Indications and outcomes of AneuRx Phase III Trial versus use of commercial AneuRx stent graft

被引:13
|
作者
Ayerdi, J [1 ]
McLafferty, RB [1 ]
Markwell, SJ [1 ]
Solis, MM [1 ]
Parra, JR [1 ]
Gruneiro, LA [1 ]
Ramsey, DE [1 ]
Hodgson, KJ [1 ]
机构
[1] So Illinois Univ, Sch Med, Div Vasc Surg, Dept Surg, Springfield, IL 62704 USA
关键词
D O I
10.1067/mva.2003.222
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: Approval by the United States Food and Drug Administration of endoluminal repair of abdominal aortic aneurysm (AAA) with the AneuRx stent graft was based on the outcome of a multicenter trial in which patients met strict inclusion and exclusion criteria. Since widespread use of the commercially available graft, little information is available as to whether indications and outcomes have evolved. We examined this important issue at our institution. Methods. Data concerning indications, repair, and follow-up for all patients undergoing endoluminal repair of AAA was prospectively entered into a patient registry. Group I comprised consecutive patients enrolled in the AneuRx Phase III clinical trial between November 1998 and September 2000. Group 2 consisted of consecutive patients who underwent implantation of the commercially available AneuRx graft between May 1999 and June 2001. Results. Group 1 included 42 patients (mean age, 72 years), and group 2 included 54 patients (mean age, 73 years). Patient demographics and risk factors were similar between the two groups. Maximum aortic aneurysm diameter was significantly greater (P = .021) in group 1 (55 mm +/- 10.9 [SD] mm) compared with group 2 (52 +/- 15.6 mm). Maximum infrarenal aortic neck length was significantly longer (P = .022) in group 1 (30 +/- 11.7 nun) than in group 2 (23 +/- 12.0 mm). Maximum left common iliac artery diameter in group 1 (13.0 +/- 3.2 mm) was significantly smaller (P =.032) than that in group 2 (14 +/- 6.5 mm). During follow-up, no differences were observed for number of endoleaks, subsequent interventions, or graft explantation between the two groups. Conclusions. In group 2 patients AAAs were significantly smaller, infrarenal aortic neck length was shorter, and left common iliac arteries were larger. Common iliac artery ectasia and aneurysmal disease has become another indication for use of the AneuRx commercial graft at our institution, with no significant differences in intermediate outcome. Given the possibility for evolving indications compared with trial inclusion and exclusion criteria, institutions that use the AneuRx commercial graft should prospectively monitor outcomes for quality assurance.
引用
收藏
页码:739 / 743
页数:5
相关论文
共 50 条
  • [1] Endovascular versus open AAA repair: AneuRx stent-graft multicenter clinical trial
    Zarins, CK
    White, RA
    Schwarten, DE
    Kinney, EV
    Diethrich, EB
    Hodgson, KJ
    Sullivan, TM
    Bersin, RM
    Matsumura, JS
    Money, SR
    Laird, J
    Moore, W
    Fogarty, TJ
    JOURNAL OF ENDOVASCULAR SURGERY, 1999, 6 (01): : 122 - 123
  • [2] AneuRx stent graft versus open surgical repair of abdominal aortic aneurysms: Multicenter prospective clinical trial
    Zarins, CK
    White, RA
    Schwarten, D
    Kinney, E
    Diethrich, EB
    Hodgson, KJ
    Fogarty, TJ
    JOURNAL OF VASCULAR SURGERY, 1999, 29 (02) : 292 - 306
  • [3] AneuRx stent graft versus open surgical repair of abdominal aortic aneurysms: Multicenter prospective clinical trial - Discussion
    Makaroun, MS
    Zarins, CK
    Ahn, SS
    Quinones-Baldrich, WJ
    White, GM
    JOURNAL OF VASCULAR SURGERY, 1999, 29 (02) : 306 - 308
  • [4] Abdominal aortic aneurysm repair with the AneuRx stent graft in cardiology practice
    Howell, MH
    Strickman, N
    Dougherty, K
    Hallman, C
    Krajcer, Z
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 75A - 75A
  • [5] Three year single centre experience with the AneuRx aortic stent graft
    Nolthenius, RPT
    van Herwaarden, JA
    van den Berg, JC
    van Marrewijk, C
    Teijink, JAW
    Moll, FL
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2001, 22 (03) : 257 - 264
  • [6] Crossing the limbs: A useful adjunct for successful deployment of the AneuRx Stent-Graft
    Ramaiah, VG
    Thompson, CS
    Shafique, S
    Rodriguez, JA
    Ravi, R
    DiMugno, L
    Diethrich, EB
    JOURNAL OF ENDOVASCULAR THERAPY, 2002, 9 (05) : 583 - 586
  • [7] AneuRx stent-graft failure four years after TAA exclusion
    Schepens, MA
    van den Berg, JC
    Moll, FL
    Dossche, KM
    Heijmen, RH
    JOURNAL OF ENDOVASCULAR THERAPY, 2004, 11 (01) : 49 - 52
  • [8] Endovascular aneurysm repair with the AneuRx stent-graft is safe, but is it effective?
    Chuter, TAM
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 38 (04) : 1047 - 1048
  • [9] Aneurysm rupture after endovascular repair using the AneuRx stent graft
    Zarins, CK
    White, RA
    Fogarty, TJ
    JOURNAL OF VASCULAR SURGERY, 2000, 31 (05) : 960 - 970
  • [10] Regarding "AneuRx stent graft versus open surgical repair of abdominal aortic aneurysms: multicenter prospective clinical trial" - Reply
    Zarins, CK
    JOURNAL OF VASCULAR SURGERY, 1999, 30 (03) : 580 - 581