Performance evaluation for pinhole collimators of small gamma camera by MTF and NNPS analysis: Monte Carlo simulation study

被引:3
|
作者
Jeon, Hosang [1 ]
Kim, Hyunduk [1 ]
Cha, Bo Kyung [1 ]
Kim, Jong Yul [1 ]
Cho, Gyuseong [1 ]
Chung, Yong Hyun [2 ]
Yun, Jong-Il [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Taejon 305701, South Korea
[2] Yonsei Univ, Dept Radiol Sci, Wonju, South Korea
关键词
MTF; NNPS; FWHM; Sensitivity; Uniformity; Gamma camera; MODULATION TRANSFER-FUNCTION;
D O I
10.1016/j.nima.2009.01.037
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Presently, the gamma camera system is widely used in various medical diagnostic, industrial and environmental fields. Hence, the quantitative and effective evaluation of its imaging performance is essential for design and quality assurance. The National Electrical Manufacturers Association (NEMA) standards for gamma camera evaluation are insufficient to perform sensitive evaluation. In this study, modulation transfer function (MTF) and normalized noise power spectrum (NNPS) will be suggested to evaluate the performance of small gamma camera with changeable pinhole collimators using Monte Carlo Simulation. We simulated the system with a cylinder and a disk source, and seven different pinhole collimators from 1- to 4-mm-diameter pinhole with lead. The MTF and NNPS data were obtained from output images and were compared with full-width at half-maximum (FWHM), sensitivity and differential uniformity. In the result, we found that MTF and NNPS are effective and novel standards to evaluate imaging performance of gamma cameras instead of conventional NEMA standards. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:93 / 96
页数:4
相关论文
共 50 条
  • [1] Monte-Carlo simulation of pinhole collimator of a small field of view gamma camera for small animal imaging
    Zhu Jie
    Ma Wenyan
    Zhu Yufeng
    Ma Hongguang
    Wu Yuelei
    Hu Huasi
    Zhang Boping
    Huo Yonggang
    Liu Silu
    Han Bin
    Wang Zhaomin
    NUCLEAR SCIENCE AND TECHNIQUES, 2009, 20 (01) : 37 - 41
  • [3] Monte Carlo study on image shrinkage of a small gamma camera
    Zeng, HN
    Xu, ZZ
    Wang, ZM
    Li, C
    Wang, CR
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2000, 24 (02): : 166 - 171
  • [4] Monte Carlo study on image shrinkage of a small gamma camera
    Zeng, Haining
    Xu, Zizong
    Wang, Zhaomin
    Li, Cheng
    Wang, Congrong
    Kao Neng Wu Li Yu Ho Wu Li/High Energy Physics and Nuclear Physics, 2000, 24 (02): : 166 - 171
  • [5] Evaluation of TlBr semiconductor detector in gamma camera imaging: Monte Carlo simulation study
    Lee, Youngjin
    Park, Chanrok
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2022, 54 (12) : 4652 - 4659
  • [6] Monte Carlo evaluation of a BGO-pinhole dedicated bremsstrahlung camera
    Hesse, M.
    Walrand, S.
    Demonceau, G.
    Lhommel, R.
    Jamar, F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S307 - S307
  • [7] Monte Carlo Simulation of Intrinsic Count Rate Performance for Scintillation Gamma Camera
    Mowlavi, A. A.
    de Denaro, M.
    Fornasier, M. R.
    Binesh, A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 209 - 211
  • [8] Gamma Camera Imaging with Rotating Multi-Pinhole Collimator. A Monte Carlo Feasibility Study
    Ilisie, Victor
    Moliner, Laura
    Morera, Constantino
    Nuyts, Johan
    Benlloch, Jose Maria
    SENSORS, 2021, 21 (10)
  • [9] Monte Carlo simulation of scatter effect for clinical gamma camera
    Saikouk, H.
    El Khayati, N.
    2014 MIDDLE EAST CONFERENCE ON BIOMEDICAL ENGINEERING (MECBME), 2014, : 305 - 308
  • [10] Optimization of pinhole collimator for small animal SPECT using Monte Carlo simulation
    Song, TY
    Choi, Y
    Chung, YH
    Jung, JH
    Choe, YS
    Lee, KH
    Kim, SE
    Kim, BT
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 811 - 815