Determination of Rhodium by Inductively Coupled Plasma Atomic Emission Spectrometry With Chemical Vapour Generation Sample Introduction Technique

被引:5
|
作者
Duan Xu-chuan [1 ]
机构
[1] Tianjin Normal Univ, Coll Chem & Life Sci, Tianjin Key Lab Cytogenet & Mol Regulat, Tianjin 300387, Peoples R China
关键词
Chemical vapour generation; Inductively coupled plasma atomic emission spectrometry; Rhodium; FLUORESCENCE SPECTROMETRY; ABSORPTION-SPECTROMETRY; SEPARATION; SN; CARBON; NICKEL; CD; AU; BI;
D O I
10.3724/SP.J.1096.2010.00421
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An inductively coupled plasma atomic emission spectrometry (ICP-AES) combined with a boron hydride vapour generation device was used for the determination of rhodium in waste catalyst. When the flow rate of sample and sodium borohydride solution was 2 mL/min, waste rate was 6 mL/min, the optimum conditions were as follows; 0. 4 L/min of carrier rate, 1. 0 mol/L of acidity of nitric acid for vapor generation and concentration of sodium borohydride 1. 2( m/V). The study showed that a two-fold increase in the efficiency of rhodium was achieved as compared to nebulization. The detection limit was 6.9 mu g/L, which is slight better than that of with nebulization. The linear range for rhodium was 20 1500 mu g/L with a correlation coefficient of 0. 9986. The mean recovery of sample analysis was 95. 5%. Excellent agreement was found between the present method and FAAS.
引用
收藏
页码:421 / 424
页数:4
相关论文
共 19 条