An Improved Method for Energy Calibration of Photon Counting X-ray Detectors

被引:0
|
作者
Lee, Jeong Seok [1 ,2 ]
Shin, Ki-Young [1 ]
Kang, Dong-Goo [1 ]
Lee, Soo Yeol [2 ]
Jin, Seung Oh [1 ]
Kim, Insoo S. [3 ]
Hegazy, Mohamed A. A. [2 ]
Kim, Kisoo [2 ]
机构
[1] Korea Electrotechnol Res Inst, RSS Ctr, Ansan 15588, South Korea
[2] Kyung Hee Univ, Dept Biomed Engn, Yongin 17104, South Korea
[3] Korea Electrotechnol Res Inst, Appl Electromagnet Wave Res Ctr, Ansan 15588, South Korea
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Photon counting X-ray detectors with energy discrimination capabilities can discriminate the energy of incident X-ray photon, thereby spectral imaging is possible with them. Energy calibration is a process to find a linear mapping function between the amplitudes of electric pulses and their corresponding photon energy. For practical and efficient energy calibration, the X-ray tube voltage can be used as a reference. However, as the exposure time increases, the threshold giving 50 % of OFF pixels also increases without converging to a point. To reduce calibration errors due to this non-convergent behavior, we present an improved method for energy calibration with applying it to two commercial photon counting X-ray detectors, i.e., a CdTe Medipix2 detector and a CdTe Pixirad2 detector with PIXIE-II read-out ASIC. With the improved mapping function estimated by the improved method, we can better identify the iodine component in a phantom consisting of iodine and other components.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] An X-Ray Fluorescence Technique for Energy Calibration of Photon Counting Detectors
    Ding, H.
    Choi, H.
    Barber, W.
    Iwanczyk, J.
    Molloi, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 476 - 476
  • [2] Novel, photon counting X-ray detectors
    Durst, RD
    Diawara, Y
    Khazins, D
    Medved, S
    Becker, B
    Thorson, T
    POWDER DIFFRACTION, 2003, 18 (02) : 103 - 105
  • [3] Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study
    Cho, H-M
    Ding, H.
    Ziemer, B. P.
    Molloi, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (23): : 7211 - 7227
  • [4] Energy response of AlGaAs soft X-ray photon counting detectors
    Lees, J. E.
    Bugby, S. L.
    Boie, L.
    Scoular, C.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [5] Correlated counting and energy resolving properties of photon counting X-ray detectors like the Medipix detectors
    Boehnel, Michael
    Bartl, Peter
    Michel, Thilo
    Anton, Gisela
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2732 - 2735
  • [6] Energy calibration of photon counting detectors using x-ray tube potential as a reference for material decomposition applications
    Das, Mini
    Kandel, Bigyan
    Park, Chan Soo
    Liang, Zhihua
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [7] Photon counting x-ray detectors as scatter probes
    Lahoud, Elias
    Zaffrani, Yael
    Yagil, Yoad
    MEDICAL PHYSICS, 2024, 51 (01) : 93 - 102
  • [8] Beyond single photon counting X-ray detectors
    Bergamaschi, A.
    Dinapoli, R.
    Henrich, B.
    Johnson, I.
    Mozzanica, A.
    Shi, X.
    Schmitt, B.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 628 (01): : 238 - 241
  • [9] Pulse Splitting in Photon Counting X-Ray Detectors
    Engel, Klaus Juergen
    Steadman, Roger
    Herrmann, Christoph
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4727 - 4734
  • [10] Photon-counting x-ray detectors for CT
    Danielsson, Mats
    Persson, Mats
    Sjolin, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03):