Research on a Novel High-Precision Methane Concentration Detection System

被引:0
|
作者
Song Lin-li [1 ]
Zhou Han-chang [1 ]
Zhang Zhi-jie [2 ]
机构
[1] North Univ China, Natl Key Lab Elect Measurement Technol, Taiyuan 030051, Peoples R China
[2] North Univ China, Key Lab Instrumentat Sci & Dynam Measurement, Minist Educ, Taiyuan 030051, Peoples R China
关键词
Spectroscopy; Differential ratio algorithm; Characteristic absorption conversion window; Methane concentration;
D O I
10.3964/j.issn.1000-0593(2014)12-3354-04
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In the gas concentration detection process using the characteristic spectrum absorption method, in order to improve the detection accuracy of the gas concentration, it often has to use the high-quality narrowband modulated laser and modulate wavelength to align with the characteristic absorption peaks of measured gas. But by this way, the cost of the laser and system requirements will be greatly increased. To use the existing portable, low-cost semiconductor laser conditions, at the same time it can obtain higher precision, conversion window differential absorption optical structure and the algorithm of differential characteristic absorption ratio was designed. Selection reason of position of the wavelength characteristic was analyzed, and steps to implement the processing algorithm were given. Systematically utilizing the combination method of conversion window and absorption gas chamber, by the method for calculating the ratio of the light intensity response, the light intensity from non-characteristic absorption peak position was divided out. So it achieved a similar detecting effect was achieved that used a narrow-band laser aligned to the feature absorption peak position. Experiments adopted MW-IR-1650 infrared laser, type SSM17-2 stepper motor control module, C30659 infrared detectors, and other devices. In the experiments, different concentrations of methane gas were tested, and experimental results show that the relative error of measurement was less than 2. 0% within the range from 200 to 5 000 ppm. In summary, it's proved that the system has high accuracy and stability.
引用
收藏
页码:3354 / 3357
页数:4
相关论文
共 13 条
  • [1] Modified point diffraction interferometer for inspection and evaluation of ophthalmic components
    Acosta, E
    Chamadoira, S
    Blendowske, R
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2006, 23 (03) : 632 - 637
  • [2] Photoacoustic spectroscopy of formaldehyde with tunable laser radiation at the parts per billion level
    Angelmahr, M.
    Miklos, A.
    Hess, P.
    [J]. APPLIED PHYSICS B-LASERS AND OPTICS, 2006, 85 (2-3): : 285 - 288
  • [3] [Anonymous], INFRARED
  • [4] [Anonymous], APPL OPTICS
  • [5] Cihelka J, 2007, ICTON MEDIT WIN CONF, P61
  • [6] All optical wavelength converter-reshaper based on multi-wavelengths spectral components generation in optical fibre
    Curti, F
    Matera, F
    Tosi-Beleffi, GM
    [J]. ICTON 2002: PROCEEDINGS OF THE 2002 4TH INTERNATIONAL CONFERENCE ON TRANSPARENT OPTICAL NETWORKS AND EUROPEAN SYMPOSIUM ON PHOTONIC CRYSTALS, VOL 1, 2002, : 66 - 68
  • [7] Detection of leak acoustic signal in buried gas pipe based on the time-frequency analysis
    Kim, Min-Soo
    Lee, Sang-Kwon
    [J]. JOURNAL OF LOSS PREVENTION IN THE PROCESS INDUSTRIES, 2009, 22 (06) : 990 - 994
  • [8] Wang Yanju, 2006, Chinese Journal of Scientific Instrument, V27, P1647
  • [9] [王玉田 Wang Yutian], 2003, [传感技术学报, Chinese Journal of Sensors and Actuators], V16, P324
  • [10] Wu Xijun, 2010, Chinese Journal of Sensors and Actuators, V23, P14