Penumbra evaluation of the synergy-S and novalis Micro-MLCs

被引:0
|
作者
Podder, T. [1 ]
Bednarz, G. [1 ]
Misic, V. [1 ]
Yu, Y. [1 ]
Galvin, J. [1 ]
机构
[1] Thomas Jefferson Univ Hosp, Philadelphia, PA 19107 USA
关键词
D O I
10.1118/1.2761003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the effect of variations in the field size, treatment depth, and treatment distance on penumbra at leaf‐end and leaf‐side of two micro‐leaf collimators (μMLCs). Method and Materials: The penumbra characteristics of the Elekta's Synergy‐S micro‐leaf beam modulator with 4mm leave width and BrainLab's Novalis μMLC with 3mm leaf width at isocenter were investigated. Kodak EDR2 film was exposed to approximately 200cGy for a variety of build‐up thicknesses (6MV photon beam, depth =1.5, 5, 10 and 15cm), field sizes (3cm2–10cm2), and treatment distances (90, 100, 105, 110, and 115cm SSD) for both square and circular fields. The penumbra information was expressed as 80% to 20% isodose distance. Results: Analysis of the measured penumbra revealed that BrainLab's μMLC had smaller effective penumbra (2.3mm–3.9mm at leaf‐side and 2.4mm–4.4mm at leaf‐end) as compared to that of Synergy‐S (3mm–6mm at leaf‐side and 4.3mm–7.8mm at leaf‐end) for square fields, at 100cm SAD. The average difference of effective penumbras at leaf‐side and at leaf‐end was about 50% for the Synergy‐S, while this difference was about 6% for BrainLab's μMLC. However, the effective penumbras for circular fields (3cm–9cm diameter) were comparable; ranging 4mm–7mm for the Synergy‐S μMLC and 3mm–6mm for the BrainLab's system. Discussion and Conclusions: Both the BrainLab's and the Elekta's μMLCs produce comparable effective penumbras for typical treatment conditions. The Synergy‐S does not use any adjustable jaw system. In this study, the adjustable jaws of BrainLab's μMLC were positioned at least 0.5cm behind the leaf‐ends to eliminate their effect on penumbra. The variation of penumbra on leaf‐end is significantly smaller for the BrainLab's μMLC. This difference could be due to several factors such as x‐ray target to leaf distance, target's shape and size, or the exact leaf tip design; these issues are being investigated. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2480 / 2480
页数:1
相关论文
共 50 条
  • [1] Physical characterization and comparison of two commercially available micro-MLCs
    Podder, Tarun K.
    Bednarz, Greg
    Yu, Yan
    Galvin, James M.
    PHYSICA MEDICA, 2011, 27 (01) : 52 - 57
  • [2] Synergy-S Stereotactic Radiosurgery for Spinal Tumors
    Strauss, Ido
    Jonas-Kimchi, Tali
    Lidar, Zvi
    Buchbut, Dana
    Shtraus, Natan
    Corn, Benjamin W.
    Kanner, Andrew A.
    ISRAEL MEDICAL ASSOCIATION JOURNAL, 2013, 15 (11): : 678 - 681
  • [3] Spinal Radiosurgery Using the Synergy-S system
    Harel, Ran
    Spiegelmann, Roberto
    ISRAEL MEDICAL ASSOCIATION JOURNAL, 2013, 15 (11): : 712 - 713
  • [4] IMRT planning comparisons between Elekta Synergy-S and Elekta Synergy
    Guo, F. Q.
    He, K.
    Wu, C.
    Yang, C.
    Liu, T.
    Stern, R.
    Purdy, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2461 - 2461
  • [5] Leakage radiation of Elekta Synergy-S LINAC machine
    Yang, C.
    Guo, F.
    Liu, T.
    Purdy, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2466 - 2466
  • [6] FEASIBILITY OF APPLYING THE SYNERGY-S SYSTEM FOR STEREOTACTIC IRRADIATION SPINAL TUMORS
    Kanner, Andrew
    Dan, Schifter
    Shtraus, Natan
    Alani, Shlomo
    Lidar, Zvi
    Corn, Benjamin W.
    NEURO-ONCOLOGY, 2008, 10 (05) : 888 - 888
  • [7] Evaluation of Photoneutron Contamination in Elekta Synergy-S High-Energy Linear Accelerator and Indigenous Novel Solution: The AIIMS Experience
    Subramani, V.
    Singh, M.
    Sharma, S.
    Bisht, R.
    Gopishankar, N.
    Rath, G.
    MEDICAL PHYSICS, 2012, 39 (06) : 3765 - 3765
  • [8] The Rigaku Oxford Diffraction XtaLAB Synergy-S, a versatile microfocus sealed tube diffractometer for weakly diffracting samples.
    Le Magueres, Pierre
    Reinheimer, Eric W.
    Meyer, Mathias
    Jones, Adrian
    Kucharczyk, Damian
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2018, 74 : A468 - A468
  • [9] Utilizing Collimator Rotation to Increase Maximum Treatable Target Dimensions Using An Elekta Synergy-S with Beam Modulator Multileaf Collimator
    Rhodes, C.
    Campbell, S.
    Shields, W.
    Fabien, J.
    Colussi, V.
    Wessels, B.
    MEDICAL PHYSICS, 2012, 39 (06) : 3844 - 3844
  • [10] 医科达Synergy-S加速器多叶准直器剂量学研究
    李军
    张西志
    张先稳
    桂龙刚
    孙新臣
    李金凯
    中国医学装备, 2018, 15 (04) : 39 - 42