogrin
    Tomaz Brajlih
    Gregor Recnik
    Archives of Orthopaedic and Trauma Surgery, 2013, 133 : 893 - 899
  • [7] A medium invasiveness multi-level patient's specific template for pedicle screw placement in the scoliosis surgery
    Azimifar, Farhad
    Hassani, Kamran
    Saveh, Amir Hossein
    Ghomsheh, Farhad Tabatabai
    BIOMEDICAL ENGINEERING ONLINE, 2017, 16
  • [8] A medium invasiveness multi-level patient’s specific template for pedicle screw placement in the scoliosis surgery
    Farhad Azimifar
    Kamran Hassani
    Amir Hossein Saveh
    Farhad Tabatabai Ghomsheh
    BioMedical Engineering OnLine, 16
  • [9] Application of a novel 3D drill template for cervical pedicle screw tunnel design: a cadaveric study
    Zhengxi Yu
    Guodong Zhang
    Xuanhuang Chen
    Xu Chen
    Changfu Wu
    Yijun Lin
    Wenhua Huang
    Haibin Lin
    European Spine Journal, 2017, 26 : 2348 - 2356
  • [10] Application of a novel 3D drill template for cervical pedicle screw tunnel design: a cadaveric study
    Yu, Zhengxi
    Zhang, Guodong
    Chen, Xuanhuang
    Chen, Xu
    Wu, Changfu
    Lin, Yijun
    Huang, Wenhua
    Lin, Haibin
    EUROPEAN SPINE JOURNAL, 2017, 26 (09) : 2348 - 2356