Minimum detectable activity concentration of radio-cesium by a LaBr3(Ce) detector for in situ measurements on the ground-surface and in boreholes

被引:3
|
作者
Hasan, Md Moudud [1 ]
Rutten, Jos [1 ]
Camps, Johan [1 ]
Huysmans, Marijke [2 ]
机构
[1] Belgian Nucl Res Ctr, SCK CEN, Boeretang 200, BE-2400 Mol, Belgium
[2] Vrije Univ Brussel VUB, Dept Hydrol & Hydraul Engn, Pleinlaan 2, BE-1050 Brussels, Belgium
关键词
MCNP; Gamma spectrometry; Site characterization; Radioactivity; GAMMA-RAY SPECTROMETRY; MONTE-CARLO-SIMULATION; SOIL CONTAMINATION; NAI DETECTOR; RADIONUCLIDES; CS-137; CALIBRATION; ENVIRONMENT; DEPOSITION; SYSTEM;
D O I
10.1016/j.apradiso.2022.110247
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A 3.81 x 3.81 cm LaBr3(Ce) detector based portable measurement setup has been developed for in situ gamma spectrometric survey of a contaminated site. This system is suitable for above-and below ground surface gamma spectrometric measurements of 137Cs. However, the minimum detectable activity concentration (MDAC), an important parameter of a measurement system, should be estimated for planning purposes of the gamma spectrometric survey. In this study, the MDAC of 137Cs for the measurement setup was investigated. The efficiency of the measurement setups was calculated from Monte Carlo simulations using MCNP code. The numerical model of the different studied set-ups, used in MCNP, performed well for the known cases. The results show that the MDAC varies with the position of the detector with respect to ground surface. A 5-20 min acquisition time, depending on the detector position, can be sufficient to get a MDAC of about 10% of the exemption limit of 137Cs (100 Bq/kg).
引用
收藏
页数:10
相关论文
共 4 条
  • [1] Assessment of radioactive cesium deposition using ground-based gamma-ray spectrometry with a LaBr3(Ce) detector
    Ji, Young-Yong
    Lim, Taehyung
    Hitomi, Keitaro
    Yajima, Tatsuo
    JOURNAL OF RADIOLOGICAL PROTECTION, 2019, 39 (04) : 1006 - 1020
  • [2] The Effect of Intrinsic Radiation from a 3 x 3-in. LaBr3(Ce) Scintillation Detector on In Situ Artificial Radiation Measurements
    Li Sangang
    Cheng Yi
    Wang Lei
    Yang Li
    Liu Huan
    Liao Jiawei
    Zeng Liyang
    Luo Yong
    Wang Xiaoyu
    Pei Qiuyan
    Wang Jie
    NUCLEAR TECHNOLOGY, 2018, 204 (02) : 195 - 202
  • [3] Simulated minimum detectable activity concentration (MDAC) for a real-time UAV airborne radioactivity monitoring system with HPGe and LaBr3 detectors
    Tang, Xiao-Bin
    Meng, Jia
    Wang, Peng
    Cao, Ye
    Huang, Xi
    Wen, Liang-Sheng
    Chen, Da
    RADIATION MEASUREMENTS, 2016, 85 : 126 - 133
  • [4] Spectrometric Estimation of Dose Rate Induced from Radioactive Cesium in the Ground Using a Mobile Gamma-Ray Spectrometry Based on a LaBr3(Ce) Detector
    Ji, Young-Yong
    Lim, Taehyung
    Hitomi, Keitaro
    Yajima, Tatsuo
    HEALTH PHYSICS, 2020, 118 (02): : 215 - 225