Comparison amongst pulse sequences for enhanced contrast to noise ratio in magnetic resonance imaging

被引:0
|
作者
Amin, Naima [1 ]
Afzal, Rao Muhammad [3 ]
Yousaf, Muhammad [2 ]
Javid, Muhammad Arshad [4 ]
机构
[1] COMSATS Inst Informat Technol, Dept Phys, Lahore, Pakistan
[2] COMSATS Inst Informat Technol, Dept Math, Lahore, Pakistan
[3] Islamia Univ Bahawalpur, Dept Phys, Bahawalpur, Pakistan
[4] Univ Engn & Technol, Dept Basic Sci, Texla, Pakistan
关键词
Contrast to noise ratio; Pulse sequences; Magnetic resonance imaging; Imaging parameters; T1-T2 weighted images; PARAMETERS; MRI; GEL; MOUSE; T-2;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To provide optimised pulse sequence and imaging protocols for contrast-to-noise ratio and for tissues that have different signal intensities in magnetic resonance imaging. Methods: A tissue equivalent material, ferrous benzoic xylenol orange gel, was prepared using gelatine, ferrous ammonium sulfate, sulfuric acid, xylenol orange tetrasodium salt and benzoic acid. The gel was irradiated using 6 MV photons from a Varian Clinac 600C linear accelerator, with a dose of 5, 10, 15, 20 and 25 gray. Experimental variations in imaging parameters were performed in echo time and repetition time. The quantitative analysis consisted of contrast-to-noise ratio. Results: Conventional spin echo and fast spin echo were equivalent for the tissues of comparable signal intensities and for entities moderate difference between signal intensities. Conventional spin echo provided remarkable contrast for tissues where signal intensity difference was extremely high in T1, T2-weighted study. An appropriate inversion time of fast fluid attenuated inversion recovery made it significant to measure contrast between tissues where signal intensity difference was the smallest and ordinary. Conclusion: Choice of pulse sequence and parameters played a vital role in developing fine image contrast.
引用
收藏
页码:225 / 232
页数:8
相关论文
共 50 条
  • [1] Pulse sequences for contrast-enhanced magnetic resonance imaging
    Graves, Martin J.
    RADIOGRAPHY, 2007, 13 : E20 - E30
  • [2] THE VARIATION OF SIGNAL, CONTRAST AND CONTRAST-TO-NOISE RATIO IN NUCLEAR-MAGNETIC-RESONANCE FAST IMAGING SEQUENCES
    LEACH, MO
    HIND, AJ
    GOWLAND, PA
    BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (716): : 788 - 788
  • [3] Signal-to-noise ratio, contrast-to-noise ratio and pharmacokinetic modeling considerations in dynamic contrast-enhanced magnetic resonance imaging
    Li, Xin
    Huang, Wei
    Rooney, William D.
    MAGNETIC RESONANCE IMAGING, 2012, 30 (09) : 1313 - 1322
  • [4] THE CONTRAST-TO-NOISE RATIO AS A MEASURE OF TISSUE CONTRAST IN MAGNETIC-RESONANCE-IMAGING
    BAIERL, P
    SEIDERER, M
    HEYWANG, S
    RATH, M
    DIGITALE BILDDIAGNOSTIK, 1986, 6 (03): : 101 - 110
  • [5] Noninvasive coronary angiography using contrast-enhanced magnetic resonance imaging and electron beam computed tomography: Comparison of the contrast-to-noise ratio
    Ropers, D
    Achenbach, S
    Regenfus, M
    Birke, S
    Daniel, WG
    Moshage, W
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (02) : 415A - 415A
  • [6] Comparison of Pulse Sequences of Magnetic Resonance Imaging for Optimization of Timing and Image Quality
    Amin N.
    Yousaf M.
    Javid M.A.
    Farooq A.
    Iranian Journal of Medical Physics, 2020, 17 (06) : 350 - 358
  • [7] Comparison of myocardial viability assessed with contrast enhanced magnetic resonance imaging and nuclear imaging
    Roes, Stijntje D.
    Kaandorp, Theodorus A.
    Henneman, Maureen M.
    Ajmone, Nina
    Westenberg, Jos J.
    Stokkel, Marcel
    Lamb, Hildo J.
    van der Wall, Ernst E.
    de Roos, Albert
    Box, Jeroen J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (10) : A171 - A171
  • [8] IMAGE-CONTRAST AND PULSE SEQUENCES IN URINARY-TRACT MAGNETIC-RESONANCE-IMAGING
    NEWHOUSE, JH
    UROLOGIC RADIOLOGY, 1986, 8 (03) : 120 - 126
  • [9] Contrast Enhanced Magnetic Resonance Imaging in Sacroiliitis
    Tosun, Ozgur
    Tangal, Gulden Saglamer
    Cay, Nurdan
    Maras, Yuksel
    Ercan, Karabekir
    Arslan, Halil
    IRANIAN JOURNAL OF RADIOLOGY, 2017, 14 (01)
  • [10] Contrast Enhanced Magnetic Resonance Imaging of the Liver
    Noterdaeme, Olivier
    Brady, Michael
    2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 831 - 834