Can you trust your cryostat? Reproducibility of cryostat section thickness

被引:5
|
作者
Hamer, Philip C. De Witt
Bleeker, Fonnet E.
Zwinderman, Aeilko H.
Van Noorden, Cornelis J. F.
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Neurosurg, NL-1100 DD Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Clin Epidemiol & Biostat, NL-1105 AZ Amsterdam, Netherlands
[3] Univ Amsterdam, Acad Med Ctr, Dept Cell Biol & Histol, NL-1105 AZ Amsterdam, Netherlands
关键词
reproducibility; section thickness; cryostat; cryomicrotome; frozen sections; quantitative microscopy; lactate dehydrogenase activity;
D O I
10.1002/jemt.20354
中图分类号
R602 [外科病理学、解剖学]; R32 [人体形态学];
学科分类号
100101 ;
摘要
Reproducibility of cryostat section thickness is required for valid quantitative microscopy. This is generally pursued by motorized sectioning using a low but constant speed. The purpose of our study was to compare variation in section thickness between motorized and manual cryostat sectioning. Serial sections were cut from a frozen block of homogenized tissue on different days. Lactate dehydrogenase activity was histochemically detected and calibrated absorbance measurements were taken. The coefficients of variation of measurements was 9.7% for motorized sectioning and 3.3% for manual sectioning. In conclusion, section thickness is similarly reproducible after manual sectioning compared with motorized sectioning, if not better.
引用
收藏
页码:835 / 838
页数:4