Application of electrical resistance tomography for thrombus visualization in blood

被引:13
|
作者
Sapkota, Achyut [1 ]
Fuse, Takuto [2 ]
Seki, Masatoshi [2 ]
Maruyama, Osamu [3 ]
Sugawara, Michiko [2 ]
Takei, Masahiro [2 ]
机构
[1] Kisarazu Coll, Natl Inst Technol, Dept Informat & Comp Engn, Kisarazu, Chiba 292041, Japan
[2] Chiba Univ, Div Artificial Syst Sci, Chiba 2638522, Japan
[3] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058564, Japan
关键词
Electrical resistance tomography; Thrombus; Blood flow monitoring; Artificial organs; Extracorporeal circulation; FLOW VISUALIZATION; ASSIST DEVICES; IMPEDANCE; PUMP; COMPATIBILITY; SUSPENSION; VISCOSITY;
D O I
10.1016/j.flowmeasinst.2015.06.023
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Visualization of a thrombus is very important in the development of various artificial organs and extracorporeal circulation devices. This paper presents an application of electrical resistance tomography (ERT) technique for the visualization of a thrombus in blood. Experiments were conducted in static and flowing bovine and swine blood samples. Artificially created thrombi were mixed in the blood samples for visualization. Eight-electrode tomography sensor was used for the measurement. Cross-sectional resistivity distribution was reconstructed using linear back projection algorithm. A thrombus was characterized by increased local resistivity. We successfully reconstructed the time, size and cross-sectional location of a thrombus, and reached a conclusion that the concentration and orientation of the RBCs in a thrombus contributed to the increase in the resistivity. The increment was relatively higher in the static blood than in flowing blood. These findings can be helpful in the development of an instrumentation system for the real-time monitoring of blood to visualize a thrombus. Developers of left ventricular assistance devices, heart-lung machines, hemodialyzer etc., and the end-users (i.e. patients) can greatly benefit from such a system. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:334 / 340
页数:7
相关论文
共 50 条
  • [1] Application of Electrical Resistance Tomography
    Su, Chenhui
    Zhao, Jianyu
    Ren, Hongwei
    Qin, Lei
    Tian, Kaige
    PROCEEDINGS OF THE 4TH INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY AND MANAGEMENT INNOVATION, 2015, 28 : 450 - 457
  • [2] Visualization of Horizontal Settling Slurry Flow Using Electrical Resistance Tomography
    Li, Kun
    Wang, Mi
    Han, Yan
    PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON SENSING AND IMAGING, 2018, 2019, 606 : 65 - 75
  • [3] Research on damage visualization of concrete structures based on electrical resistance tomography
    Su, Chenhui
    Zhang, Wenchao
    Zhao, Jianyu
    Yang, Hongjuan
    Li, Chengdong
    FRONTIERS IN PHYSICS, 2022, 10
  • [4] Visualization investigation of defects in structural steel materials with electrical resistance tomography
    Zhao, Shuanfeng
    Wei, Zheng
    Xing, Zhizhong
    Wu, Yuyao
    Guo, Shuai
    Liu, Bo
    CONSTRUCTION AND BUILDING MATERIALS, 2022, 356
  • [5] Application of Electrical Resistance Tomography to Damage Detection in Concrete
    Zhao, Jianyu
    Su, Chenhui
    Ren, Hongwei
    Qin, Lei
    Dong, Biqin
    Xing, Feng
    PROCEEDINGS OF THE 28TH CHINESE CONTROL AND DECISION CONFERENCE (2016 CCDC), 2016, : 4746 - 4749
  • [6] Qualitative visualization and quantitative analysis of milk flow using electrical resistance tomography
    Sharifi, Mohadeseh
    Young, Brent
    JOURNAL OF FOOD ENGINEERING, 2012, 112 (03) : 227 - 242
  • [7] Electrical resistance tomography: A review of the application of conducting vessel walls
    Aw, Suzanna Ridzuan
    Rahim, Ruzairi Abdul
    Rahiman, Mohd Hafiz Fazalul
    Yunus, Fazlul Rahman Mohd
    Goh, Chiew Loon
    POWDER TECHNOLOGY, 2014, 254 : 256 - 264
  • [8] Visualization of asymmetric solids distribution in horizontal swirling flows using electrical resistance tomography
    Wang, M
    Jones, TF
    Williams, RA
    CHEMICAL ENGINEERING RESEARCH & DESIGN, 2003, 81 (A8): : 854 - 861
  • [9] Application of electrical resistance tomography to interrogate mixing processes at plant scale
    Mann, R
    Dickin, FJ
    Wang, M
    Dyakowski, T
    Williams, RA
    Edwards, RB
    Forrest, AE
    Holden, PJ
    CHEMICAL ENGINEERING SCIENCE, 1997, 52 (13) : 2087 - 2097
  • [10] Application of electrical resistance tomography to detect deposition in hydraulic conveying systems
    Fangary, YS
    Williams, RA
    Neil, WA
    Bond, J
    Faulks, I
    POWDER TECHNOLOGY, 1998, 95 (01) : 61 - 66