Online gamma dose-rate measurements by means of a self-powered gamma detector

被引:8
|
作者
Van Nieuwenhove, R [1 ]
Vermeeren, L [1 ]
机构
[1] SCK CEN, B-2400 Mol, Belgium
关键词
dosimetry; gamma ray detectors; nuclear measurements; self-powered detectors;
D O I
10.1109/TNS.2002.801486
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Gamma dose-rate profiles have been measured by means of a self-powered gamma detector (SPGD) in a gamma irradiation facility using Co-60 sources, situated in the side pool of the BR2 materials testing reactor. Because of its small size (diameter of only 4 mm), the SPGD is ideally suited to perform very localized gamma dose-rate measurements. It has been shown that the SPGD can provide precise relative measurements of the local gamma dose rate in the range 0.5-20 kGy/h. A very detailed Monte Carlo code has been developed to relate the gamma dose rate (for a given gamma spectrum) to the current induced by the SPGD. The obtained conversion factor could thus be used to determine the absolute gamma dose rate.
引用
收藏
页码:1914 / 1918
页数:5
相关论文
共 50 条
  • [1] Development of a self-powered gamma detector
    Takeuchi, Tomoaki
    Ohtsuka, Noriaki
    Shibata, Akira
    Tsuchiya, Kunihiko
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2014, 51 (7-8) : 939 - 943
  • [2] On-line gamma flux measurements by means of a Self Powered Gamma Detector
    Van Nieuwenhove, R
    Vermeeren, L
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 827 - 831
  • [3] COMBINED SCINTILLATION DETECTOR FOR GAMMA DOSE-RATE MEASUREMENT
    VIERERBL, L
    NOVAKOVA, O
    JURSOVA, L
    RADIATION PROTECTION DOSIMETRY, 1990, 32 (04) : 273 - 277
  • [4] ANALYSIS OF SELF-POWERED GAMMA-RAY DETECTOR WITH DIRECTIONAL DISCRIMINATION
    LEVERT, FE
    BEYERLEIN, RA
    COX, SA
    NUCLEAR INSTRUMENTS & METHODS, 1979, 160 (01): : 61 - 72
  • [6] RESPONSE OF SELF-POWERED GAMMA DETECTORS
    STRINDEHAG, O
    ATOMKERNENERGIE, 1971, 17 (04): : 267 - +
  • [7] Study on the self-powered detectors for measuring the gamma-ray dose
    Li, Mingfei
    Jin, Shengren
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 1996, 30 (03): : 262 - 265
  • [8] Vanadium Self-Powered Detector Gamma Response in Commercial Pressurized Water Reactor
    Chen, Jianwei
    REACTOR DOSIMETRY: 16TH INTERNATIONAL SYMOSIUM, 2018, 1608 : 175 - 189
  • [9] SENSITIVITY OF SELF-POWERED DETECTOR PROBES TO ELECTRON AND GAMMA-RAY FIELDS
    LONE, MA
    WONG, PY
    AJMANI, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 349 (2-3): : 563 - 576
  • [10] ON CALCULATION OF DOSE-RATE OF SCATTERED GAMMA RADIATION
    WAECHTER, KH
    KERNTECHNIK ISOTOPENTECHNIK UND CHEMIE, 1967, 9 (02): : 62 - &