A Novel Respiratory Motion Perturbation Model Adaptable to Patient Breathing Irregularities

被引:10
|
作者
Yuan, Amy [1 ]
Wei, Jie [2 ]
Gaebler, Carl P. [1 ]
Huang, Hailiang [1 ]
Olek, Devin [1 ]
Li, Guang [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, 1275 York Ave, New York, NY 10065 USA
[2] CUNY City Coll, Dept Comp Sci, New York, NY 10031 USA
基金
美国国家卫生研究院;
关键词
BODY RADIATION-THERAPY; CELL LUNG-CANCER; PANCREATIC-CANCER; DIAPHRAGM MOTION; TUMOR TRACKING; MRI; RADIOTHERAPY; ACCURACY; FEASIBILITY; PREDICTION;
D O I
10.1016/j.ijrobp.2016.08.044
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop a physical, adaptive motion perturbation model to predict tumor motion using feedback from dynamic measurement of breathing conditions to compensate for breathing irregularities. Methods and Materials: A novel respiratory motion perturbation (RMP) model was developed to predict tumor motion variations caused by breathing irregularities. This model contained 2 terms: the initial tumor motion trajectory, measured from 4-dimensional computed tomography (4DCT) images, and motion perturbation, calculated from breathing variations in tidal volume (TV) and breathing pattern (BP). The motion perturbation was derived from the patient-specific anatomy, tumor-specific location, and time-dependent breathing variations. Ten patients were studied, and 2 amplitude-binned 4DCT images for each patient were acquired within 2 weeks. The motion trajectories of 40 corresponding bifurcation points in both 4DCT images of each patient were obtained using deformable image registration. An in-house 4D data processing toolbox was developed to calculate the TV and BP as functions of the breathing phase. The motion was predicted from the simulation 4DCT scan to the treatment 4DCT scan, and vice versa, resulting in 800 predictions. For comparison, noncorrected motion differences and the predictions from a published 5-dimensional model were used. Results: The average motion range in the superoinferior direction was 9.4 +/- 4.4 mm, the average DTV ranged from 10 to 248 mm 3 (-26% to 61%), and the DBP ranged from 0 to 0.2 (-71% to 333%) between the 2 4DCT scans. The mean noncorrected motion difference was 2.0 +/- 2.8 mm between 2 4DCT motion trajectories. After applying the RMP model, the mean motion difference was reduced significantly to 1.2 +/- 1.8 mm (P=.0018), a 40% improvement, similar to the 1.2 +/- 1.8 mm (P=.72) predicted with the 5-dimensional model. Conclusions: A novel physical RMP model was developed with an average accuracy of 1.2 +/- 1.8 mm for interfraction motion prediction, similar to that of a published lung motion model. This physical RMP was analytically derived and is able to adapt to breathing irregularities. Further improvement of this RMP model is under investigation. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:1087 / 1096
页数:10
相关论文
共 50 条
  • [1] Novel breathing motion model for radiotherapy
    Low, DA
    Parikh, PJ
    Lu, W
    Dempsey, JF
    Wahab, SH
    Hubenschmidt, JP
    Nystrom, MM
    Handoko, M
    Bradley, JD
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (03): : 921 - 929
  • [2] Novel prospective respiratory motion correction approach for free-breathing coronary MR angiography using a patient-adapted affine motion model
    Manke, D
    Nehrke, K
    Börnert, P
    MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (01) : 122 - 131
  • [3] An Analytical Respiratory Perturbation Model for Lung Motion Prediction
    Li, G.
    Yuan, A.
    Wei, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 473 - 473
  • [4] Patient-derived breathing patterns for a hardware respiratory motion simulator
    Wong, K. H.
    Dieterich, S.
    Tang, J.
    Cleary, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S619 - S619
  • [5] Investigation of a Respiratory Motion Model Based on an External Surrogate Signal With a Digital Phantom and Realistic Patient Breathing
    Hurwitz, M.
    Williams, C. L.
    Mishra, P.
    Dhou, S.
    Lewis, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S846 - S846
  • [6] Free Breathing Motion Model: First Report of Patient Model Parameters
    Low, D.
    Zhao, T.
    Yang, D.
    Lu, W.
    Mutic, S.
    Noel, C.
    Hubenschmidt, J.
    Parikh, P.
    Bradley, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S23 - S23
  • [7] A Theoretical Model for Respiratory Motion Artifacts in Free Breathing CT Scans
    Lewis, J.
    Jiang, S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [8] Characterization of continuous bed motion effects on patient breathing and respiratory motion correction in PET/CT imaging
    Meier, Joseph G.
    Erasmus, Jeremy J.
    Gladish, Gregory W.
    Peterson, Christine B.
    Diab, Radwan H.
    Mawlawi, Osama R.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (01): : 158 - 165
  • [9] Effect of breathing irregularities on the quantitative accuracy of respiratory gated PET/CT
    Teo, Boon-Keng
    Saboury, Babak
    Scheuermann, Joshua
    Torigian, Drew
    Karp, Joel
    Alavi, Abass
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [10] The effect of breathing irregularities on quantitative accuracy of respiratory gated PET/CT
    Teo, Boon-Keng
    Saboury, Babak
    Munbodh, Reshma
    Scheuermann, Joshua
    Torigian, Drew A.
    Zaidi, Habib
    Alavi, Abass
    MEDICAL PHYSICS, 2012, 39 (12) : 7390 - 7397