Monte Carlo simulations versus experimental measurements in a small animal PET system. A comparison in the NEMA NU 4-2008 framework

被引:2
|
作者
Popota, F. D. [1 ,2 ]
Aguiar, P. [3 ,4 ,5 ]
Espana, S. [6 ,7 ]
Lois, C. [8 ,9 ]
Udias, J. M. [10 ]
Ros, D. [1 ,11 ,12 ]
Pavia, J. [11 ,12 ,13 ]
Gispert, J. D. [11 ,14 ]
机构
[1] Univ Barcelona, Unitat Biofis & Bioengn, Barcelona, Spain
[2] Univ Pompeu Fabra, Barcelona, Spain
[3] Santiago Compostela CHUS, Fundac Ramon Dominguez, CHUS, Galicia, Spain
[4] IDIS, Grp Imaxe Mol, Galicia, Spain
[5] Santiago de Compostela USC, Dept Psiquiatria Radiol & Saude Publ, USC, Galicia, Spain
[6] CNIC, Adv Imaging Unit, Madrid, Spain
[7] Ciber Enfermedades Resp CIBERES, Madrid, Spain
[8] Grp Invest Radiofis, Dept Fis Particulas, Santiago De Compostela, Spain
[9] MIT, Madrid MIT M Vis Consortium, Cambridge, MA 02139 USA
[10] Univ Complutense Madrid, CEI Moncloa, Dept Fis Atom, Grp Fis Nucl Mol & Nucl, Madrid, Spain
[11] Ciber Bioengn Biomet & Nanomed CIBER BBN, Barcelona, Spain
[12] IDIBAPS, Barcelona, Spain
[13] Hosp Clin Barcelona, Serv Med Nucl, Barcelona, Spain
[14] Fdn Pasqual Maragall, Barcelona, Spain
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 01期
关键词
Monte Carlo simulation; molecular imaging; NEMA NU standards; PERFORMANCE EVALUATION; MICROPET R4; SCATTER CORRECTION; VALIDATION; GATE; SCANNER; PENELOPET; SOFTWARE; DESIGN;
D O I
10.1088/0031-9155/60/1/151
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this work a comparison between experimental and simulated data using GATE and PeneloPET Monte Carlo simulation packages is presented. All simulated setups, as well as the experimental measurements, followed exactly the guidelines of the NEMA NU 4-2008 standards using the microPET R4 scanner. The comparison was focused on spatial resolution, sensitivity, scatter fraction and counting rates performance. Both GATE and PeneloPET showed reasonable agreement for the spatial resolution when compared to experimental measurements, although they lead to slight underestimations for the points close to the edge. High accuracy was obtained between experiments and simulations of the system's sensitivity and scatter fraction for an energy window of 350-650 keV, as well as for the counting rate simulations. The latter was the most complicated test to perform since each code demands different specifications for the characterization of the system's dead time. Although simulated and experimental results were in excellent agreement for both simulation codes, PeneloPET demanded more information about the behavior of the real data acquisition system. To our knowledge, this constitutes the first validation of these Monte Carlo codes for the full NEMA NU 4-2008 standards for small animal PET imaging systems.
引用
收藏
页码:151 / 162
页数:12
相关论文
共 31 条
  • [1] NEMA NU 4-2008 Comparison of Preclinical PET Imaging Systems
    Goertzen, Andrew L.
    Bao, Qinan
    Bergeron, Melanie
    Blankemeyer, Eric
    Blinder, Stephan
    Canadas, Mario
    Chatziioannou, Arion F.
    Dinelle, Katherine
    Elhami, Esmat
    Jans, Hans-Sonke
    Lage, Eduardo
    Lecomte, Roger
    Sossi, Vesna
    Surti, Suleman
    Tai, Yuan-Chuan
    Jose Vaquero, Juan
    Vicente, Esther
    Williams, Darin A.
    Laforest, Richard
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (08) : 1300 - 1309
  • [2] NEMA NU 4-2008 validation and applications of the PET-SORTEO Monte Carlo simulations platform for the geometry of the Inveon PET preclinical scanner
    Boisson, F.
    Wimberley, C. J.
    Lehnert, W.
    Zahra, D.
    Pham, T.
    Perkins, G.
    Hamze, H.
    Gregoire, M-C
    Reilhac, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (19): : 6749 - 6763
  • [3] Performance Evaluation of a PEM Scanner Using the NEMA NU 4-2008 Small Animal PET Standards
    Luo, Weidong
    Anashkin, Edward
    Matthews, Christopher G.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 94 - 103
  • [4] Performance evaluation of the small-animal PET scanner ClairvivoPET using NEMA NU 4-2008 Standards
    Sato, K.
    Shidahara, M.
    Watabe, H.
    Watanuki, S.
    Ishikawa, Y.
    Arakawa, Y.
    Nai, Y. H.
    Furumoto, S.
    Tashiro, M.
    Shoji, T.
    Yanai, K.
    Gonda, K.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 696 - 711
  • [5] Technical Note: Performance evaluation of a small-animal PET/CT system based on NEMA NU 4-2008 standards
    Liu, Qiong
    Li, Chaofan
    Liu, Jiguo
    Krish, Kishore
    Fu, Xinlei
    Zhao, Jie
    Chen, Jyh-Cheng
    MEDICAL PHYSICS, 2021, 48 (09) : 5272 - 5282
  • [6] NEMA NU 4-2008 Performance Measurements of Two Commercial Small-Animal PET Scanners: ClearPET and rPET-1
    Canadas, Mario
    Embid, Miguel
    Lage, Eduardo
    Desco, Manuel
    Jose Vaquero, Juan
    Manuel Perez, Jose
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (01) : 58 - 65
  • [7] Performance evaluation of a small-animal PET scanner with 213 mm axis using NEMA NU 4-2008
    Wei, Qing
    Wang, Yingjie
    Huang, Xianchao
    Li, Daowu
    Zheng, Yushuang
    Wang, Peilin
    Sun, Xiaoli
    Chai, Pei
    Han, Xiaorou
    Liu, Shuangquan
    Feng, Baotong
    Zhou, Wei
    Zeng, Xiangtao
    Zhu, Meiling
    Zhang, Zhiming
    Wei, Long
    MEDICAL PHYSICS, 2024,
  • [8] First Test Results of a Commercially Available Clinical PET Scanner Using the NEMA NU 4-2008 Small Animal PET Standards
    Luo, Weidong
    Anashkin, Edward
    Matthews, Chris G.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3991 - 3996
  • [9] Comments on the NEMA NU 4-2008 Standard on Performance Measurement of Small Animal Positron Emission Tomographs
    Patrick Hallen
    David Schug
    Volkmar Schulz
    EJNMMI Physics, 7
  • [10] Comments on the NEMA NU 4-2008 Standard on Performance Measurement of Small Animal Positron Emission Tomographs
    Hallen, Patrick
    Schug, David
    Schulz, Volkmar
    EJNMMI PHYSICS, 2020, 7 (01)