Measuring Temperature of Reflected Shock Wave Using a Standard Chemical Reaction

被引:0
|
作者
Singh, M. Kiran [1 ]
Rajakumar, B. [2 ]
Arunan, E. [1 ]
机构
[1] Indian Inst Sci, Dept Inorgan & Phys Chem, Bangalore 560012, Karnataka, India
[2] Indian Inst Technol Madras, Dept Chem, Madras, Tamil Nadu, India
关键词
Single pulse shock tube; Chemical kinetics; Chemical thermometry; Gas Chromatography; THERMAL-DECOMPOSITION; TUBE; KINETICS; ISOMERIZATION;
D O I
暂无
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The importance of temperature calibration of a single pulse shock tube for obtaining precise kinetics data is well recognized. Some of the commonly used standard reactions for chemical thermometric measurements and their uses in our recent studies are discussed in this article. The chemical standards listed include cyclopropane-carbonitrile, 1,1,1-trifluoroethane, cyclohexene, and ethyl chloride. A recenttemperature calibration study of one of our chemical shock tubes performed using ethyl chloride as external standard for the temperature range from 982-1183 K is presented. The reflected shock temperatures calculated using mach number, T-5(Ms), and those using ethyl chloride as external standard (T-5kin) were found to differ by similar to 1.3-4.5 % in the studied temperature range. This is an improvement compared to our previously reported calibration data where the corresponding difference was in the range similar to 8-14 %. The difference in the calibration factor is due to the various changes made in the shock tube and it highlights the importance of calibration, if and when the shock tube is modified.