Simultaneous measurement of noise and spatial resolution in PET phantom images

被引:28
|
作者
Lodge, Martin A. [1 ]
Rahmim, Arman [1 ]
Wahl, Richard L. [1 ,2 ]
机构
[1] Johns Hopkins Univ, Sch Med, Div Nucl Med, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ, Sch Med, Sidney Kimmel Comprehens Canc Ctr Johns Hopkins, Baltimore, MD USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2010年 / 55卷 / 04期
关键词
ACQUISITION;
D O I
10.1088/0031-9155/55/4/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As an aid to evaluating image reconstruction and correction algorithms in positron emission tomography, a phantom procedure has been developed that simultaneously measures image noise and spatial resolution. A commercially available Ge-68 cylinder phantom (20 cm diameter) was positioned in the center of the field-of-view and two identical emission scans were sequentially performed. Image noise was measured by determining the difference between corresponding pixels in the two images and by calculating the standard deviation of these difference data. Spatial resolution was analyzed using a Fourier technique to measure the extent of the blurring at the edge of the phantom images. This paper addresses the noise aspects of the technique as the spatial resolution measurement has been described elsewhere. The noise measurement was validated by comparison with data obtained from multiple replicate images over a range of noise levels. In addition, we illustrate how simultaneous measurement of noise and resolution can be used to evaluate two different corrections for random coincidence events: delayed event subtraction and singles-based randoms correction. For a fixed number of iterations of the maximum-likelihood expectation-maximization algorithm, the singles-based correction gave rise to higher noise than delayed event subtraction. However, when noise and resolution were measured simultaneously it was shown that singles-based randoms correction gave rise to lower noise than delayed event subtraction for a fixed spatial resolution. The proposed method of simultaneously measuring image noise and spatial resolution is useful for evaluating reconstruction algorithms and may aid standardization of data collection between centers.
引用
收藏
页码:1069 / 1081
页数:13
相关论文
共 50 条
  • [1] Simultaneous estimation of bias and resolution in PET images with a single long-lived phantom
    Byrd, Darrin
    Wangerin, Kristen
    Helba, Brian
    Liu, Xiaoxiao
    Kinahan, Paul
    Avila, Rick
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [2] Simultaneous estimation of bias and resolution in PET images with a single long-lived phantom
    Byrd, Darrin
    Wangerin, Kristen
    Helba, Brian
    Liu, Xiaoxiao
    Kinahan, Paul
    Avila, Rick
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [3] Simultaneous Estimation of Bias and Resolution in PET Images With a Long-Lived "Pocket" Phantom System
    Kinahan, Paul E.
    Byrd, Darrin W.
    Helba, Brian
    Wangerin, Kristen A.
    Liu, Xiaoxiao
    Levy, Joshua R.
    Allberg, Keith C.
    Krishnan, Karthik
    Avila, Ricardo S.
    TOMOGRAPHY, 2018, 4 (01) : 33 - 41
  • [4] About Measurement of PET Spatial Resolution
    Lodge, Martin A.
    Sunderland, John J.
    Rahmim, Arman
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [5] PET SCANNER SPATIAL RESOLUTION FROM CONTRAST PHANTOM MEASUREMENTS
    O'Keefe, G.
    Dore, V.
    Williams, R.
    Johnston, L.
    Rowe, C.
    INTERNAL MEDICINE JOURNAL, 2019, 49 : 39 - 39
  • [6] Noise and physical limits to maximum resolution of PET images
    Herraiz, J. L.
    Espana, S.
    Vicente, E.
    Vaquero, J. J.
    Desco, M.
    Udias, J. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (02): : 934 - 937
  • [7] Satisfying simultaneous resolution and noise criteria in digital images
    Shaw, R
    PICS 2000: IMAGE PROCESSING, IMAGE QUALITY, IMAGE CAPTURE, SYSTEMS CONFERENCE, PROCEEDINGS, 2000, : 71 - 74
  • [8] Quantification of gamma camera spatial resolution by means of bar phantom images
    Wasserman, HJ
    NUCLEAR MEDICINE COMMUNICATIONS, 1998, 19 (11) : 1089 - 1097
  • [9] Feasibility of Total Variation Noise Reduction Algorithm According to Various MR-Based PET Images in a Simultaneous PET/MR System: A Phantom Study
    Park, Chan-Rok
    Kang, Seong-Hyeon
    Lee, Young-Jin
    DIAGNOSTICS, 2021, 11 (02)
  • [10] Development of a computational phantom for validation of automated noise measurement in CT images
    Anam, Choirul
    Sutanto, Heri
    Adi, Kusworo
    Budi, Wahyu Setia
    Muhlisin, Zaenul
    Haryanto, Freddy
    Matsubara, Kosuke
    Fujibuchi, Toshioh
    Dougherty, Geoff
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (06)