摘要
Stripping voltammetry analysis using a hemispheroidal mercury microelectrode has been previously assessed theoretically as a possible candidate 'primary method' for amount of substance measurement. It was shown that the measurement methodology can be described completely by a set of measurement equations when a novel quantification process is employed to measure the steady-state diffusion-limited deposition current over an optimum sampling time, instead of employing an analyte stripping step. The corollary is that, for solutions of known composition, the method may be used for highly accurate determination of diffusion coefficients. This investigation uses experimental data from steady-state diffusion-limited current measurements with sampling periods optimised using Allan deviation techniques, together with a full uncertainty analysis to estimate feasibility and likely accuracy of this method.