On the determination of dynamic errors for rise time measurement with an oscilloscope

被引:52
|
作者
Mittermayer, C [1 ]
Steininger, A [1 ]
机构
[1] Vienna Tech Univ, Dept Measurement Technol, A-1040 Vienna, Austria
关键词
dynamic response; error analysis; measurement errors; oscilloscopes; time domain measurements;
D O I
10.1109/19.816121
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The dynamic behavior of an oscilloscope is usually described by its bandwidth, which is not sufficient for well-founded error estimation for rise time measurement. Two rules of thumb have become most common that describe a) a relation between bandwidth and rise time of an oscilloscope and b) an estimation of the error resulting from the rise time limitation of the oscilloscope. Although it is well understood that these rules are approximations, no information on their accuracy and extent of validity appears in the literature. In this paper, we analyze these rules of thumb with their underlying assumptions and respect a more realistic model for oscilloscope behavior. We give accurate error estimates for a selection of typical input signals, show how well the rules perform for these signals and that, although the rules provide good upper bounds for the error, they can not be reliably used for error correction.
引用
收藏
页码:1103 / 1107
页数:5
相关论文
共 50 条
  • [1] Sampling-oscilloscope Rise Time.
    Bonizzato, Maurizio
    Schiaffino, Paolo
    Energia Elettrica, 1982, 59 (09): : 397 - 401
  • [2] Versatile oscilloscope saves measurement time
    不详
    MICROWAVES & RF, 1997, 36 (06) : 128 - 128
  • [3] Compensation for dynamic measurement errors of time grating sensor
    School of Instrument Science and Optoelectronic Engineering, Hefei University of Technology, Hefei
    230009, China
    不详
    400054, China
    不详
    400054, China
    Guangxue Jingmi Gongcheng, 4 (1114-1121):
  • [4] ELIMINATING TIME-BASE ERRORS FROM OSCILLOSCOPE MEASUREMENTS
    MURPHY, JB
    ELECTRONICS, 1978, 51 (13): : 138 - 144
  • [5] Calibration of Oscilloscope Rise Time based on Pulse Synthesis Technique
    Yeung, C. F. Au
    Yang, S. L.
    Tsang, H. F.
    2024 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CPEM 2024, 2024,
  • [6] Numerical errors of the real time dynamic correction for the current measurement
    Nabielec, J
    Moronczyk, A
    COMPUTER STANDARDS & INTERFACES, 1999, 21 (04) : 335 - 340
  • [7] Oscilloscope influence on the calibration uncertainty of the pulse rise time of ESD simulators
    Sroka, J
    2003 IEEE International Symposium on Electromagnetic Compatibility (EMC), Vols 1 and 2, Symposium Record, 2003, : 378 - 381
  • [8] ERRORS OF TIME DETERMINATION
    SYMMS, LST
    PENNY, CJA
    MONTHLY NOTICES OF THE ROYAL ASTRONOMICAL SOCIETY, 1946, 106 (05) : 390 - &
  • [9] Errors in dynamic force measurement
    Dixon, M.
    Strain, 1988, 24 (04) : 139 - 142
  • [10] Influence of measurement errors on temperature-based death time determination
    Michael Hubig
    Holger Muggenthaler
    Gita Mall
    International Journal of Legal Medicine, 2011, 125 : 503 - 517