Finger arterial pressure measurement with Finapres

被引:0
|
作者
Wesseling, KH
机构
来源
关键词
finger arterial pressure; pulse waveform distortion; distortion correction; generalized inverse filter; automatic level correction; quality control points;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Finger arterial pressure measurement with Finapres has been available since a decade. Its availability has promoted at least 300 methodological and research papers over these years, outlining the usefulness and the limitations of the method and the device. Finapres is based on the volume clamp method of Penaz and the Physiocal criteria of Wesseling. Tracking of intraarterial pressure is usually satisfactory even under conditions of strongly changing hemodynamics and high and very low blood pressures. Finapres accuracy is similar to that of other noninvasive methods. Systolic pressure levels scatter more than mean and diastolic levels. One source of error is physiologic and determined by the peripheral measurement site of the finger, causing pulse waveform distortion and a pressure gradient. The Finapres waveform can be filtered, however, to obtain a brachial pressure wave. This decreases systolic scatter under vase-constrictive drug infusion and dynamic exercise to exhaustion, conditions where precision of systolic tracking has been criticized in the literature. Recently, level correction techniques were found which shift finger pressure up or down based on a regression equation with finger systolic and diastolic pressures. This procedure requires no additional measurements yet improves systolic, diastolic and mean level accuracy and precision remarkably. Finally, we show how to judge the quality of a Finapres recording from the behavior of Physiocal.
引用
收藏
页码:38 / 44
页数:7
相关论文
共 50 条
  • [1] NONINVASIVE, CONTINUOUS MEASUREMENT OF THE FINGER ARTERIAL-PRESSURE WITH THE SERVO-PLETHYSMO-MANOMETER FINAPRES
    HARTMANN, B
    BASSENGE, E
    HERZ, 1989, 14 (04) : 251 - 259
  • [2] MEASUREMENT OF BLOOD-PRESSURE IN OBESE PERSONS - COMPARISON OF INTRAARTERIAL AND FINGER ARTERIAL BLOOD-PRESSURE (FINAPRES)
    PONCELET, P
    DURAND, P
    LECHANTRE, R
    MOUNIERVEHIER, C
    FOUQUOIRE, B
    PETETIN, N
    CARRE, A
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1992, 85 (08): : 1165 - 1168
  • [3] NONINVASIVE MEASUREMENT OF SYSTEMIC ARTERIAL-PRESSURE BY FINAPRES IN ANESTHETIZED DOGS
    TABARU, H
    WATANABE, H
    TANAKA, M
    NAKAMA, S
    JAPANESE JOURNAL OF VETERINARY SCIENCE, 1990, 52 (02): : 427 - 430
  • [4] ACCURACY OF BEAT-TO-BEAT NONINVASIVE MEASUREMENT OF FINGER ARTERIAL-PRESSURE USING THE FINAPRES - A SPECTRAL-ANALYSIS APPROACH
    NOVAK, V
    NOVAK, P
    SCHONDORF, R
    JOURNAL OF CLINICAL MONITORING, 1994, 10 (02): : 118 - 126
  • [5] A CENTURY OF NONINVASIVE ARTERIAL-PRESSURE MEASUREMENT - FROM MAREY TO PENAZ AND FINAPRES
    WESSELING, KH
    HOMEOSTASIS IN HEALTH AND DISEASE, 1995, 36 (2-3): : 50 - 65
  • [6] COMPARISON OF INVASIVE AND NONINVASIVE MEASUREMENT OF CONTINUOUS ARTERIAL-PRESSURE USING THE FINAPRES
    STOKES, DN
    CLUTTONBROCK, T
    PATIL, C
    THOMPSON, JM
    HUTTON, P
    BRITISH JOURNAL OF ANAESTHESIA, 1991, 67 (01) : 26 - 35
  • [7] ARTERIAL-PRESSURE MEASUREMENT IN THE CRITICALLY ILL - COMPARISON OF FINAPRES AND INTRA-ARTERIAL CANNULA
    FARQUHAR, I
    EVANS, D
    BRITISH JOURNAL OF ANAESTHESIA, 1989, 63 (05) : P619 - P620
  • [8] EFFECT OF HYDROSTATIC-PRESSURE ON FINGER PRESSURE MEASURED NONINVASIVELY BY FINAPRES(TM)
    GIZDULICH, P
    ASCHERO, G
    GUERRISI, M
    WESSELING, KH
    HOMEOSTASIS IN HEALTH AND DISEASE, 1995, 36 (2-3): : 120 - 129
  • [9] THE FINAPRES 2300E FINGER CUFF - THE INFLUENCE OF CUFF APPLICATION ON THE ACCURACY OF BLOOD-PRESSURE MEASUREMENT
    JONES, RDM
    KORNBERG, JP
    ROULSON, CJ
    VISRAM, AR
    IRWIN, MG
    ANAESTHESIA, 1993, 48 (07) : 611 - 615
  • [10] Relationship between the ascending aortic pressure waveform and the pressure wave determined in the finger by Finapres
    Plunkett, BT
    Lehmann, ED
    Kelly, RP
    Hayward, CS
    Avolio, AP
    ORourke, MF
    ZEITSCHRIFT FUR KARDIOLOGIE, 1996, 85 : 136 - 136