An investigation into the effect of body mass index on the agreement between whole-body fat mass determined by MRI and air-displacement plethysmography

被引:3
|
作者
Weedall, Andrew D. [1 ]
Wilson, Adrian J. [2 ,3 ]
Wayte, Sarah C. [1 ]
机构
[1] Univ Hosp Coventry & Warwickshire, Dept Clin Phys & Bioengn, Radiol Phys, Coventry, W Midlands, England
[2] Univ Hosp Coventry & Warwickshire, Res & Dev, Human Metab Res Unit, Coventry, W Midlands, England
[3] Univ Warwick, Dept Phys, Coventry, W Midlands, England
来源
BRITISH JOURNAL OF RADIOLOGY | 2019年 / 92卷 / 1103期
关键词
BROWN ADIPOSE-TISSUE; MAGNETIC-RESONANCE; QUANTIFICATION; MUSCLE; WATER;
D O I
10.1259/bjr.20190300
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To validate MRI fat measurement protocols using purpose built test objects and by comparison with air-displacement plethysmography (ADP) whole-body fat measurements in non-obese subjects. Methods: Test objects of known fat concentration were used to quantify the accuracy of the MRI measurements. 10 participants with a body mass index in the range 18-30 underwent whole-body MRI using two different Dixon-based sequences (LAVA Flex and IDEAL IQ) to obtain an estimate of their whole-body fat mass. The MRI determined fat mass was compared to the fat mass determined by ADP. Results: MRI test object measurements showed a high correlation to expected fat percentage (r > 098). The participant MRI and ADP results were highly correlated (r = 0.99) but on average (mean +/- standard deviation) MRI determined a higher fat mass than ADP (3.8 +/- 3.1kg for LAVA Flex and 1.9 +/- 3.2kg for IDEAL IQ). There was no trend in the difference between MRI and ADP with total fat mass. Conclusion: The good agreement between MRI and ADP shows that Dixon-based MRI can be used effectively as a tool in physiological research for non-obese adults. Advances In knowledge: This work found that for ten non-obese subjects body mass index had no effect on the MRI determination of whole-body fat mass.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Precision of measurement and body size in whole-body air-displacement plethysmography
    JCK Wells
    NJ Fuller
    International Journal of Obesity, 2001, 25 : 1161 - 1167
  • [2] Precision of measurement and body size in whole-body air-displacement plethysmography
    Wells, JCK
    Fuller, NJ
    INTERNATIONAL JOURNAL OF OBESITY, 2001, 25 (08) : 1161 - 1167
  • [3] Validation of air-displacement plethysmography for estimation of body fat mass in healthy elderly subjects
    A. Bosy-Westphal
    M. Mast
    C. Eichhorn
    C. Becker
    D. Kutzner
    M. Heller
    Manfred J. Müller
    European Journal of Nutrition, 2003, 42 : 207 - 216
  • [4] Validation of air-displacement plethysmography for estimation of body fat mass in healthy elderly subjects
    Bosy-Westhal, A
    Mast, M
    Eichhorn, C
    Becker, C
    Kutzner, D
    Heller, M
    Müller, MJ
    EUROPEAN JOURNAL OF NUTRITION, 2003, 42 (04) : 207 - 216
  • [5] Interaction of clothing and body mass index affects validity of air-displacement plethysmography in adults
    Shafer, Kimberly J.
    Siders, William A.
    Johnson, LuAnn K.
    Lukaski, Henry C.
    NUTRITION, 2008, 24 (02) : 148 - 154
  • [6] Validation of air-displacement plethysmography to measure body fat
    Garrow, JS
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 2003, 77 (05): : 1338 - 1339
  • [7] Effect of body hair on air-displacement plethysmography
    Potter, C
    King, GA
    Thompson, DL
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2004, 36 (05): : S72 - S72
  • [8] Whole-Body MRI-Based Fat Quantification: A Comparison to Air Displacement Plethysmography
    Ludwig, Ute A.
    Klausmann, Florian
    Baumann, Sandra
    Honal, Matthias
    Hoevener, Jan-Bernd
    Koenig, Daniel
    Deibert, Peter
    Buechert, Martin
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 40 (06) : 1437 - 1444
  • [9] Whole-Body Computed Tomography-Based Body Mass and Body Fat Quantification: A Comparison to Hydrostatic Weighing and Air Displacement Plethysmography
    Gibby, Jacob T.
    Njeru, Dennis K.
    Cvetko, Steve T.
    Heiny, Eric L.
    Creer, Andrew R.
    Gibby, Wendell A.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (02) : 302 - 308
  • [10] Monitoring body fat in the elderly: application of air-displacement plethysmography
    Fields, DA
    Hunter, GR
    CURRENT OPINION IN CLINICAL NUTRITION AND METABOLIC CARE, 2004, 7 (01): : 11 - 14