Geometric feature-based multimodal image registration of contrast-enhanced cardiac CT with gated myocardial perfusion SPECT

被引:21
|
作者
Woo, Jonghye [1 ]
Slomka, Piotr J. [1 ]
Dey, Damini [1 ]
Cheng, Victor Y. [1 ]
Hong, Byung-Woo [1 ]
Ramesh, Amit [1 ]
Berman, Daniel S. [1 ]
Karlsberg, Ronald P. [1 ]
Kuo, C. -C. Jay [1 ]
Germano, Guido [1 ]
机构
[1] Cedars Sinai Med Ctr, Dept Imaging, AIM Program, Los Angeles, CA 90048 USA
关键词
multimodal image registration; myocardial perfusion SPECT; coronary CTA; EMISSION COMPUTED-TOMOGRAPHY; EJECTION FRACTION; QUANTIFICATION; VALIDATION; FUSION; SEGMENTATION; QUANTITATION; MOTION;
D O I
10.1118/1.3253301
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Cardiac computed tomography (CT) and single photon emission computed tomography (SPECT) provide clinically complementary information in the diagnosis of coronary artery disease (CAD). Fused anatomical and physiological data acquired sequentially on separate scanners can be coregistered to accurately diagnose CAD in specific coronary vessels. Methods: A fully automated registration method is presented utilizing geometric features from a reliable segmentation of gated myocardial perfusion SPECT (MPS) volumes, where regions of myocardium and blood pools are extracted and used as an anatomical mask to de-emphasize the inhomogeneities of intensity distribution caused by perfusion defects and physiological variations. A multiresolution approach is employed to represent coarse-to-fine details of both volumes. The extracted voxels from each level are aligned using a similarity measure with a piecewise constant image model and minimized using a gradient descent method. The authors then perform limited nonlinear registration of gated MPS to adjust for phase differences by automatic cardiac phase matching between CT and MPS. For phase matching, they incorporate nonlinear registration using thin-plate-spline-based warping. Rigid registration has been compared with manual alignment (n = 45) on 20 stress/rest MPS and coronary CTA data sets acquired from two different sites and five stress CT perfusion data sets. Phase matching was also compared to expert visual assessment. Results: As compared with manual alignment obtained from two expert observers, the mean and standard deviation of absolute registration errors of the proposed method for MPS were 4.3 +/- 3.5, 3.6 +/- 2.6, and 3.6 +/- 2.1 mm for translation and 2.1 +/- 3.2 degrees, 0.3 +/- 0.8 degrees, and 0.7 +/- 1.2 degrees for rotation at site A and 3.8 +/- 2.7, 4.0 +/- 2.9, and 2.2 +/- 1.8 mm for translation and 1.1 +/- 2.0 degrees, 1.6 +/- 3.1 degrees, and 1.9 +/- 3.8 degrees for rotation at site B. The results for CT perfusion were 3.0 +/- 2.9, 3.5 +/- 2.4, and 2.8 +/- 1.0 mm for translation and 3.0 +/- 2.4 degrees, 0.6 +/- 0.9 degrees, and 1.2 +/- 1.3 degrees for rotation. The registration error shows that the proposed method achieves registration accuracy of less than 1 voxel (6.4 x 6.4 x 6.4 mm) misalignment. The proposed method was robust for different initializations in the range from -80 to 70, -80 to 70, and -50 to 50 mm in the x-, y-, and z-axes, respectively. Validation results of finding best matching phase showed that best matching phases were not different by more than two phases, as visually determined. Conclusions: The authors have developed a fast and fully automated method for registration of contrast cardiac CT with gated MPS which includes nonlinear cardiac phase matching and is capable of registering these modalities with accuracy <10 mm in 87% of the cases. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3253301]
引用
收藏
页码:5467 / 5479
页数:13
相关论文
共 50 条
  • [1] ANALYSIS OF GATED MYOCARDIAL PERFUSION SPECT IMAGES BASED ON COMPUTATIONAL IMAGE REGISTRATION
    Alves, Raquel S.
    Faria, Diogo Borges
    Costa, Durval Campos
    Tavares, Joao Manuel R. S.
    2015 IEEE 4TH PORTUGUESE MEETING ON BIOENGINEERING (ENBENG), 2015,
  • [2] Incorporating global information in feature-based multimodal image registration
    Li, Yong
    Stevenson, Robert
    JOURNAL OF ELECTRONIC IMAGING, 2014, 23 (02)
  • [3] Advances in Contrast-Enhanced Cardiovascular CT for the Evaluation of Myocardial Perfusion
    Nasis A.
    Seneviratne S.
    DeFrance T.
    Current Cardiovascular Imaging Reports, 2010, 3 (6) : 372 - 381
  • [4] Feature-based GDLOH deformable registration for CT lung image
    Yang, Yuhan
    Xie, Yaoqin
    MEASUREMENT TECHNOLOGY AND ENGINEERING RESEARCHES IN INDUSTRY, PTS 1-3, 2013, 333-335 : 969 - 973
  • [5] Automated quantification of myocardial perfusion based on segmentation and non-rigid registration of contrast-enhanced cardiac magnetic resonance images
    Giacomo Tarroni
    Cristiana Corsi
    Federico Veronesi
    James Walter
    Claudio Lamberti
    Roberto M Lang
    Victor Mor-Avi
    Amit R Patel
    Journal of Cardiovascular Magnetic Resonance, 13 (Suppl 1)
  • [6] Multimodal Medical Image Registration Based on Feature Spheres in Geometric Algebra
    Cao, Wenming
    Lyu, Fangfang
    He, Zhihai
    Cao, Guitao
    He, Zhiquan
    IEEE ACCESS, 2018, 6 : 21164 - 21172
  • [7] Infarct detection and sizing by myocardial perfusion SPECT correlates well with contrast-enhanced MRI
    Di Carli, MF
    Dorbala, S
    Averbach, M
    Sulaiman, A
    Kwong, R
    CIRCULATION, 2003, 108 (17) : 456 - 456
  • [8] Non-rigid Registration of Multiphase Contrast-enhanced Abdomen CT Image
    Zhu, Zhiliang
    Li, Ji
    Jiang, Huiyan
    Zhang, Libo
    Yang, Benqiang
    2011 6TH INTERNATIONAL CONFERENCE ON COMPUTER SCIENCES AND CONVERGENCE INFORMATION TECHNOLOGY (ICCIT), 2012, : 909 - 914
  • [9] Automatic Alignment of Myocardial Perfusion Images With Contrast-Enhanced Cardiac Computed Tomography
    Faber, Tracy L.
    Santana, Cesar A.
    Piccinelli, Marina
    Nye, Jonathon A.
    Votaw, John R.
    Garcia, Ernest V.
    Haber, Eldad
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (05) : 2296 - 2302
  • [10] Co-registration of cardiac MRI and rest gated SPECT in the assessment of myocardial perfusion, function and viability
    Misko, J
    Dziuk, M
    Skrobowska, E
    Szalus, N
    Pietrzykowski, J
    Warczynska, A