Design and Evaluation of a Capacitively Coupled Sensor Readout Circuit, toward Contact-less ECG and EEG

被引:18
|
作者
Svard, Daniel [1 ]
Cichocki, Andrzej [2 ]
Alvandpour, Atila [1 ]
机构
[1] Linkoping Univ, Dept Elect Engn, Div Elect Devices, S-58183 Linkoping, Sweden
[2] Riken Brain Sci Inst, Lab Adv Brain Signal Proc, Wako, Saitama, Japan
关键词
Biomedical transducers; Electrocardiography; Electroencephalograhpy; Biomedical signal detection;
D O I
10.1109/BIOCAS.2010.5709631
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Electrophysiological signal acquisition such as ECG and EEG play an important part in modern medical monitoring and diagnostics. The measurement of these very low-level, lowfrequency signals are normally made from the skin with a directly coupled sensor utilizing a conductive gel to create a low resistance path for the charge. The application of the gel is tedious and time consuming as well as requiring a clinical environment and prevents long period measurements. In this paper, a contact-less, capacitively coupled sensor without any need for gel-together with an electronic readout circuit using a PCB is presented. A design with a very high input impedance allows for measurements of signals with amplitudes down to a few tens of microvolts and at frequencies between a few hertz to tens of hertz. Measurements show that the circuit could detect an input signal of 25 mu V at 10 Hz with an SNR of 9.7 dB.
引用
收藏
页码:302 / 305
页数:4
相关论文
共 6 条
  • [1] An Integrated Micromachined Thermopile Sensor With a Chopper Interface Circuit for Contact-Less Temperature Measurements
    Moisello, Elisabetta
    Vaiana, Michele
    Castagna, Maria Eloisa
    Bruno, Giuseppe
    Malcovati, Piero
    Bonizzoni, Edoardo
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2019, 66 (09) : 3402 - 3413
  • [2] Active integrated antenna design using a contact-less, proximity coupled, differentially fed technique
    Lee, Ee
    Chan, Kin Meng
    Gardner, Peter
    Dodgson, Terence E.
    IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2007, 55 (02) : 267 - 276
  • [3] An Area-Efficient Capacitively-Coupled Sensor Readout Circuit with Current-Splitting OTA and FIR-DAC
    Tu, Chih-Chan
    Lee, Feng-Wen
    Chen, Han-Chun
    Wang, Yu-Kai
    Lin, Tsung-Hsien
    2017 IEEE ASIAN SOLID-STATE CIRCUITS CONFERENCE (A-SSCC), 2017, : 57 - 60
  • [4] FORENSIC SENSOR QUALITY EVALUATION: TOWARDS OBJECTIVE VISUAL QUALITY INDICES FOR CONTACT-LESS FINGERPRINT AND FIBER TRACES
    Kiltz, Stefan
    Dittmann, Jana
    Vielhauer, Claus
    2ND INTERNATIONAL WORKSHOP ON BIOMETRICS AND FORENSICS (IWBF2014), 2014,
  • [5] The Design and Evaluation of an End-User-Deployable, Whole House, Contact less Power Consumption Sensor
    Patel, Shwetak N.
    Gupta, Sidhant
    Reynolds, Matthew S.
    CHI2010: PROCEEDINGS OF THE 28TH ANNUAL CHI CONFERENCE ON HUMAN FACTORS IN COMPUTING SYSTEMS, VOLS 1-4, 2010, : 2471 - +
  • [6] Readout circuit based on double voltage feedback loops in the two-dimensional resistive sensor array: design, modelling and simulation evaluation
    Wu, Jian-Feng
    Li, Jian-Qing
    Song, Ai-Guo
    IET SCIENCE MEASUREMENT & TECHNOLOGY, 2017, 11 (03) : 288 - 296