Positioning of a plane-parallel ionization chamber in clinical electron beams and the impact on perturbation factors

被引:24
|
作者
Zink, K. [1 ]
Wulff, J. [1 ]
机构
[1] Univ Appl Sci Giessen Friedberg, Inst Med Phys & Strahlenschutz IMPS, D-35390 Giessen, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 08期
关键词
MONTE-CARLO-SIMULATION; STOPPING-POWER RATIOS; QUALITY CORRECTION FACTORS; ION CHAMBERS; MEGAVOLTAGE PHOTON; DOSIMETRY; PLATE; TRANSPORT; PENELOPE; PROTOCOL;
D O I
10.1088/0031-9155/54/8/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Current dosimetry protocols recommend the use of plane-parallel ionization chambers for the dosimetry of clinical electron beams. The necessary perturbation corrections p(wall) and p(cav) are assumed to be unity, independent of the depth of measurement and the energy of the primary electrons. To verify these assumptions detailed Monte Carlo studies of a Roos chamber in clinical electron beams with energies in the range of 6-21 MeV are performed at different depths in water and analyzed in terms of Spencer-Attix cavity theory. Separate simulations for the perturbation corrections p(wall) and p(cav) indicate quite different properties of both correction factors with depth. Dose as well as fluence calculations show a nearly depth-independent wall correction factor for a shift of the Roos chamber Delta z = -0.017 cm toward the focus. This value is in good agreement with the positioning recommendation given in all dosimetry protocols. Regarding the fluence perturbation p(cav) the simulation of the electron fluence inside the air cavity in comparison to water unambiguously reveals an in-scattering of low energy electrons, despite the fact, that the cavity is 'well guarded'. For depths beyond the reference depth z(ref) this effect is superimposed by an increased loss of primary electrons from the beam resulting in p(cav) > 1. This effect is largest for low electron energies but present for all electron energies involved in this study. Based on the different depth dependences of p(wall) and p(cav) it is possible to choose a chamber shift Delta z in a way to minimize the depth dependence of the overall perturbation factor p. For the Roos chamber this shift is Delta z = -0.04 cm independent of electron energy.
引用
收藏
页码:2421 / 2435
页数:15
相关论文
共 50 条
  • [1] Perturbation factors for the plane-parallel NACP02 ionisation chamber in electron beams
    Zakhikani, R
    Seuntjens, J
    DuSautoy, A
    Shipley, D
    Bostock, G
    Verhaegen, F
    MEDICAL PHYSICS, 2005, 32 (06) : 2007 - 2007
  • [2] Perturbation correction factors for the NACP-02 plane-parallel ionization chamber in water in high-energy electron beams
    Verhaegen, F
    Zakikhani, R
    DuSautoy, A
    Palmans, H
    Bostock, G
    Shipley, D
    Seuntjens, J
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (05): : 1221 - 1235
  • [3] Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams
    Berg, M
    Noerrevang, O
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (23): : 5309 - 5318
  • [4] Development of plane-parallel ionization chamber
    Chen, Yi-Zhen
    Lin, Min
    Su, Dan
    Meng, Xian-Fang
    Xia, Wen
    Chen, Ke-Sheng
    Xu, Li-Jun
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2008, 42 (SUPPL.): : 380 - 384
  • [5] Calculation of water and graphite perturbation factors for NACP-02 plane-parallel chamber in high-energy electron beams
    Chin, E.
    Shipley, D.
    Seuntfens, J.
    Palmans, H.
    DuSautoy, A.
    Bailey, M.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S54 - S54
  • [6] Investigating the impact of the effective point of measurement for plane-parallel ionization chambers in clinical proton beams
    Baumann, Kilian-Simon
    Lourenco, Ana
    Wulff, Joerg
    Vilches-Freixas, Gloria
    Palmans, Hugo
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (01):
  • [7] Plane-parallel ionization chamber response in the buildup region of obliquely incident photon beams
    Gerbi, BJ
    Khan, FM
    MEDICAL PHYSICS, 1997, 24 (06) : 873 - 878
  • [8] Analysis of Dose Perturbation Factors for An NACP-02 Parallel-Plate Ionization Chamber in Clinical Electron Beams
    Chin, E.
    Seuntjens, J.
    Bailey, M.
    Shipley, D.
    Palmans, H.
    Dusautoy, A.
    Verhaegen, F.
    MEDICAL PHYSICS, 2008, 35 (06) : 2953 - +
  • [9] THE CALIBRATION AND USE OF PLANE-PARALLEL IONIZATION CHAMBERS FOR DOSIMETRY OF ELECTRON-BEAMS
    ALMOND, PR
    XU, ZG
    PARK, HC
    MEDICAL PHYSICS, 1995, 22 (08) : 1307 - 1314
  • [10] A prototype for a plane-parallel ionization chamber for β radiation detection
    Bercea, S.
    Radulescu, L.
    Celarel, A.
    Cenusa, C.
    Miron, C.
    Scarlat, F.
    Scarisoreanu, A.
    OPTOELECTRONICS AND ADVANCED MATERIALS-RAPID COMMUNICATIONS, 2011, 5 (3-4): : 305 - 308