Evaluation of auto-planning in IMRT and VMAT for head and neck cancer

被引:29
|
作者
Ouyang, Zi [1 ]
Shen, Zhilei Liu [1 ]
Murray, Eric [1 ]
Kolar, Matt [1 ]
LaHurd, Danielle [1 ]
Yu, Naichang [1 ]
Joshi, Nikhil [1 ]
Koyfman, Shiomo [1 ]
Bzdusek, Karl [2 ]
Xia, Ping [1 ]
机构
[1] Cleveland Clin, Dept Radiat Oncol, Cleveland, OH 44106 USA
[2] Philips Healthcare Inc, Fitchburg, WI USA
来源
关键词
auto-planning; head and neck; IMRT; step and shoot; VMAT; QUALITY; DVH;
D O I
10.1002/acm2.12652
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purposes of this work are to (a) investigate whether the use of auto-planning and multiple iterations improves quality of head and neck (HN) radiotherapy plans; (b) determine whether delivery methods such as step-and-shoot (SS) and volumetric modulated arc therapy (VMAT) impact plan quality; (c) report on the observations of plan quality predictions of a commercial feasibility tool. Materials and methods Twenty HN cases were retrospectively selected from our clinical database for this study. The first ten plans were used to test setting up planning goals and other optimization parameters in the auto-planning module. Subsequently, the other ten plans were replanned with auto-planning using step-and-shoot (AP-SS) and VMAT (AP-VMAT) delivery methods. Dosimetric endpoints were compared between the clinical plans and the corresponding AP-SS and AP-VMAT plans. Finally, predicted dosimetric endpoints from a commercial program were assessed. Results All AP-SS and AP-VMAT plans met the clinical dose constraints. With auto-planning, the dose coverage of the low dose planning target volume (PTV) was improved while the dose coverage of the high dose PTV was maintained. Compared to the clinical plans, the doses to critical organs, such as the brainstem, parotid, larynx, esophagus, and oral cavity were significantly reduced in the AP-VMAT (P < 0.05); the AP-SS plans had similar homogeneity indices (HI) and conformality indices (CI) and the AP-VMAT plans had comparable HI and improved CI. Good agreement in dosimetric endpoints between predictions and AP-VMAT plans were observed in five of seven critical organs. Conclusion With improved planning quality and efficiency, auto-planning module is an effective tool to enable planners to generate HN IMRT plans that are meeting institution specific planning protocols. DVH prediction is feasible in improving workflow and plan quality.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 50 条
  • [1] Assessment of DVH Prediction Model and Auto-Planning Module for Head and Neck VMAT Planning
    Tao, C.
    Liu, B.
    Li, C.
    Zhu, J.
    Lu, J.
    Yin, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E788 - E788
  • [2] Head & Neck VMAT Auto-Planning in Pinnacle. A class solution.
    Van Kollenburg, P. G. M.
    Bouwmans, L. C. W.
    Kusters, J. M. A. M.
    Brunenberg, E. J. L.
    Kunze-Busch, M. C.
    Dijkema, T.
    Kaanders, J. H. A. M.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S602 - S603
  • [3] Automatic IMRT and VMAT Treatment Planning for Head and Neck Cancer
    Shen, Z. L.
    Murray, E.
    Yu, N.
    Kolar, M. D.
    Greskovich, J. F., Jr.
    Koyfman, S.
    Bzdusek, K.
    Xia, P.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E560 - E561
  • [4] Evaluation of auto-planning in VMAT for locally advanced nasopharyngeal carcinoma
    Chen Jihong
    Chen Kaiqiang
    Dai Yitao
    Zhang Xiuchun
    Chen Yanyu
    Bai Penggang
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [5] Evaluation of auto-planning in VMAT for locally advanced nasopharyngeal carcinoma
    Chen Jihong
    Chen Kaiqiang
    Dai Yitao
    Zhang Xiuchun
    Chen Yanyu
    Bai Penggang
    Scientific Reports, 12
  • [6] Evaluation of the clinical application of Auto-Planning module for IMRT plans of left breast cancer
    Liu, Fang-Yu
    Dong, Zhi-Wei
    Yang, Hui-Bin
    Shi, Hong-Yun
    RADIATION PHYSICS AND CHEMISTRY, 2020, 166
  • [7] Evaluation of Novel IMRT Auto-Planning Tool for Nasopharyngeal Carcinoma Cases
    Jeong, K.
    Bzdusek, K.
    Kumar, P.
    Tome, W.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] A Comparison of IMRT and VMAT Treatment Planning for Head and Neck Cancer With NTCP/TCP Analysis
    Kryger, M.
    Wang, W.
    Wu, S. P. P.
    Xue, J.
    Cheng, P.
    Das, I. J.
    Hu, K. S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E680 - E680
  • [9] Auto-planning for VMAT accelerated partial breast irradiation
    Marrazzo, Livia
    Meattini, Icro
    Arilli, Chiara
    Calusi, Silvia
    Casati, Marta
    Talamonti, Cinzia
    Livi, Lorenzo
    Pallotta, Stefania
    RADIOTHERAPY AND ONCOLOGY, 2019, 132 : 85 - 92
  • [10] Head-And-Neck IMRT Auto-Planning Through Fluence Map Prediction Using Progressive Growing of Generative Adversarial Networks
    Li, X.
    Wu, Q. J.
    Wu, Q.
    Wang, C.
    Sheng, Y.
    Wang, W.
    Stephens, H.
    Yin, F.
    Ge, Y.
    MEDICAL PHYSICS, 2021, 48 (06)