Highly sensitive amperometric glucose biosensor based on glassy carbon electrode with copper/palladium coating

被引:24
|
作者
Dai, YQ [1 ]
Shiu, KK [1 ]
机构
[1] Hong Kong Baptist Univ, Dept Chem, Kowloon, Hong Kong, Peoples R China
关键词
amperometric biosensor; glucose determination; copper undercoating; palladium deposition; poly(o-phenylenediamine) film;
D O I
10.1002/elan.200303037
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive amperometric glucose biosensor based on immobilizing glucose oxidase in electropolymerized poly(o-phenylenediamine) film on glassy carbon electrode coated sequentially with copper and palladium layers has been developed. The steady-state amperometric response to glucose was determined by means of the oxidation of hydrogen peroxide generated by the enzymatic reaction at a potential of either + 0.70 or + 0.40 V (vs. Ag \ AgCl reference). The deposited copper/palladium layer showed great enhancement in the performance of the enzyme electrode, possibly due to its better electrocatalytic activity for hydrogen peroxide oxidation and large surface area. Effects of the relative loading of palladium, enzyme and polymer on the electrode performance were examined in detail. Sensitivity and detection limit for glucose determinations at + 0.70 V were about 7.3 muA/mM and 0.1 muM, respectively. A wide linear range up to 6.0 mM glucose could be achieved. Electrode performance was superior to similar works reported in the literature. The response time was less than 2 s and its lifetime was longer than three months. The permeable polyphenylenediamine film also offered good anti-interference ability to ascorbic acid, uric acid and acetaminophen, especially when a detection potential of + 0.40 V was employed.
引用
收藏
页码:1806 / 1813
页数:8
相关论文
共 50 条
  • [1] A Highly Sensitive Amperometric Glucose Biosensor Based on a Nano-cube Cu2O Modified Glassy Carbon Electrode
    Ma Zhen-Zhen
    Yu Hui-Cheng
    Wu Zhao-Yang
    Wu Yan
    Xiao Fu-Bing
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 44 (05) : 822 - 826
  • [2] Amperometric glucose biosensor based on the deposition of copper and glucose oxidase onto glassy carbon transducer
    Rodriguez, MC
    Rivas, GA
    ANALYTICAL LETTERS, 2000, 33 (12) : 2373 - 2389
  • [3] Amperometric Glucose Biosensor Based on Palladium Nanoparticles Combined Aligned Carbon Nanotubes Electrode
    Xu Ying
    Zhao Kun
    Zhang Xiao-Yan
    He Pin-Gang
    Fang Yu-Zhi
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2010, 31 (04): : 672 - 678
  • [4] Amperometric glucose biosensor based on immobilization of glucose oxidase on a magnetic glassy carbon electrode modified with a novel magnetic nanocomposite
    Baghayeri, Mehdi
    Veisi, Hojat
    Ghanei-Motlagh, Masoud
    SENSORS AND ACTUATORS B-CHEMICAL, 2017, 249 : 321 - 330
  • [5] Development of Silver Nanowires Based Highly Sensitive Amperometric Glucose Biosensor
    Ahmad, Mashkoor
    Sun, Hongyu
    Hussain, Muhammad
    Karim, Shafqat
    Nisar, Amjad
    Khan, Maaz
    ELECTROANALYSIS, 2015, 27 (06) : 1498 - 1506
  • [6] Design of an Amperometric Glucose Biosensor Based on Glucose Oxidase/Arginated-Fe3O4/Glassy Carbon Electrode
    Krishna, Rahul
    Chandra, Sudeshna
    Bardhan, Neelkanth
    Salimian, Maryam
    Yang, Y. Y.
    Titus, Elby
    Gracio, Jose
    Bahadur, Dhirendra
    SCIENCE OF ADVANCED MATERIALS, 2013, 5 (04) : 333 - 340
  • [7] Amperometric glucose biosensor based on glucose oxidase encapsulated in carbon nanotube-titania-nafion composite film on platinized glassy carbon electrode
    Choi, Han Nim
    Han, Jee Hoon
    Park, Ji Ae
    Lee, Joong Min
    Lee, Won-Yong
    ELECTROANALYSIS, 2007, 19 (17) : 1757 - 1763
  • [8] A Single ZnO Nanofiber-Based Highly Sensitive Amperometric Glucose Biosensor
    Ahmad, Mashkoor
    Pan, Caofeng
    Luo, Zhixiang
    Zhu, Jing
    JOURNAL OF PHYSICAL CHEMISTRY C, 2010, 114 (20): : 9308 - 9313
  • [9] Bienzymatic amperometric biosensor for glucose based on polypyrrole/ceramic carbon as electrode material
    Tian, FM
    Zhu, GY
    ANALYTICA CHIMICA ACTA, 2002, 451 (02) : 251 - 258
  • [10] A glassy carbon electrode modified with FeS nanosheets as a highly sensitive amperometric sensor for hydrogen peroxide
    Jin, Jiayi
    Wu, Wenqin
    Min, Hui
    Wu, Huimin
    Wang, Shengfu
    Ding, Yu
    Yang, Shuijin
    MICROCHIMICA ACTA, 2017, 184 (05) : 1389 - 1396