Square-wave voltammetric stripping analysis of thallium(III) at a poly(4-vinylpyridine)/mercury film electrode

被引:20
|
作者
Zen, JM
Wu, JW
机构
[1] Department of Chemistry, National Chung-Hsing University, Taichung
关键词
thallium; square-wave anodic stripping voltammetry; mercury film electrode; poly(4-vinylpyridine);
D O I
10.1002/elan.1140090408
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A poly(4-vinylpyridine)/mercury film electrode (PVP/MFE) was used for the determination of trace thallium(III) by square-wave anodic stripping voltammetry (SWASV). Thallium(III) is preconcentrated onto the PVP/MFE as the anionic forms in chloride medium by the ion-exchange effect of the PVP. The high solubility of thallium in mercury further facilitates the accumulation effect. Various factors influencing the determination of thallium(III) were thoroughly investigated. This modified electrode displayed good resistance to interferences from surface-active compounds and common ions and increased sensitivity when used in conjunction with SWASV. In addition, detection can be achieved without deoxygenation and the electrode can be easily renewed. Applicability to various water samples is illustrated.