Analytical solution to 3D SPECT reconstruction with non-uniform attenuation, scatter, and spatially-variant resolution variation for variable focal-length fan-beam collimators

被引:2
|
作者
Wen, JH [1 ]
Lu, HB [1 ]
Li, TF [1 ]
Liang, ZR [1 ]
机构
[1] SUNY Stony Brook, Dept Radiol, Stony Brook, NY 11794 USA
关键词
analytical reconstruction; non-uniform attenuation; Poisson noise reduction; scatter; distance-dependent resolution variation; attenuated Radon transform; and variable focal length fan-beam geometry;
D O I
10.1117/12.480472
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the past decades, analytical (non-iterative) methods have been extensively investigated and developed for the reconstruction of three-dimensional (3D) single-photon emission computed tomography (SPECT). However, it becomes possible only recently when the exact analytic non-uniform attenuation reconstruction algorithm was derived. Based on the explicit inversion formula for the attenuated Radon transform discovered by Novikov (2000), we extended the previous researches of inverting the attenuated Radon transform of parallel-beam collimation geometry to fan-beam and variable focal-length fan-beam (VFF) collimators and proposed an efficient, analytical solution to 3D SPECT reconstruction with VFF collimators, which compensates simultaneously for non-uniform attenuation, scatter, and spatially-variant or distance-dependent resolution variation (DDRV), as well as suppression of signal dependent non-stationary Poisson noise. In this procedure, to avoid the reconstructed images being corrupted by the presence of severe noise, we apply a Karhune-Loeve (K-L) domain adaptive Wiener filter, which accurately treats the non-stationary Poisson noise. The scatter is then removed by our scatter estimation method, which is based on the energy spectrum and modified from the triple-energy-window acquisition protocol. For the correction of DDRV, a distance-dependent deconvolution is adapted to provide a solution that realistically characterizes the resolution kernel in a real SPECT system. Finally image is reconstructed using our VFF non-uniform attenuation inversion formula.
引用
收藏
页码:1858 / 1867
页数:10
相关论文
共 5 条
  • [1] A ray-driven approach to analytical SPECT reconstruction of non-uniform attenuation with variable focal-length fan-beam collimators
    Wen, JH
    Li, TF
    Liang, ZR
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 731 - 735
  • [2] Analytical compensation for spatially variant detector response in SPECT with varying focal-length fan-beam collimators
    Li, TF
    Wen, JH
    Liang, ZR
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (03) : 398 - 404
  • [3] Compensation for non-stationary detector response in analytical varying focal-length fan-beam SPECT reconstruction
    Li, TF
    Wen, JH
    Liang, ZR
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1686 - 1690
  • [4] Analytical solution to Cone-beam SPECT reconstruction with non-uniform attenuation and distance-dependent resolution variation
    Zhang, Hao
    Wen, Junhai
    Yang, Jing
    Hu, Peng
    Liang, Zhengrong
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [5] Towards analytical solution for 3D SPECT reconstruction with non-uniform attenuation and distance-dependent resolution variation: A Monte Carlo simulation study
    Lu, HB
    Wen, JH
    Li, X
    Li, TF
    Han, GP
    Liang, ZG
    MEDICAL IMAGING 2002: IMAGE PROCESSING, VOL 1-3, 2002, 4684 : 20 - 28