Measurement of end-expiratory lung volume in intubated children without interruption of mechanical ventilation

被引:12
|
作者
Bikker, Ido G. [1 ]
Scohy, Thierry V. [2 ]
Bogers, Ad J. J. C. [3 ]
Bakker, Jan [1 ]
Gommers, Diederik [1 ,2 ]
机构
[1] Erasmus MC, Dept Intens Care Med, NL-3015 CE Rotterdam, Netherlands
[2] Erasmus MC, Dept Anesthesiol, NL-3015 CE Rotterdam, Netherlands
[3] Erasmus MC, Dept Cardiothorac Surg, NL-3015 CE Rotterdam, Netherlands
关键词
Functional residual capacity; Pediatric; Positive-pressure respiration; Lung volume measurements; Anesthesiology; FUNCTIONAL RESIDUAL CAPACITY; COMPUTED-TOMOGRAPHY; HELIUM DILUTION; NITROGEN WASHOUT/WASHIN; REFERENCE VALUES; PRESSURE; INFANTS;
D O I
10.1007/s00134-009-1579-0
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Monitoring end-expiratory lung volume (EELV) is a valuable tool to optimize respiratory settings that could be of particular importance in mechanically ventilated pediatric patients. We evaluated the feasibility and precision of an intensive care unit (ICU) ventilator with an in-built nitrogen washout/washin technique in mechanically ventilated pediatric patients. Duplicate EELV measurements were performed in 30 patients between 5 kg and 43 kg after cardiac surgery (age, median + range: 26, 3-141 months). All measurements were taken during pressure-controlled ventilation at 0 cm H2O of positive end-expiratory pressure (PEEP). Linear regression between duplicate measurements was excellent (R (2) = 0.99). Also, there was good agreement between duplicate measurements, bias +/- A SD: -0.3% (-1.5 mL) +/- A 5.9% (19.2 mL). Mean EELV +/- A SD was 19.6 +/- A 5.1 mL/kg at 0 cm H2O PEEP. EELV correlated with age (p < 0.001, r = 0.92, R (2) = 0.78), body weight (p < 0.001, r = 0.91, R (2) = 0.82) and height (p < 0.001, r = 0.94, R (2) = 0.75). This ICU ventilator with an in-built nitrogen washout/washin EELV technique can measure EELV with precision, and can easily be used for mechanically ventilated pediatric patients.
引用
收藏
页码:1749 / 1753
页数:5
相关论文
共 50 条
  • [1] Measurement of end-expiratory lung volume in intubated children without interruption of mechanical ventilation
    Ido G. Bikker
    Thierry V. Scohy
    Jan Ad J. J. C. Bogers
    Diederik Bakker
    Intensive Care Medicine, 2009, 35 : 1749 - 1753
  • [2] Measurement of the end-expiratory lung volume without interruption of mechanical ventilation in pediatric patients
    IG Bikker
    T Scohy
    D Gommers
    Critical Care, 13 (Suppl 1):
  • [3] The effect of flexible bronchoscopy on end-expiratory lung volumes in intubated patients on mechanical ventilation
    Shakespeare, E
    Won, C
    Matsubayashi, T
    Murashige, N
    Sundaram, S
    Mayo, P
    CHEST, 2003, 124 (04) : 78S - 78S
  • [4] Variations in end-expiratory pressure during partial liquid ventilation - Impact on gas exchange, lung compliance, and end-expiratory lung volume
    Manaligod, JM
    Bendel-Stenzel, EM
    Meyers, PA
    Bing, DR
    Connett, JE
    Mammel, MC
    CHEST, 2000, 117 (01) : 184 - 190
  • [5] End-expiratory lung volume during mechanical ventilation: a comparison with reference values and the effect of positive end-expiratory pressure in intensive care unit patients with different lung conditions
    Ido G Bikker
    Jasper van Bommel
    Dinis Reis Miranda
    Jan Bakker
    Diederik Gommers
    Critical Care, 12
  • [6] End-expiratory lung volume during mechanical ventilation: a comparison with reference values and the effect of positive end-expiratory pressure in intensive care unit patients with different lung conditions
    Bikker, Ido G.
    van Bommel, Jasper
    Miranda, Dinis Reis
    Bakker, Jan
    Gommers, Diederik
    CRITICAL CARE, 2008, 12 (06):
  • [7] The oxygenation ratio during mechanical ventilation in children: The role of tidal volume and positive end-expiratory pressure
    Halbertsma, Feico J. J.
    Vaneker, Michiel
    Pickkers, Peter
    Hoeven, Johannes G.
    JOURNAL OF CRITICAL CARE, 2009, 24 (02) : 220 - 226
  • [8] The effect of endotracheal suction on regional tidal ventilation and end-expiratory lung volume
    D. G. Tingay
    B. Copnell
    C. A. Grant
    P. A. Dargaville
    K. R. Dunster
    A. Schibler
    Intensive Care Medicine, 2010, 36 : 888 - 896
  • [9] The effect of endotracheal suction on regional tidal ventilation and end-expiratory lung volume
    Tingay, D. G.
    Copnell, B.
    Grant, C. A.
    Dargaville, P. A.
    Dunster, K. R.
    Schibler, A.
    INTENSIVE CARE MEDICINE, 2010, 36 (05) : 888 - 896
  • [10] End-expiratory lung impedance change enables bedside monitoring of end-expiratory lung volume change
    J. Hinz
    G. Hahn
    P. Neumann
    M. Sydow
    P. Mohrenweiser
    G. Hellige
    H. Burchardi
    Intensive Care Medicine, 2003, 29 : 37 - 43