Development of CMM auxiliary equipment for tiny hole measuring

被引:1
|
作者
Chen Feng [1 ]
Bing Zhi-gang [1 ]
Li Shang-hui [1 ]
Cheng Li [1 ]
Zhang Zhi-quan [1 ]
Chen Jia [1 ]
Yang Yi [1 ]
机构
[1] Zhejiang Inst Qual Inspect Sci, Hangzhou 310027, Peoples R China
来源
关键词
Coordinate Measuring Machine (CMM); Tiny Hole; Auxiliary Equipment; Reverse Test; Precision Error;
D O I
10.4028/www.scientific.net/AMR.797.543
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
According to the limited measurement range of specific coordinate measuring machine (CMM), a reverse test method was put forward, and an auxiliary measuring equipment for tiny hole was developed. A measuring method for tiny hole was formulated, which using gauge head trigger signal of CMM as the test signal. The principle which was the tiny hole moving but not moving gauge head was used and the corresponding measurement software was designed so that the small measurement range was expanded. The measurement precision influence factors were analyzed and the measurement scheme was optimized. The results indicate that: then the measurement range of tiny hole is expanded to 1.0mm based on the combination of auxiliary equipment and CMM. The precision error is 0.004 mm which meets the measurement requirements.
引用
收藏
页码:543 / 548
页数:6
相关论文
共 50 条