Cold cathode gauges for ultrahigh vacuum measurements

被引:18
|
作者
Kendall, BRF [1 ]
Drubetsky, E [1 ]
机构
[1] FREDERICKS CO,TELEVAC DIV,HUNTINGDON VALLEY,PA 19006
来源
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS | 1997年 / 15卷 / 03期
关键词
D O I
10.1116/1.580813
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Nine cold cathode gauges have been evaluated on ion-pumped ultrahigh vacuum systems operating at pressures down to the 10(-11) Torr range. The test gauges included magnetrons, inverted magnetrons, and double inverted magnetrons from four different manufacturers as well as experimental variable-geometry gauges built especially for this project. Spinning-rotor and extractor gauges were used for calibration. The investigation covered repeated calibration over the 10(-10) to 10(-4) Torr range, stability over periods of up to 35 000 h of low pressure operation, tests for discontinuities in the current-pressure characteristics, stray magnetic field measurements, susceptibility to external magnetic fields, outgassing effects, and starting behavior at very low pressures. Our conclusion is that modern cold-cathode gauges are capable of giving far more accurate results than were possible with earlier Penning-type designs. Because of their extremely low outgassing rates, and their relative freedom from x-ray and electron stimulated desorption errors, they may in practice give results at low pressures which are more accurate than those obtained with typical hot-cathode gauges. (C) 1997 American Vacuum Society.
引用
收藏
页码:740 / 746
页数:7
相关论文
共 50 条