Error in measurement of oxygen and carbon dioxide concentrations by the DeltatracII metabolic monitor in the presence of desflurane

被引:9
|
作者
Scheeren, TWL
Krossa, M
Meriläinen, P
Arndt, JO
机构
[1] Univ Dusseldorf, Inst Expt Anaesthesiol, D-40225 Dusseldorf, Germany
[2] Instrumentarium Corp, Datex Engstrom Div, Helsinki, Finland
关键词
anaesthetics volatile; desflurane; monitoring; oxygen; measurement techniques; calorimetry; equipment; gas analysers;
D O I
10.1093/bja/80.4.521
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In experiments in dogs on the metabolic effects of inhalation anaesthetics, we noticed that in the presence of desflurane, oxygen uptake ((V) over dot O-2) measured with the Deltatracll metabolic monitor seemingly increased whereas it decreased when determined independently by the Fick principle. This difference remained even after correction for changes in gas concentration on addition of an inhalation anaesthetic. Therefore, we suspected that desflurane interferes with the measurement of gas concentrations. Using different precision gases, we found that desflurane disturbed both the paramagnetic oxygen sensor and the infrared carbon dioxide detector so that the measured oxygen (when Fi(O2) was >0.21) and carbon dioxide concentrations were greater than expected. These errors multiply in the computing process of oxygen uptake by the Deltatracll. When the Deltatracll is to be used during inhalation anaesthesia, its results should be corrected for the presence of an anaesthetic gas. More importantly, corrections must also be made for measurement errors of the oxygen and carbon dioxide sensors, unless the device has been equipped with a modified (nickel membrane) oxygen sensor insensitive to the presence of volatile agents.
引用
收藏
页码:521 / 524
页数:4
相关论文
共 50 条