An Ankle-Foot Prosthesis Emulator With Control of Plantarflexion and Inversion-Eversion Torque

被引:28
|
作者
Kim, Myunghee [1 ,2 ]
Chen, Tianjian [1 ,3 ]
Chen, Tianyao [1 ,4 ,5 ]
Collins, Steven H. [1 ,6 ]
机构
[1] Carnegie Mellon Univ, Dept Mech Engn, Pittsburgh, PA 15213 USA
[2] Harvard Univ, John A Paulson Sch Engn & Appl Sci, Cambridge, MA 02138 USA
[3] Columbia Univ, Dept Mech Engn, New York, NY 10025 USA
[4] Catholic Univ Amer, Dept Biomed Engn, Washington, DC 20064 USA
[5] HuMoTech, Pittsburgh, PA 15206 USA
[6] Stanford Univ, Dept Mech Engn, Stanford, CA 94305 USA
基金
美国国家科学基金会;
关键词
Locomotion; mechanism design; medical robots and systems; prosthetics; rehabilitation robotics; LATERAL BALANCE; ENERGY-COST; WALKING; WORK; MOTION; JOINT; STABILIZATION; STABILITY; BENEFITS; DESIGN;
D O I
10.1109/TRO.2018.2830372
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
Ankle inversion-eversion compliance is an important feature of conventional prosthetic feet, and control of inversion, or roll, in active prostheses could improve balance for people with amputation. We designed a tethered ankle-foot prosthesis with two independently actuated toes that are coordinated to provide plantarflexion and inversion-eversion torques. A Bowden cable tether provides series elasticity. The prosthesis is simple and lightweight, with a mass of 0.72 kg. Strain gauges on the toes measure torque with less than 1% root mean squared (RMS) error. Benchtop tests demonstrated a step response rise time of less than 33 ms, peak torques of 250 N.m in plantarflexion and +/- 30 N.m in inversion-eversion, and peak power above 3 kW. The phase-limited closed-loop torque bandwidth is 20 Hz with a chirp from 10 to 90 N.m in plantarflexion, and 24 Hz with a chirp from -20 to 20 N.m in inversion. The system has low sensitivity to toe position disturbances at frequencies of up to 18 Hz. Walking trials with an amputee subject demonstrated RMS torque tracking errors of less than 5.1 N.m in plantarflexion and less than 1.5 N-m in inversion-eversion. These properties make the platform suitable for testing inversion-related prosthesis features and controllers in experiments with humans.
引用
收藏
页码:1183 / 1194
页数:12
相关论文
共 50 条
  • [1] An Ankle-Foot Prosthesis Emulator with Control of Plantarflexion and Inversion-Eversion Torque
    Collins, Steven H.
    Kim, Myunghee
    Chen, Tianjian
    Chen, Tianyao
    2015 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2015, : 1210 - 1216
  • [2] Design and Preliminary evaluation of a Two DOFs cable-Driven ankle-Foot Prosthesis with active Dorsiflexion-Plantarflexion and inversion-eversion
    Ficanha, Evandro Maicon
    Ribeiro, Guilherme Aramizo
    Dallali, Houman
    Rastgaar, Mohammad
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2016, 4
  • [3] Conceptual Design of a Powered Ankle-Foot Prosthesis for Walking with Inversion and Eversion
    Masum, Habib
    Bhaumik, Subhasis
    Ray, Ranjit
    2ND INTERNATIONAL CONFERENCE ON INNOVATIONS IN AUTOMATION AND MECHATRONICS ENGINEERING, ICIAME 2014, 2014, 14 : 228 - 235
  • [4] Design, Control, and Evaluation of a Robotic Ankle-Foot Prosthesis Emulator
    Anderson, Anthony J.
    Hudak, Yuri F.
    Muir, Brittney C.
    Aubin, Patrick M.
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2023, 5 (03): : 741 - 752
  • [5] Design and development of ankle-foot prosthesis with delayed release of plantarflexion
    Mitchell, Michael
    Craig, Katelynn
    Kyberd, Peter
    Biden, Edmund
    Bush, Greg
    JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2013, 50 (03): : 409 - 422
  • [6] GAIT EMULATOR FOR EVALUATION OF A POWERED ANKLE-FOOT PROSTHESIS
    Ficanha, Evandro
    Dallali, Houman
    Rastgaar, Mo
    PROCEEDINGS OF THE ASME 10TH ANNUAL DYNAMIC SYSTEMS AND CONTROL CONFERENCE, 2017, VOL 1, 2017,
  • [7] A Universal Ankle-Foot Prosthesis Emulator for Human Locomotion Experiments
    Caputo, Joshua M.
    Collins, Steven H.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (03):
  • [8] An Ankle-Foot Prosthesis Emulator Capable of Modulating Center of Pressure
    Chiu, Vincent L.
    Voloshina, Alexandra S.
    Collins, Steven H.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2020, 67 (01) : 166 - 176
  • [9] INSTRUMENTED WALKWAY FOR ESTIMATION OF THE ANKLE IMPEDANCE IN DORSIFLEXION-PLANTARFLEXION AND INVERSION-EVERSION DURING STANDING AND WALKING
    Ficanha, Evandro M.
    Ribeiro, Guilherme
    Aagaah, Mohammad Rastgaar
    PROCEEDINGS OF THE ASME 8TH ANNUAL DYNAMIC SYSTEMS AND CONTROL CONFERENCE, 2015, VOL 3, 2016,
  • [10] Simulating the effect of ankle plantarflexion and inversion-eversion exoskeleton torques on center of mass kinematics during walking
    Bianco, Nicholas A.
    Collins, Steven H.
    Liu, Karen
    Delp, Scott L.
    PLOS COMPUTATIONAL BIOLOGY, 2023, 19 (08)