A compact PET detector array using wavelength shifting fibres

被引:0
|
作者
Herbert, DJ [1 ]
Meng, LJ [1 ]
Dallimore, M [1 ]
Ramsden, D [1 ]
机构
[1] Univ Southampton, Dept Phys & Astron, Southampton SO14 1BJ, Hants, England
来源
1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3 | 1999年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The basis for a design of a high-resolution PET imaging system, utilising a wavelength-shifting (WLS) fibre readout technique is presented here. Scintillation light is transported from the individual LSO pixels in the detector array through the WLS fibres to a multi-pixel hybrid photodiode (M-HPD). Each detector array consists of a number of 2x2x10mm LSO crystals viewed by two orthogonal ribbons of WLS fibres. An array of as many as 30x31 crystals can thus be read-out by a single, small 61-pixel M-HPD tube. The design is both compact, inexpensive and suitable for application not only in the field of small-animal PET or in a dedicated PEM system, but could also operate in a combined PET/MRI system. Initial tests have shown that the signal level generated at the detector by a single 511 keV energy-deposit, is 20 similar to 25 photoelectrons. Separate tests have indicated that, at this signal-level, a resolving time of similar to 25ns could be achievable.
引用
收藏
页码:1304 / 1308
页数:3
相关论文
共 50 条
  • [1] A compact high resolution pet detector using an APD array.
    Shao, Y
    Silverman, RW
    Farrell, R
    Cirignano, L
    Grazioso, R
    Shah, KS
    Cherry, SR
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 75P - 75P
  • [2] A compact neutron detector using scintillators with wavelength shifting fibers by backside readout method
    Katagiri, M
    Matsubayashi, M
    Sakasai, K
    Nakamura, T
    Ebine, M
    Birumachi, A
    Rhodes, N
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 529 (1-3): : 313 - 316
  • [3] Performance Demonstration for PET Detector Using 0.2 mm phi Wavelength-Shifting Fibers
    Kobayashi, A.
    Ito, H.
    Kawai, H.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [4] Development of a high-resolution PET detector using LSO and wavelength-shifting fibers
    Worstell, W
    Johnson, O
    Zawarzin, V
    1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1756 - 1760
  • [5] Studies of a 3D PET Detector with Wavelength Shifting Fibers
    Han, S.
    Kawai, H.
    Ito, H.
    Ijima, S.
    Kodama, S.
    Kumogoshi, D.
    Mase, K.
    Suzuki, S.
    Tabata, M.
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [6] First results with high-resolution PET detector modules using wavelength-shifting fibers
    Boston Univ Physics Dep and Cent for, Photonics, Boston, United States
    IEEE Trans Nucl Sci, 6 pt 2 (2993-2999):
  • [7] First results with high-resolution PET detector modules using wavelength-shifting fibers
    Worstell, W
    Johnson, O
    Kudrolli, H
    Zavarzin, V
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (06) : 2993 - 2999
  • [8] First results with high-resolution PET detector modules using wavelength-shifting fibers
    Worstell, W
    Johnson, O
    Kudrolli, H
    Zavarzin, V
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 929 - 933
  • [9] A compact gamma-ray detector using wavelength-shifting fibers coupled to YAP scintillation crystal
    Zhu Jie
    Ma Hongguang
    Ma Wenyan
    Zeng Hui
    Wang Zhaomin
    Xu Zizhong
    NUCLEAR SCIENCE AND TECHNIQUES, 2008, 19 (01) : 45 - 49