High accuracy measurement of the relative efficiency curve and determination of gamma-ray relative intensity for 38Cl

被引:6
|
作者
Hayashi, N
Miyahara, H
Mori, C
Takeuchi, N
Iwamoto, S
Ishikawa, I
机构
[1] Nagoya Univ, Grad Sch Engn, Dept Nucl Engn, Chikusa Ku, Nagoya, Aichi 4648603, Japan
[2] Nagoya Univ, Sch Hlth Sci, Dept Radiol Technol, Higashi Ku, Nagoya, Aichi 4618673, Japan
[3] Aichi Inst Technol, Dept Elect, Toyota 4700356, Japan
[4] Japan Atom Energy Res Inst, Dept Radioisotopes, Tokai, Ibaraki 3191195, Japan
[5] Univ Michigan, Ann Arbor, MI 48109 USA
[6] Nagoya Univ, Radioisotope Ctr, Nagoya, Aichi, Japan
关键词
Chlorine - Gamma rays - Germanium - Particle detectors - Radioactivity measurement;
D O I
10.1016/S0969-8043(99)00237-7
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Relative efficiency curves were accurately determined for Ge gamma-ray detectors from measurements for radionuclides which emit two gamma-rays with essentially the same emission probabilities. The relative intensity of the 1643 and 2167 keV gamma-rays from Cl-38 was determined to be 0.7496 (11) by using these highly accurate efficiency curves, (C) 2000 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:733 / 737
页数:5
相关论文
共 29 条
  • [1] Precise measurement of the gamma-ray emission probabilities of 38Cl
    Miyahara, Hiroshi
    Yoshida, Atsushi
    Wurdiyanto, Gatot
    Hara, Kei
    Yanagida, Katsuo
    Mori, Chizuo
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996, 369 (2-3): : 494 - 497
  • [2] Gamma-ray energies in the decay of 38Cl
    Krane, K. S.
    Keck, M. L.
    Norman, E. B.
    Shivprasad, A. P.
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (04) : 740 - 742
  • [3] Highly precise measurement of the relative gamma-ray detection efficiency curve
    Hayashi, N
    Miyahara, H
    Mori, C
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2000, 37 (02) : 139 - 143
  • [4] BETA- AND GAMMA-RAY MEASUREMENTS ON DECAY OF 38CL
    VANKLINKEN, J
    PLEITER, F
    DIJKSTRA, HT
    NUCLEAR PHYSICS A, 1968, A112 (02) : 372 - +
  • [5] RELATIVE INTENSITY CALIBRATION OF A GE(LI) GAMMA-RAY SPECTROMETER
    JARDINE, LJ
    NUCLEAR INSTRUMENTS & METHODS, 1971, 96 (02): : 259 - &
  • [6] DETERMINATION OF RELATIVE PHOTOPEAK EFFICIENCY AND SYSTEM LINEARITY FOR GE(LI) GAMMA-RAY SPECTROMETERS
    GEHRKE, RJ
    CLINE, JE
    HEATH, RL
    NUCLEAR INSTRUMENTS & METHODS, 1971, 91 (03): : 349 - +
  • [7] DETERMINATION OF RELATIVE PHOTOPEAK EFFICIENCY AND SYSTEM LINEARITY FOR GE(LI) GAMMA-RAY SPECTROMETERS
    GEHRKE, RJ
    CLINE, JE
    HEATH, RL
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1969, 14 (12): : 1243 - &
  • [8] RELATIVE GAMMA-RAY INTENSITIES FOR GE(LI) EFFICIENCY DETERMINATIONS
    GEHRKE, RJ
    HELMER, RG
    GREENWOOD, RC
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1977, 22 (01): : 93 - 93
  • [9] Determination of the relative age for an intrusive body by gamma-ray spectrometry
    Qiu, Yuan-de
    Zhou, Rong-sheng
    Hou, Xin-sheng
    Fang, Fang
    Ma, Ying-jie
    Jia, Wen-yi
    Chengdu Ligong Xueyuan Xuebao/Journal of Chengdu Institute of Technology, 2000, 27 (02): : 193 - 196
  • [10] RELATIVE GAMMA-RAY INTENSITY MEASUREMENTS IN THE DECAY OF IR-192
    HAMILTON, JH
    HULTBERG, S
    FREY, WF
    ARKIV FOR FYSIK, 1962, 22 (06): : 485 - 490