Biomechanical analysis of arthroscopically assisted latissimus dorsi transfer fixation for irreparable posterosuperior rotator cuff tears-Knotless versus knotted anchors

被引:6
|
作者
Ernstbrunner, Lukas [1 ]
Borbas, Paul [1 ]
Rohner, Marco [1 ]
Brun, Sascha [1 ]
Bachmann, Elias [1 ]
Bouaicha, Samy [1 ]
Wieser, Karl [1 ]
机构
[1] Univ Zurich, Balgrist Univ Hosp, Dept Orthoped, Forchstr 340, CH-8008 Zurich, Switzerland
关键词
arthroscopically assisted; biomechanics; latissimus dorsi transfer; rotator cuff; TENDON TRANSFER; FAILURE; BONE;
D O I
10.1002/jor.24963
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We compared the construct stability of traditional knotted techniques with modern knotless anchor systems used in arthroscopically assisted latissimus dorsi transfers (aLDTs) for irreparable posterosuperior rotator cuff tears. Eighteen cadaveric shoulders were age- and gender-matched to the following three groups: knotted group (two knotted anchors); knotless group (two knotless anchors); and triple anchor group (two knotless anchors; one all-suture anchor at the apex of the humeral head). All tendons were cyclically loaded in line of the aLDT over 400 cycles followed by a load to failure test. Outcome measures were the ultimate load to failure, elongation, construct stiffness, and failure modes. The triple anchor group revealed the highest ultimate load to failure (431 +/- 78 N) compared with the knotted (326 +/- 52 N; p = .022) and knotless (353 +/- 105 N; p = .129) groups. Total elongation and construct stiffness were not significantly different comparing all three groups. The failure modes were: anchor pull-out in all specimen of the knotted group; three (50%) anchor pull-out and three suture pull-out in the knotless group (p = .046); four (67%) anchor pull-out, one (17%) suture pull-out and one tendon pull-out in the triple anchor group. Biomechanical analyses of knotless fixation techniques for aLDTs show similar construct stability and elongation compared with the traditional, knotted techniques. Bone fixation seems to be improved with the knotless anchor systems. In our practice, we continue to use the knotless fixation technique for aLDT and in the situation of weak, osteoporotic bone, we add a third (all-suture) anchor to improve construct stability.
引用
收藏
页码:2234 / 2242
页数:9
相关论文
共 50 条
  • [1] Arthroscopically Assisted Latissimus Dorsi Transfer for Irreparable Rotator Cuff Tears
    Millett, Peter J.
    Yen, Yi-Meng
    Huang, Michael J.
    TECHNIQUES IN SHOULDER AND ELBOW SURGERY, 2008, 9 (02): : 76 - 79
  • [2] Mid-term results of arthroscopically assisted latissimus dorsi transfer for irreparable posterosuperior rotator cuff tears
    Waltenspul, Manuel
    Jochum, Benedikt
    Filli, Lukas
    Ernstbrunner, Lukas
    Wieser, Karl
    Meyer, Dominik
    Gerber, Christian
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2021, 30 (11) : E676 - E688
  • [3] Biomechanics of latissimus dorsi transfer for irreparable posterosuperior rotator cuff tears
    Ling, H. Y.
    Angeles, I. G.
    Horodyski, M. B.
    CLINICAL BIOMECHANICS, 2009, 24 (03) : 261 - 266
  • [4] Mid-Term Results of Arthroscopically-Assisted Latissimus Dorsi Transfer for Irreparable Posterosuperior Rotator Cuff Tears
    Waltenspul, Manuel
    Jochum, Benedikt
    Filli, Lukas
    Ernstbrunner, Lukas
    Wieser, Karl
    Meyer, Dominik
    Gerber, Christian
    SWISS MEDICAL WEEKLY, 2021, 151 : 4S - 5S
  • [5] Clinical and Radiographic Outcomes With Assessment of the Learning Curve in Arthroscopically Assisted Latissimus Dorsi Tendon Transfer for Irreparable Posterosuperior Rotator Cuff Tears
    Yamakado, Kotaro
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2017, 33 (12): : 2144 - 2151
  • [6] Efficacy of latissimus dorsi and teres major tendon transfer in anterosuperior irreparable rotator cuff tears and posterosuperior irreparable rotator cuff tears
    Baek, Chang Hee
    Lim, Chaemoon
    Kim, Jung Gon
    Kim, Bo Taek
    Kim, Seung Jin
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2025, 20 (01):
  • [7] Clinic and electromyographic results of latissimus dorsi transfer for irreparable posterosuperior rotator cuff tears
    De Casas, Ricardo
    Lois, Matias
    Cidoncha, Myriam
    Valadron, Miguel
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2014, 9 : 83
  • [8] Clinic and electromyographic results of latissimus dorsi transfer for irreparable posterosuperior rotator cuff tears
    Ricardo De Casas
    Matías Lois
    Myriam Cidoncha
    Miguel Valadron
    Journal of Orthopaedic Surgery and Research, 9
  • [9] Tendon fixation in arthroscopic latissimus dorsi transfer for irreparable posterosuperior cuff tears: An in vitro biomechanical comparison of interference screw and suture anchors
    Diop, Amadou
    Maurel, Nathalie
    Chang, Vivian K.
    Kany, Jean
    Duranthon, Louis-Denis
    Grimberg, Jean
    CLINICAL BIOMECHANICS, 2011, 26 (09) : 904 - 909
  • [10] Association of Arthroscopically-Assisted Latissimus Dorsi Tendon Transfer with Implantation of a Subacromial Balloon Spacer for Patients with Irreparable Posterosuperior Rotator Cuff Tears
    Gervasi, E.
    Vigni, G. E.
    Vezeridis, P. S.
    Tomasi, A.
    Sabbioni, G.
    Fazzari, F.
    Camarda, L.
    MLTJ-MUSCLES LIGAMENTS AND TENDONS JOURNAL, 2022, 12 (04): : 497 - 510