Direct determination of Cr and Ni in oil samples by isotope dilution and external standard calibration using inductively coupled plasma mass spectrometry

被引:9
|
作者
Viana, Lais Nascimento [1 ]
Saint'Pierre, Tatiana Dillenburg [1 ]
机构
[1] Pontifical Catholic Univ Rio De Janeiro PUC Rio, Dept Chem, Rua Marques Sao Vicente 225, BR-22451900 Rio De Janeiro, RJ, Brazil
关键词
ICP-MS; Isotope dilution; External standard calibration; Nickel; Chromium; Oil samples; TRACE-ELEMENT DETERMINATION; ICP-MS; ULTRASONIC NEBULIZATION; ACCURATE DETERMINATION; CRUDE-OIL; METALS; PETROLEUM; FUEL; SULFUR; INJECTION;
D O I
10.1016/j.microc.2019.104219
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the present work, a method for the determination of Ni and Cr in crude oil samples and derivatives (petroleum asphalt cement (CAP) and malthene) through direct analysis of samples diluted in a mixture of organic solvents (xylene/butanol), by isotope dilution inductively coupled plasma mass spectrometry (ICP-MS), was developed. An exciting advantage of the proposed method was the addition of a polar solvent (butanol) to xylene, which has allowed the dilution of the samples, making possible the use of aqueous standards for the isotope dilution, without phase separation. For comparison, external standard calibration was carried out with organic standards and dilution in the same solvent. The spectrometer was set up and the instrumental parameters were optimized for the introduction of organic samples. Oxygen was introduced into the plasma to avoid carbon build-up at the interface. For interference correction from (ArC+)-Ar-40-C-12 over Cr-52(+), the dynamic reaction cell was employed with CH4 as reaction gas. Accuracy evaluation was performed by analyzing the standard reference material NIST SRM 1084a (wear-metals in lubricant oil). The agreement between the obtained results and the SRM values were within the range of 95% to 106%. The RSDs were lower than 12% for the isotope dilution, and 10% for external standard calibration. The results obtained for Ni and Cr in crude oil and derivatives have shown that both calibration methods are statistically in agreement. Although isotope dilution is not commonly used for routine analysis, it has the potential to be. The time necessary for sample preparation could be reduced with the employment of an autodiluter, and the cost of enriched isotopes is a relative issue. The ID can also be employed as a primary method to verify the accuracy of external standard calibration methods.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Determination of thallium in environmental samples by inductively coupled plasma mass spectrometry:: comparison and validation of isotope dilution and external calibration methods
    Mestek, O
    Koplík, R
    Fingerová, H
    Suchánek, M
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (04) : 403 - 407
  • [2] Iodine determination in food samples using inductively coupled plasma isotope dilution mass spectrometry
    Rädlinger, G
    Heumann, KG
    ANALYTICAL CHEMISTRY, 1998, 70 (11) : 2221 - 2224
  • [3] Comparison of external calibration and isotope dilution methods for iodine determination in foods by inductively coupled plasma mass spectrometry
    Photi, Juntima
    Srisakda, Premmin
    Jurkvon, Tiyanun
    Zeder, Christophe
    Judprasong, Kunchit
    FOOD CHEMISTRY-X, 2025, 25
  • [4] Determination of zinc in plant samples by isotope dilution inductively coupled plasma mass spectrometry
    Mestek, O
    Komínková, J
    Koplík, R
    Suchánek, M
    TALANTA, 2001, 54 (05) : 927 - 934
  • [5] The determination of chlorine in plastic using isotope dilution inductively coupled plasma mass spectrometry
    Ohata, Masaki
    Wada, Ayaka
    ANALYTICAL METHODS, 2017, 9 (26) : 4004 - 4010
  • [6] Determination of hexavalent Cr in river sediments by speciated isotope dilution inductively coupled plasma mass spectrometry
    Drincic, Ana
    Zuliani, Tea
    Scancar, Janez
    Milacic, Radmila
    SCIENCE OF THE TOTAL ENVIRONMENT, 2018, 637 : 1286 - 1294
  • [7] Validation protocol for the determination of copper in plant samples by isotope dilution inductively coupled plasma mass spectrometry
    Koplik, R
    Mestek, O
    Fingerová, H
    Suchánek, M
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (02) : 241 - 245
  • [8] Determination of mercury in biological and environmental samples by inductively coupled plasma mass spectrometry with the isotope dilution technique
    Yoshinaga, J
    Morita, M
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (04) : 417 - 420
  • [9] Determination of selected nickel isotopes in biological samples by inductively coupled plasma mass spectrometry with isotope dilution
    Patriarca, M
    Lyon, TDB
    McGaw, B
    Fell, GS
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (04) : 297 - 302
  • [10] Direct determination of mercury in cosmetic samples by isotope dilution inductively coupled plasma mass spectrometry after dissolution with formic acid
    Gao, Ying
    Shi, Zeming
    Zong, Qinxia
    Wu, Peng
    Su, Jing
    Liu, Rui
    ANALYTICA CHIMICA ACTA, 2014, 812 : 6 - 11