Image distortion correction for MRI in low field permanent magnet systems with strong B0 inhomogeneity and gradient field nonlinearities

被引:27
|
作者
Koolstra, Kirsten [1 ]
O'Reilly, Thomas [2 ]
Bornert, Peter [2 ,3 ]
Webb, Andrew [2 ]
机构
[1] Leiden Univ, Div Image Proc, Radiol, Med Ctr, Albinusdreef 2, NL-2333 ZA Leiden, Netherlands
[2] Leiden Univ, Med Ctr, CJ Gorter Ctr High Field MRI, Radiol, Albinusdreef 2, NL-2333 ZA Leiden, Netherlands
[3] Philips Res, Rontgenstr 24-26, D-22335 Hamburg, Germany
基金
欧洲研究理事会;
关键词
Distortion correction; B-0; mapping; Conjugate phase reconstruction; Model-based reconstruction; Low field MRI;
D O I
10.1007/s10334-021-00907-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To correct for image distortions produced by standard Fourier reconstruction techniques on low field permanent magnet MRI systems with strong B-0 inhomogeneity and gradient field nonlinearities. Materials and methods Conventional image distortion correction algorithms require accurate Delta B-0 maps which are not possible to acquire directly when the B-0 inhomogeneities also produce significant image distortions. Here we use a readout gradient time-shift in a TSE sequence to encode the B-0 field inhomogeneities in the k-space signals. Using a non-shifted and a shifted acquisition as input, Delta B-0 maps and images were reconstructed in an iterative manner. In each iteration, Delta B-0 maps were reconstructed from the phase difference using Tikhonov regularization, while images were reconstructed using either conjugate phase reconstruction (CPR) or model-based (MB) image reconstruction, taking the reconstructed field map into account. MB reconstructions were, furthermore, combined with compressed sensing (CS) to show the flexibility of this approach towards undersampling. These methods were compared to the standard fast Fourier transform (FFT) image reconstruction approach in simulations and measurements. Distortions due to gradient nonlinearities were corrected in CPR and MB using simulated gradient maps. Results Simulation results show that for moderate field inhomogeneities and gradient nonlinearities, Delta B-0 maps and images reconstructed using iterative CPR result in comparable quality to that for iterative MB reconstructions. However, for stronger inhomogeneities, iterative MB reconstruction outperforms iterative CPR in terms of signal intensity correction. Combining MB with CS, similar image and Delta B-0 map quality can be obtained without a scan time penalty. These findings were confirmed by experimental results. Discussion In case of B-0 inhomogeneities in the order of kHz, iterative MB reconstructions can help to improve both image quality and Delta B-0 map estimation.
引用
收藏
页码:631 / 642
页数:12
相关论文
共 50 条
  • [1] Image distortion correction for MRI in low field permanent magnet systems with strong B0 inhomogeneity and gradient field nonlinearities
    Kirsten Koolstra
    Thomas O’Reilly
    Peter Börnert
    Andrew Webb
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2021, 34 : 631 - 642
  • [2] Iterative correction of subject-dependent B0 inhomogeneity field maps for geometric distortion correction
    Matakos, A.
    Balter, J.
    Cao, Y.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S33 - S34
  • [3] Novel Method for Geometrically Undistorted B0 Inhomogeneity Field Map Estimation and Image Correction
    Matakos, A.
    Balter, J.
    Cao, Y.
    MEDICAL PHYSICS, 2014, 41 (06) : 528 - 528
  • [4] CORRECTION OF B0 INHOMOGENEITY DISTORTION IN MAGNETIC RESONANCE SPECTROSCOPIC IMAGING
    Eslami, Ramin
    Jacob, Mathews
    2009 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH, AND SIGNAL PROCESSING, VOLS 1- 8, PROCEEDINGS, 2009, : 405 - 408
  • [5] A new gradient shimming method based on undistorted field map of B0 inhomogeneity
    Bao, Qingjia
    Chen, Fang
    Chen, Li
    Song, Kan
    Liu, Zao
    Liu, Chaoyang
    JOURNAL OF MAGNETIC RESONANCE, 2016, 265 : 25 - 32
  • [6] B0 concomitant field compensation for MRI systems employing asymmetric transverse gradient coils
    Weavers, Paul T.
    Tao, Shengzhen
    Trzasko, Joshua D.
    Frigo, Louis M.
    Shu, Yunhong
    Frick, Matthew A.
    Lee, Seung-Kyun
    Foo, Thomas K-F
    Bernstein, Matt A.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (03) : 1538 - 1544
  • [7] Joint B0 and image estimation integrated with model based reconstruction for field map update and distortion correction in prostate diffusion MRI
    Usman, Muhammad
    Kakkar, Lebina
    Matakos, Antonis
    Kirkham, Alex
    Arridge, Simon
    Atkinson, David
    MAGNETIC RESONANCE IMAGING, 2020, 65 : 90 - 99
  • [8] Comparison of B0 versus B0 and B1 field inhomogeneity correction for glycosaminoglycan chemical exchange saturation transfer imaging
    Mueller-Lutz, Anja
    Ljimani, Alexandra
    Stabinska, Julia
    Zaiss, Moritz
    Boos, Johannes
    Wittsack, Hans-Joerg
    Schleich, Christoph
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2018, 31 (05) : 645 - 651
  • [9] Comparison of B0 versus B0 and B1 field inhomogeneity correction for glycosaminoglycan chemical exchange saturation transfer imaging
    Anja Müller-Lutz
    Alexandra Ljimani
    Julia Stabinska
    Moritz Zaiss
    Johannes Boos
    Hans-Jörg Wittsack
    Christoph Schleich
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2018, 31 : 645 - 651
  • [10] 3D gradient system for two B0 field directions in earth’s field MRI
    Steffen Lother
    Uvo Hoelscher
    Thomas Kampf
    Peter Jakob
    Florian Fidler
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2013, 26 : 565 - 573