Biomimetic collagen/elastin meshes for ventral hernia repair in a rat model

被引:40
|
作者
Minardi, Silvia [1 ,2 ]
Taraballi, Francesca [1 ]
Wang, Xin [1 ]
Cabrera, Fernando J. [1 ]
Van Eps, Jeffrey L. [1 ]
Robbins, Andrew B. [3 ]
Sandri, Monica [2 ]
Moreno, Michael R. [1 ,3 ,4 ,5 ]
Weiner, Bradley K. [1 ,5 ]
Tasciotti, Ennio [1 ,5 ]
机构
[1] HMRI, Ctr Biomimet Med, 6670 Bertner Ave,Suite R10-116, Houston, TX 77030 USA
[2] Natl Res Council Italy, Inst Sci & Technol Ceram ISTEC CNR, Via Granarolo 64, I-48018 Faenza, RA, Italy
[3] Texas A&M Univ, Dept Biomed Engn, 401 Joe Routt Blvd, College Stn, TX 77843 USA
[4] Texas A&M Univ, Dept Mech Engn, 3123 TAMU, College Stn, TX 77843 USA
[5] Houston Methodist Hosp, Dept Orthoped, 6565 Fannin St, Houston, TX 77030 USA
关键词
Hernia; Collagen; Elastin; Biomimicry; Meshes; Tissue engineering; CALCIUM-PHOSPHATE CEMENT; MECHANICAL-PROPERTIES; INCISIONAL HERNIAS; BIOLOGICAL MESHES; PORCINE MODEL; SCAFFOLDS; ELASTIN; STIFFNESS; FASCIA; FIBERS;
D O I
10.1016/j.actbio.2016.11.032
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ventral hernia repair remains a major clinical need. Herein, we formulated a type I collagenielastin cross linked blend (CollE) for the fabrication of biomimetic meshes for ventral hernia repair. To evaluate the effect of architecture on the performance of the implants, CollE was formulated both as flat sheets (CollE Sheets) and porous scaffolds (CollE Scaffolds). The morphology, hydrophylicity and in vitro degradation were assessed by SEM, water contact angle and differential scanning calorimetry, respectively. The stiffness of the meshes was determined using a constant stretch rate uniaxial tensile test, and compared to that of native tissue. CollE Sheets and Scaffolds were tested in vitro with human bone marrow-derived mesenchymal stem cells (h-BM-MSC), and finally implanted in a rat ventral hernia model. Neovascularization and tissue regeneration within the implants was evaluated at 6 weeks, by histology, immunofluorescence, and q-PCR. It was found that CollE Sheets and Scaffolds were not only biomechanically sturdy enough to provide immediate repair of the hernia defect, but also promoted tissue restoration in only 6 weeks. In fact, the presence of elastin enhanced the neovascularization in both sheets and scaffolds. Overall, CollE Scaffolds displayed mechanical properties more closely resembling those of native tissue, and induced higher gene expression of the entire marker genes tested, associated with de novo matrix deposition, angiogenesis, adipogenesis and skeletal muscles, compared to CollE Sheets. Altogether, this data suggests that the improved mechanical properties and bioactivity of CollE Sheets and Scaffolds make them valuable candidates for applications of ventral hernia repair. Statement of Significance Due to the elevated annual number of ventral hernia repair in the US, the lack of successful grafts, the design of innovative biomimetic meshes has become a prime focus in tissue engineering, to promote the repair of the abdominal wall, avoid recurrence. Our meshes (CollE Sheets and Scaffolds) not only showed promising mechanical performance, but also allowed for an efficient neovascularization, resulting in new adipose and muscle tissue formation within the implant, in only 6 weeks. In addition, our meshes allowed for the use of the same surgical procedure utilized in clinical practice, with the commercially available grafts. This study represents a significant step in the design of bioactive acellular off-the shelf biomimetic meshes for ventral hernia repair. (C) 2016 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:165 / 177
页数:13
相关论文
共 50 条
  • [1] Evaluation of resorbable polydioxanone and polyglycolic acid meshes in a rat model of ventral hernia repair
    Fatkhudinov, Timur
    Tsedik, Larisa
    Arutyunyan, Irina
    Lokhonina, Anastasia
    Makarov, Andrey
    Korshunov, Aleksey
    Elchaninov, Andrey
    Kananykhina, Evgeniya
    Vasyukova, Olesya
    Usman, Natalia
    Uvarova, Elena
    Chuprynin, Vladimir
    Eremina, Irina
    Degtyarev, Dmitry
    Sukhikh, Gennady
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2019, 107 (03) : 652 - 663
  • [2] Dermal collagen matrices for ventral hernia repair: comparative analysis in a rat model
    G. Broderick
    J. McIntyre
    M. Noury
    H. M. Strom
    C. Psoinos
    A. Christakas
    K. Billiar
    Z. M. Hurwitz
    J. F. Lalikos
    R. A. Ignotz
    R. M. Dunn
    Hernia, 2012, 16 : 333 - 343
  • [3] Dermal collagen matrices for ventral hernia repair: comparative analysis in a rat model
    Broderick, G.
    McIntyre, J.
    Noury, M.
    Strom, H. M.
    Psoinos, C.
    Christakas, A.
    Billiar, K.
    Hurwitz, Z. M.
    Lalikos, J. F.
    Ignotz, R. A.
    Dunn, R. M.
    HERNIA, 2012, 16 (03) : 333 - 343
  • [4] Testing meshes in a computer model of a laparoscopic ventral hernia repair
    Haitham Qandeel
    Cindy Chew
    K. E. Tanner
    Patrick J. O’Dwyer
    Surgical Endoscopy, 2022, 36 : 4124 - 4128
  • [5] Testing meshes in a computer model of a laparoscopic ventral hernia repair
    Qandeel, Haitham
    Chew, Cindy
    Tanner, K. E.
    O'Dwyer, Patrick J.
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2022, 36 (06): : 4124 - 4128
  • [6] Evaluation of Biologic Meshes in a Porcine Model of Ventral Incisional Hernia Repair
    Delaney, John P.
    Gaertner, Wolfgang B.
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2011, 213 (05) : 691 - 691
  • [7] Ventral hernia repair in rat using nanofibrous polylactic acid/polypropylene meshes
    Ye Xiaolong
    Han Xiaoyan
    Wei Bo
    Huang Jianglong
    Yang Xiaofeng
    Tang Xiao
    Zheng Zongheng
    Luo Linbo
    Zhan Zefeng
    Wei Hongbo
    NANOMEDICINE, 2018, 13 (17) : 2187 - 2200
  • [8] Evaluation of new prosthetic meshes for ventral hernia repair
    J. W. A. Burger
    J. A. Halm
    A. R. Wijsmuller
    S. ten Raa
    J. Jeekel
    Surgical Endoscopy And Other Interventional Techniques, 2006, 20
  • [9] Evaluation of new prosthetic meshes for ventral hernia repair
    Burger, J. W. A.
    Halm, J. A.
    Wijsmuller, A. R.
    ten Raa, S.
    Jeekel, J.
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (08): : 1320 - 1325
  • [10] Evaluation of Biologic Meshes in a Porcine Model of Ventral Incisional Hernia Repair Reply
    Deeken, Corey R.
    Melman, Lora
    Jenkins, Eric D.
    Greco, Suellen C.
    Frisella, Margaret M.
    Matthews, Brent D.
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2011, 213 (05) : 692 - 692