Design of a multi-detector, single line-of-sight, time-of-flight system to measure time-resolved neutron energy spectra

被引:2
|
作者
Schlossberg, D. J. [1 ]
Moore, A. S. [1 ]
Kallman, J. S. [1 ]
Lowry, M. [1 ]
Eckart, M. J. [1 ]
Hartouni, E. P. [1 ]
Hilsabeck, T. J. [1 ]
Kerr, S. M. [1 ]
Kilkenny, J. D. [2 ]
机构
[1] Lawrence Livermore Natl Lab, Livermore, CA 94551 USA
[2] Gen Atom, San Diego, CA 92121 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2022年 / 93卷 / 11期
关键词
Dynamic environments - Fusion plasmas - Inertial-confinement fusions - Line of Sight - Lines-of-sight - Multi-detectors - Neutron energy spectrum - Spectra's - Time-of-flight system - Time-resolved;
D O I
10.1063/5.0101874
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In the dynamic environment of burning, thermonuclear deuterium-tritium plasmas, diagnosing the time-resolved neutron energy spectrum is of critical importance. Strategies exist for this diagnosis in magnetic confinement fusion plasmas, which presently have a lifetime of similar to 10(12) longer than inertial confinement fusion (ICF) plasmas. Here, we present a novel concept for a simple, precise, and scale-able diagnostic to measure time-resolved neutron spectra in ICF plasmas. The concept leverages general tomographic reconstruction techniques adapted to time-of-flight parameter space, and then employs an updated Monte Carlo algorithm and National Ignition Facility-relevant constraints to reconstruct the time-evolving neutron energy spectrum. Reconstructed spectra of the primary 14.028 MeV n(DT) peak are in good agreement with the exact synthetic spectra. The technique is also used to reconstruct the time-evolving downscattered spectrum, although the present implementation shows significantly more error.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] MULTI-DETECTOR NEUTRON TIME-OF-FLIGHT SPECTROMETER
    DAVIS, JC
    MOELLER, TG
    POHL, BA
    ANDERSON, JD
    GRIMES, SM
    WONG, C
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1973, 18 (10): : 1424 - 1424
  • [2] Reconstruction of Time-Resolved Neutron Energy Spectra in Z-Pinch Experiments Using Time-of-flight Method
    Rezac, K.
    Klir, D.
    Kubes, P.
    Kravarik, J.
    DENSE Z-PINCHES, 2009, 1088 : 211 - 214
  • [3] Passive optical time-of-flight for non line-of-sight localization
    Jeremy Boger-Lombard
    Ori Katz
    Nature Communications, 10
  • [4] Passive optical time-of-flight for non line-of-sight localization
    Boger-Lombard, Jeremy
    Katz, Ori
    NATURE COMMUNICATIONS, 2019, 10 (1)
  • [5] The five line-of-sight neutron time-of-flight (nToF) suite on the National Ignition Facility (NIF)
    Moore, A. S.
    Hartouni, E. P.
    Schlossberg, D.
    Kerr, S.
    Eckart, M.
    Carrera, J.
    Ma, L.
    Waltz, C.
    Barker, D.
    Gjemso, J.
    Mariscal, E.
    Grim, G.
    Kilkenny, J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2021, 92 (02):
  • [7] A METHOD FOR CORRECTING TIME-OF-FLIGHT SPECTRA FOR NEUTRON EMISSION TIME AND DETECTOR EFFICIENCY
    DOULTREM.P
    NUCLEAR INSTRUMENTS & METHODS, 1970, 77 (01): : 1 - &
  • [8] DATA ACQUISITION-SYSTEM OF A TIME-OF-FLIGHT ANALYZER FOR USE IN A MULTI-DETECTOR ASSEMBLY FOR NEUTRON-SCATTERING EXPERIMENTS
    ONO, M
    TASAKI, S
    OKAMOTO, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 285 (03): : 477 - 483
  • [9] Fusion neutron energy spectra measured by time-of-flight spectrometers
    Elevant, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 476 (1-2): : 485 - 489
  • [10] MULTI-ANGLE DETECTOR SYSTEM FOR FAST-NEUTRON TIME-OF-FLIGHT SPECTROSCOPY
    ECKSTEIN, P
    HELFER, H
    KATZMER, D
    KAYSER, J
    KRAUSE, R
    LEHMANN, D
    MEILING, W
    PILZ, W
    RUMPF, J
    SCHMIDT, D
    SEELIGER, D
    STREIL, T
    NUCLEAR INSTRUMENTS & METHODS, 1980, 169 (03): : 533 - 538